The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04953170050992), received between Jan 24, 2020 and Apr 29, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Adverse Event Without Identified Device or Use Problem1
- Material Frayed1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Injury1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO LARYNGOFIBERSCOPE COATING ON THE CONNECTING TUBE WAS PEELING OFF. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25031938 ↗MalfunctionDec 13, 2022Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE INSERTION PART WAS PEELING ON THE RHINO-LARYNGOFIBERSCOPE. INSPECTION AND TESTING OF THE RETURNED DEVICE FOUND THE RESIN PART OF THE INSERTION TUBE HAD FALLEN OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15972799 ↗MalfunctionSep 14, 2022Material Frayed; Material Frayed
THE CUSTOMER REPORTED TO OLYMPUS, THE LINING ON THE INSIDE CHANNEL OF THE RHINO-LARYNGOFIBERSCOPE WAS FRAYING/UNRAVELING. THE FRAYED PORTION WAS BLOCKING PART OF THE SCOPE IMAGE. THE ISSUE WAS DETECTED DURING REPROCESSING POST PROCEDURE. THE INTENDED LARYNGOSCOPY WITH INJECTION OF KENALOG PROCEDURE WAS COMPLETED WITH THE SAME DEVICE AND CHANNEL NEEDLE. THE DEVICE HAD BEEN INSPECTED PRIOR TO USE WITH NO ABNORMALITIES NOTED. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15416913 ↗MalfunctionAug 30, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
CUSTOMER RETURNED THE DEVICE FOR EVALUATION AND REPAIR OF AN IMPROPER IMAGE ISSUE OF THE DEVICE. THERE IS NO REPORTED HARM TO ANY PATIENT OR PERSONS. UPON EVALUATION OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THE IMAGE GUIDE HOB HAD FALLEN OFF. THIS MEDWATCH IS BEING SUBMITTED FOR THE REPORTABLE ISSUE OF THE IMAGE GUIDE HOB FALLING OFF AS OBSERVED DURING DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15322394 ↗InjuryJan 24, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING A POLYPECTOMY OF THE LEFT VOCAL CORD USING THE SUBJECT DEVICE, THE PATIENT HAD A NOSE BLEEDING BECAUSE THE FOLDS AND PROTRUSION OF THE SUBJECT DEVICE SUCH AS THE LENS DAMAGED THE PATIENT'S NASAL MUCOSA. THE PATIENT COMPLAINED OF NASAL PAIN AND DISCOMFORT AFTER THE PROCEDURE. THERE WAS NO REPORT OF ADDITIONAL TREATMENT WAS PERFORMED FOR THE PATIENT. THERE WAS NO REPORT OF FURTHER PATIENT INJURY WITH THE EVENT.
Patient problems: Injury
MAUDE report 9626653 ↗
Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)785 reports
- NA 242268Medos International Sàrl · Nasopharyngoscope (flexible or rigid)573 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Olympus Medical Systems Corp. · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- VIVIDEO VNL9-CPHoya Corporation · Nasopharyngoscope (Flexible Or Rigid)389 reports
- PENTAX FNL-10RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)322 reports
- SINGLE USE BIOPSY VALVE MAJ-210Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)1 recall221 reports
- SINGLE USE SUCTION VALVE MAJ-209Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)228 reports
- PENTAX FNL-7RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)227 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page