OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE N20Recalls & reported problems
Olympus Medical Systems Corp. · BF-N20 2.2MM DIA W/O CH 55CM LENGTH · Class 2
All models of OES BRONCHOFIBERSCOPE (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04953170051142), received between Sep 14, 2020 and Jul 1, 2024 — classified as 4 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 27, 2023Class IIOpen, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2502-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Contamination /Decontamination Problem2
- Mechanical Problem2
- Adverse Event Without Identified Device or Use Problem1
- Component Missing1
- Contamination1
- Detachment of Device or Device Component1
Patient problems
- Fungal Infection3
- No Clinical Signs, Symptoms or Conditions2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Patient Involvement1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2024Mechanical Problem; Mechanical Problem
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOFIBERSCOPE EXHIBITED THE DOWN ANGULATION NOT WORKING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19643002 ↗MalfunctionAug 15, 2022Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
CUSTOMER REPORTED WITH AN ISSUE OF AIR/WATER LEAKAGES. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT. DEVICE EVALUATION FOUND THERE IS A MISSING PART (DISTAL END ) OF THE SCOPE. THIS REPORT IS BEING SUBMITTED FOR MISSING PART OF THE SCOPE (DISTAL END) WHICH WAS NOTED DURING DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15229379 ↗InjurySep 10, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER
Patient problems: Fungal Infection
MAUDE report 12455473 ↗InjurySep 10, 2021Adverse Event Without Identified Device or Use Problem
IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER
Patient problems: Fungal Infection
MAUDE report 12455467 ↗InjurySep 10, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER
Patient problems: Fungal Infection
MAUDE report 12454134 ↗InjurySep 10, 2021Contamination
IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12454092 ↗MalfunctionSep 14, 2020Component Missing; Component Missing
IT WAS REPORTED THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AS PER THE CUSTOMER IT WAS IDENTIFIED THAT DURING REPROCESSING THE FIBER OPTICS ARE BAD [SIC]. DURING ESTIMATION THE DISTAL SHEATH OF THE BENDING SECTION WAS IDENTIFIED TO BE MISSING.
Patient problems: No Patient Involvement
MAUDE report 10531733 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page