RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF TYPE XP (UDI 04953170059018)Recalls & reported problems

Olympus Medical Systems Corp. · ENF-XP RHINOLARYNNGOSCOPE · Class 2

All models of RHINO-LARYNGOFIBERSCOPE (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus rhinolaryngoscope is a flexible scope used to examine the nose and throat. The 26 reports mostly describe parts detaching or breaking, peeling coatings, and damaged or missing lens covers, with no deaths or injuries reported. There is 1 open recall because Olympus updated cleaning instructions, removing reusable brushes while keeping single-use brushes compatible. Reports rose from 3 to 13 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 04953170059018), received between Jan 15, 2020 and Sep 18, 2026 — classified as 29 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 4, 2023Class IIOpen, Classified

    Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.

    Z-0688-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component7
  • Device Reprocessing Problem4
  • Break3
  • Mechanical Problem3
  • Peeled/Delaminated3
  • Component Missing2
  • Corroded2
  • Failure to Clean Adequately2

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2026Detachment of Device or Device Component

    The customer reported there was an unprotected Distal Tip during reprocessing. The issue involved an Olympus Rhino Laryngofiberscope. Additionally, during the initial device analysis, the Olympus Service Repair Technician indicated that the Insertion Section Distal rubber was detached. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50818804 ↗
  • MalfunctionSep 11, 2026Device Reprocessing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50737639 ↗
  • MalfunctionSep 4, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50623729 ↗
  • MalfunctionAug 31, 2026Peeled/Delaminated

    It was observed that during the device evaluation, the rhino-laryngofiberscope insertion tube coating was peeled off. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50525103 ↗
  • MalfunctionApr 29, 2026Device Reprocessing Problem

    NO NEW INFORMATION HAS BEEN PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25037730 ↗
  • MalfunctionApr 28, 2026Component Missing

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED THE LIGHT GUIDE COVER GLASS MISSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25022986 ↗
  • MalfunctionApr 7, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED THE CONNECTOR ¿ LIGHT GUIDE COVER GLASS WAS DETACHED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24800092 ↗
  • MalfunctionMar 24, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED A CHIP ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24673443 ↗
  • MalfunctionFeb 24, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO LARYNGOFIBERSCOPE HAD A MISSING COVER GLASS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428811 ↗
  • MalfunctionJan 2, 2026Device Reprocessing Problem; Degraded

    NO NEW INFORMATION HAS BEEN PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23952201 ↗
  • MalfunctionDec 30, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED MISSING ADHESIVE ON THE DISTAL END, THE PLASTIC DISTAL END, THE OBJECTIVE LENS, AND THE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23926732 ↗
  • MalfunctionDec 18, 2025Corroded; Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED CORROSION INSIDE THE LENS OR THE LIGHT GUIDE PRONG. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23839408 ↗
  • MalfunctionDec 8, 2025Component Missing; Component Missing

    NO NEW INFORMATION HAS BEEN PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23742130 ↗
  • MalfunctionNov 19, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE LARYNGOFIBERSCOPE EXHIBITED DIRT ON THE IMAGE GUIDE COVER LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23597993 ↗
  • MalfunctionNov 11, 2025Unstable; Unstable

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE'S LIGHT GUIDE COVER GLASS WAS LOOSE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23526912 ↗
  • MalfunctionNov 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE SUBJECT DEVICE DISTAL END SHEATH DISCONNECTED FROM THE BENDING ARM. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23524783 ↗
  • MalfunctionOct 24, 2025Unstable; Unstable

    IT WAS REPORTED THAT THE RHINO-LARYNGO VIDEOSCOPE DISTAL HEAD CONTROL POSITION WAS LOOSE. THE ISSUE OCCURRED DURING SERVICE OR MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23380179 ↗
  • MalfunctionOct 16, 2025Break; Detachment of Device or Device Component; Material Twisted/Bent; Unintended Movement; Break; Detachment of Device or Device Component; Material Twisted/Bent; Unintended Movement

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE TIP WAS ROTATED AND DETACHED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23311506 ↗
  • MalfunctionJan 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE SCOPE OF THE RHINO-LARYNGO FIBERSCOPE FELL APART DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21072562 ↗
  • MalfunctionOct 31, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SUBJECT DEVICE EXHIBITED FOREIGN MATERIAL ON THE CURVED RUBBER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20579670 ↗
  • MalfunctionJun 26, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED, THE FIBERSCOPE'S ANGULATION BECAME LOCKED- CANNOT DISENGAGE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19619006 ↗
  • MalfunctionDec 6, 2023Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO OLYMPUS, THE RHINO-LARYNGO FIBERSCOPE HAS LOCKED TIE ROD. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18280223 ↗
  • MalfunctionJul 28, 2023Peeled/Delaminated; Peeled/Delaminated

    THE CUSTOMER REPORTED TO OLYMPUS MEDICAL THAT THE RHINO-LARYNGO FIBERSCOPE FAILED THE LEAK TEST DUE TO THE DISTAL TIP PEELING OFF. NO ADVERSE EFFECTS TO PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17417323 ↗
  • MalfunctionJul 28, 2023Peeled/Delaminated; Peeled/Delaminated

    THE CUSTOMER REPORTED TO OLYMPUS MEDICAL THAT THE RHINO-LARYNGO FIBERSCOPE FAILED THE LEAK TEST DUE TO THE DISTAL TIP PEELING OFF. NO ADVERSE EFFECTS TO PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17417281 ↗
  • MalfunctionApr 28, 2023Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE RHINO-LARYNGO FIBERSCOPE EXPERIENCED AN ANGULATION MALFUNCTION. THE EVENT OCCURRED DURING PREPARATION FOR USE. THE INTENDED PROCEDURE WAS LISTED AS DIAGNOSTIC. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF HEALTH HAZARD TO THE PATIENT OR USER OF THE DEVICE. THE SUBJECT DEVICE WAS SUBSEQUENTLY EVALUATED AND CONFIRMED BY OLYMPUS, AND DETECTED THE ANGLE MIGHT BE LOCKED DUE TO CUT ANGULATION WIRE. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16832811 ↗

Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices

All NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices →