BRONCHOFIBERSCOPE BF-PE2 (UDI 04953170062988)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of BRONCHOFIBERSCOPE (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170062988), received between May 3, 2023 and Apr 11, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 27, 2023Open, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2490-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Failure to Clean Adequately1
  • Microbial Contamination of Device1
  • Peeled/Delaminated1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2026Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24851563 ↗
  • MalfunctionDec 30, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THE BRONCHOFIBERSCOPE INSERTION TUBE HAD COLOR LOSS AND PARTIAL DISAPPEARANCE OF THE MARKINGS. THE EVENT OCCURRED DURING SET UP IN ROOM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23925710 ↗
  • MalfunctionJun 20, 2023Microbial Contamination of Device; Microbial Contamination of Device

    THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOFIBERSCOPE TESTED POSITIVE FOR UNSPECIFIED MICROBIAL CONTAMINATION. THE ISSUE WAS FOUND AT REPROCESSING DURING A ROUTINE CULTURE OF THE SCOPE. THERE WAS NO PATIENT/USER HARM OR INJURY REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17165998 ↗
  • MalfunctionMay 3, 2023Scratched Material; Failure to Clean Adequately; Scratched Material; Failure to Clean Adequately

    A FIELD SERVICE ENGINEER REPORTED ON BEHALF OF THE CUSTOMER, BLACK SPOTS VISIBLE IN THE IMAGE ON BRONCHOFIBERSCOPE. THE CUSTOMER REPORTED, THE ISSUE TO BE FOUND DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE DEVICE WAS RETURNED, AND OLYMPUS REPAIR CENTER FOUND THE FORCEPS CHANNEL PORT SHAVED. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE EVALUATION OF THE RETURNED DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16865461 ↗