BRONCHOFIBERSCOPE BF-PE2 (UDI 04953170062988)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of BRONCHOFIBERSCOPE (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170062988), received between May 3, 2023 and Apr 11, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 27, 2023Open, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2490-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Failure to Clean Adequately1
- Microbial Contamination of Device1
- Peeled/Delaminated1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 11, 2026Detachment of Device or Device Component
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24851563 ↗MalfunctionDec 30, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THE BRONCHOFIBERSCOPE INSERTION TUBE HAD COLOR LOSS AND PARTIAL DISAPPEARANCE OF THE MARKINGS. THE EVENT OCCURRED DURING SET UP IN ROOM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23925710 ↗MalfunctionJun 20, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOFIBERSCOPE TESTED POSITIVE FOR UNSPECIFIED MICROBIAL CONTAMINATION. THE ISSUE WAS FOUND AT REPROCESSING DURING A ROUTINE CULTURE OF THE SCOPE. THERE WAS NO PATIENT/USER HARM OR INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17165998 ↗MalfunctionMay 3, 2023Scratched Material; Failure to Clean Adequately; Scratched Material; Failure to Clean Adequately
A FIELD SERVICE ENGINEER REPORTED ON BEHALF OF THE CUSTOMER, BLACK SPOTS VISIBLE IN THE IMAGE ON BRONCHOFIBERSCOPE. THE CUSTOMER REPORTED, THE ISSUE TO BE FOUND DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE DEVICE WAS RETURNED, AND OLYMPUS REPAIR CENTER FOUND THE FORCEPS CHANNEL PORT SHAVED. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE EVALUATION OF THE RETURNED DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16865461 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page