DISPOSABLE BIOPSY FORCEPS FB-211D (UDI 04953170158124)Recalls & reported problems
Olympus Medical Systems Corp. · DISPOSABLE BIOPSY FORCEPS · Class 2
All models of DISPOSABLE BIOPSY FORCEPS (23)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170158124), received between Dec 21, 2018 and Oct 26, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Known Impact Or Consequence To Patient3
- Bradycardia1
- Chest Pain1
- Low Blood Pressure/ Hypotension1
- Pneumothorax1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTS, JUST AFTER A FIBEROSCOPY FOR LUNG BIOPSY USING A DISPOSABLE BIOPSY FORCEPS, THE PATIENT EXPERIENCED SEVERE DETERIORATION OF THE PATIENT'S HEALTH CONDITION CHARACTERIZED BY CHEST PAIN FOLLOWED BY VAGAL DISCOMFORT RESULTING IN HYPOTENSION (BLOOD PRESSURE 70/50) AND BRADYCARDIA (HEART RATE 40BPM). A CHEST X-RAY WAS TAKEN AND SHOWED. NON-COMPRESSIVE LEFT PNEUMOTHORAX. THE PATIENT NECESSITATED 24-HOUR HOSPITALIZATION FOR PAIN MANAGEMENT (MORPHINE ADMINISTRATION) AND MONITORING. NO ADDITIONAL CONSEQUENCES TO THE PATIENT WERE REPORTED. THE SUSPECT DEVICE CANNOT BE RETURNED TO OLYMPUS FOE EVALUATION (IT WAS RETAINED BY THE FACILITY).
Patient problems: Bradycardia; Chest Pain; Low Blood Pressure/ Hypotension; Pneumothorax
MAUDE report 15674315 ↗MalfunctionMay 30, 2019Break; Break
THE USER FACILITY REPORTED THAT DURING A BRONCHOSCOPY PROCEDURE THE JAWS BROKE OFF INTO THE SCOPE WHILE THE DEVICE WAS BEING WITHDRAWN; IT NEVER FELL INTO THE PATIENT. THERE WAS NO PATIENT HARM. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. IN ADDITION, USER FACILITY REPORTED THAT THE BEFORE THE BREAKAGE, THE BRONCHOSCOPE WAS FLEXED TO THE RIGHT UPPER LOBE (RUL) APICAL SEGMENT IN ORDER TO OBTAIN A BIOPSY. THE FORCEPS WOULD NOT ADVANCE AND WAS WITHDRAWN. THE BRONCHOSCOPE WAS THEN BACKED OUT SLIGHTLY AND STRAIGHTENED OUT. THE FORCEPS WAS ADVANCED AGAIN AND THEN THE SCOPE WAS FLEXED INTO THE RUL APICAL SEGMENT AGAIN. THE SCOPE COULD NOT FLEX ADEQUATELY AND THE SCOPE WAS AGAIN STRAIGHTENED OUT. THE BRONCHOSCOPE WAS RETRACTED AND FORCEPS REMOVED. A NEW BIOPSY LOCATION IN THE LINGUAL WAS THEN CHOSEN AS AN ALTERNATE SITE. PRIOR TO RE-INSERTING THE FORCEPS THE TECHNICIAN NOTED THAT ONE OF THE FORCEPS ARMS WAS MISSING. A SEARCH WAS PERFORMED FOR THE MISSING PART. IT WAS FOUND IN THE FILTER OF THE SUCTION CANISTER. THE PROCEDURE WAS THEN COMPLETED WITHOUT DIFFICULTY AND WITH NO ADVERSE PATIENT EFFECT. THE ONLY PROCEDURAL ISSUE WAS NOT BEING ABLE TO GET BIOPSIES FROM THE ORIGINAL RUL APICAL SEGME
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8654595 ↗MalfunctionDec 21, 2018Break; Break
OLYMPUS WAS INFORMED THAT DURING A DIAGNOSTIC BRONCHOSCOPY PROCEDURE, THE JAWS FROM TWO FORCEPS BROKE ALLOWING THE WIRES TO PROTRUDE FROM THE DISTAL END OF THE DEVICES. NO DEVICE FRAGMENT FELL INTO THE PATIENT. IT WAS REPORTED THAT THERE WAS A POPPING NOISE NOTED WHEN THE JAWS CLOSED. THE DOCTOR WAS ABLE TO RETRIEVE THE INTENDED SPECIMENS. THERE WAS NO UNUSUAL BLEEDING OBSERVED. THE INTENDED PROCEDURE WAS COMPLETED WITH A THIRD SIMILAR DEVICE. THERE WAS NO PATIENT INJURY. THIS IS REPORT 2 OF 2.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8192359 ↗MalfunctionDec 21, 2018Break; Break
OLYMPUS WAS INFORMED THAT DURING A DIAGNOSTIC BRONCHOSCOPY PROCEDURE, THE JAWS FROM TWO FORCEPS BROKE ALLOWING THE WIRES TO PROTRUDE FROM THE DISTAL END OF THE DEVICES. NO DEVICE FRAGMENT FELL INTO THE PATIENT. IT WAS REPORTED THAT THERE WAS A POPPING NOISE NOTED WHEN THE JAWS CLOSED. THE DOCTOR WAS ABLE TO RETRIEVE THE INTENDED SPECIMENS. THERE WAS NO UNUSUAL BLEEDING OBSERVED. THE INTENDED PROCEDURE WAS COMPLETED WITH A THIRD SIMILAR DEVICE. THERE WAS NO PATIENT INJURY. THIS IS REPORT 1 OF 2.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8192345 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page