HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12ERecalls & reported problems

Olympus Medical Systems Corp. · "OTV-S7PROH-HD-12E CAM HEAD 1.2X EYEPIECE" · Class 2

All models of HD AUTOCLAVABLE CAMERA HEAD (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus HD camera head for surgical laparoscopes has 1,876 adverse event reports on record, with no deaths and 1 injury. Most reports describe no image on the screen, with others describing poor picture quality, a display that cuts in and out, and split or torn casings. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,886
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,886 adverse event reports that list this exact device (DI 04953170228834), received between Apr 7, 2020 and Sep 30, 2026 — classified as 1 injury, 1,885 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (1,022 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 45 reportsDec 2023: 46 reportsJan 2024: 44 reportsFeb 2024: 73 reportsMar 2024: 49 reportsApr 2024: 59 reportsMay 2024: 60 reportsJun 2024: 71 reportsJul 2024: 56 reportsAug 2024: 43 reportsSep 2024: 43 reportsOct 2024: 49 reportsNov 2024: 23 reportsDec 2024: 24 reportsJan 2025: 17 reportsFeb 2025: 7 reportsMar 2025: 11 reportsApr 2025: 16 reportsMay 2025: 7 reportsJun 2025: 6 reportsJul 2025: 14 reportsAug 2025: 13 reportsSep 2025: 9 reportsOct 2025: 15 reportsNov 2025: 17 reportsDec 2025: 6 reportsJan 2026: 17 reportsFeb 2026: 7 reportsMar 2026: 14 reportsApr 2026: 15 reportsMay 2026: 20 reportsJun 2026: 13 reportsJul 2026: 10 reportsAug 2026: 9 reportsSep 2026: 11 reportsOct 2026: 0 reports20242025202673
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image1,022
  • Poor Quality Image432
  • Erratic or Intermittent Display135
  • Material Split, Cut or Torn82
  • Break52
  • Loose or Intermittent Connection46
  • Display or Visual Feedback Problem39
  • Leak/Splash33

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,822
  • No Patient Involvement33
  • No Consequences Or Impact To Patient32
  • Insufficient Information3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 1,886 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Leak/Splash; Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954947 ↗
  • MalfunctionSep 30, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952628 ↗
  • MalfunctionSep 30, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949906 ↗
  • MalfunctionSep 29, 2026Poor Quality Image; Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935668 ↗
  • MalfunctionSep 29, 2026Electrical /Electronic Property Problem; Poor Quality Image

    The customer reported the camera head had a static that will not clear during reprocessing. During the initial evaluation of the device, it was observed that it was due to a faulty cable. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934110 ↗
  • MalfunctionSep 21, 2026Poor Quality Image; Material Split, Cut or Torn

    The customer reported crack on the screen during reprocessing involving an Olympus camera head. During initial evaluation it was observer, one lines on the screen. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50843129 ↗
  • MalfunctionSep 17, 2026No Display/Image

    The customer reported black screen, no picture during an unknown event using an Olympus camera head. There was no patient harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50797585 ↗
  • MalfunctionSep 14, 2026No Display/Image

    The customer reported no image during reprocessing involving an Olympus Camera Head. There was no patient involvement reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755854 ↗
  • MalfunctionSep 8, 2026Erratic or Intermittent Display

    The customer reported image noise when the cable was moved during a pre-use inspection with an Olympus camera head. No patients were involved.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50667780 ↗
  • MalfunctionSep 7, 2026Poor Quality Image

    During initial evaluation of the device the corrosion on the coupler unit was observed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50649728 ↗
  • MalfunctionSep 1, 2026Failure to Clean Adequately

    The customer returned the Camera Head for a separate issue. During the initial evaluation of the device, it was observed that the cable had a stain. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50552539 ↗
  • MalfunctionAug 31, 2026Erratic or Intermittent Display; Device Emits Odor

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50529876 ↗
  • MalfunctionAug 18, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50307068 ↗
  • MalfunctionAug 18, 2026No Display/Image; Loose or Intermittent Connection; Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50304574 ↗
  • MalfunctionAug 10, 2026Loose or Intermittent Connection

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50194038 ↗
  • MalfunctionAug 7, 2026Leak/Splash; Poor Quality Image; Output Problem

    The customer returned the Camera Head, which is an Olympus asset with no reported complaint. During the evaluation of the device, Image anomaly, Leakage test failed and Operation failure was found.was observed. There was no patient involvement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50181574 ↗
  • MalfunctionAug 7, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50175427 ↗
  • MalfunctionAug 4, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50136566 ↗
  • MalfunctionAug 3, 2026Display or Visual Feedback Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50116621 ↗
  • MalfunctionAug 3, 2026Loose or Intermittent Connection

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50113557 ↗
  • MalfunctionJul 30, 2026Poor Quality Image

    The customer reported bad image during inspection for use involving an Olympus camera head. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50091927 ↗
  • MalfunctionJul 28, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50060663 ↗
  • MalfunctionJul 27, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50038942 ↗
  • MalfunctionJul 20, 2026Poor Quality Image

    THE CUSTOMER REPORTED FLICKERING IMAGE AND A RED LINE DURING DEVICE INSPECTION FOR USE INVOLVING AN OLYMPUS CAMERA HEAD. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25882087 ↗
  • MalfunctionJul 14, 2026Erratic or Intermittent Display; Electrical Shorting

    THE CUSTOMER REPORTED A SHORT IN THE MIDDLE OF THE CORD AND SCREEN COMES IN AND OUT DURING A DIAGNOSTIC ENDOSCCOPIC CARPAL TUNNEL PROCEDURE INVOLVING AN OLYMPUS CAMERA HEAD. THE INTENDED PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25812648 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →