This Covidien round balloon is a surgical balloon used in hernia repair surgery. Most of the 859 reports describe balloons that could not be inflated, often because they had broken or developed a hole, and in several cases surgeons switched to another balloon to continue. The reports include 4 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
850
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 850 adverse event reports that list this exact device (DI 10884521067929), received between Dec 13, 2016 and Aug 5, 2026 — classified as 4 injuries, 846 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (566 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 9 reportsJan 2024: 9 reportsFeb 2024: 3 reportsMar 2024: 5 reportsApr 2024: 9 reportsMay 2024: 6 reportsJun 2024: 21 reportsJul 2024: 10 reportsAug 2024: 10 reportsSep 2024: 18 reportsOct 2024: 11 reportsNov 2024: 10 reportsDec 2024: 13 reportsJan 2025: 13 reportsFeb 2025: 8 reportsMar 2025: 12 reportsApr 2025: 14 reportsMay 2025: 12 reportsJun 2025: 11 reportsJul 2025: 22 reportsAug 2025: 9 reportsSep 2025: 5 reportsOct 2025: 11 reportsNov 2025: 8 reportsDec 2025: 22 reportsJan 2026: 7 reportsFeb 2026: 13 reportsMar 2026: 14 reportsApr 2026: 7 reportsMay 2026: 1 reportsJun 2026: 2 reportsJul 2026: 4 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202622
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inflation Problem566
  • Break422
  • Gas/Air Leak34
  • Material Split, Cut or Torn34
  • Detachment of Device or Device Component31
  • Leak/Splash24
  • Noise, Audible14
  • Detachment Of Device Component10

Patient problems

  • No Clinical Signs, Symptoms or Conditions553
  • No Known Impact Or Consequence To Patient169
  • No Consequences Or Impact To Patient108
  • Insufficient Information24
  • No Patient Involvement10
  • No Information5
  • Foreign Body In Patient3
  • No Code Available2

Recent adverse event reports

The 25 most recent of 850 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2026Break; Inflation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50138614 ↗
  • MalfunctionAug 4, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50131739 ↗
  • MalfunctionJul 29, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50065438 ↗
  • MalfunctionJul 29, 2026Inflation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50065390 ↗
  • MalfunctionJul 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50061043 ↗
  • MalfunctionJul 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50059717 ↗
  • MalfunctionJun 24, 2026Inflation Problem

    ACCORDING TO THE REPORTER, DURING PRE-OPERATIVE PREPARATION FOR A TEP (TOTALLY EXTRAPERITONEAL) HERNIOPLASTY, THE BALLOON WAS TESTED AND COULD NOT BE INFLATED. A NEW BALLOON WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25605075 ↗
  • MalfunctionJun 9, 2026Inflation Problem

    IT WAS REPORTED THAT DURING THE TOTALLY EXTRAPERITONEAL HERNIOPLASTY FOR COLON CANCER, THE BALLOON COULD NOT BE INFLATED. A NEW BALLOON WAS USED TO CONTINUE THE SURGERY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25451474 ↗
  • MalfunctionMay 1, 2026Inflation Problem

    ACCORDING TO THE REPORTER, DURING A TOTALLY EXTRAPERITONEAL HERNIOPLASTY PROCEDURE, THE BALLOON COULD NOT BE INFLATED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25055832 ↗
  • MalfunctionApr 27, 2026Break

    IT WAS REPORTED THAT PREOPERATIVELY, THE BALLOON ROUND SHAPE WAS UNUSABLE BECAUSE THE BALLOON PART BURST AND THERE WAS A HOLE IN THE BALLOON. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25001028 ↗
  • MalfunctionApr 21, 2026Break

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC TOTALLY EXTRAPERITONEAL (TEP) HERNIA PROCEDURE USING A ROUND SHAPE BALLOON, THE SURGEON NOTICED THAT THE BALLOON HAD RUPTURED WHILE BEING INFLATED AND THAT THE BALLOON PHYSICALLY BROKE. THE BALLOON WAS REMOVED AND THE OPERATION CONTINUED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24948663 ↗
  • MalfunctionApr 21, 2026Inflation Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC INGUINAL HERNIA REPAIR, WHEN A ROUND SHAPE BALLOON WAS INSERTED AND, AT COMMENCEMENT OF INFLATION, THE BALLOON DID NOT INFLATE. ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE WITHOUT ISSUES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24935142 ↗
  • MalfunctionApr 14, 2026Break

    IT WAS REPORTED THAT PREOPERATIVELY OF A LAPAROSCOPIC INGUINAL HERNIA REPAIR, THE BALLOON THAT WAS DEPLOYED RUPTURED AND COULD NOT BE USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24873343 ↗
  • MalfunctionApr 12, 2026Break

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC INGUINAL HERNIA PROCEDURE BUT PRIOR TO PATIENT USE, THE BALLOON TROCAR THAT WAS DEPLOYED RUPTURED AND COULD NOT BE USED. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24853238 ↗
  • MalfunctionApr 9, 2026Break; Inflation Problem

    IT WAS REPORTED THAT DURING A TOTALLY EXTRAPERITONEAL (TEP) HERNIOPLASTY, THE BALLOON RUPTURED AND COULD NOT BE INFLATED. THE OPERAT OR USED A NEW BALLOON TO CONTINUE THE SURGERY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24831558 ↗
  • MalfunctionApr 7, 2026Break

    IT WAS REPORTED THAT PREOPERATIVELY, THE BALLOON PORTION OF THE TROCAR RUPTURED, AND DID NOT INFLATE. THE PRODUCT WAS REPLACED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24797325 ↗
  • MalfunctionMar 31, 2026Break

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC NEPHRECTOMY, WHEN THE BALLOON WAS INFLATED, THERE WAS A HOLE. A NEW BALLOON WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742014 ↗
  • MalfunctionMar 24, 2026Break

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC INGUINAL HERNIA REPAIR, WHILE BEING INFLATED TO DISSECT THE TOTALLY EXTRAPERITONEAL SPACE, THE BALLOON PHYSICALLY BROKE. THE ISSUE WAS RESOLVED BY USING A NEW BALLOON TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24678039 ↗
  • MalfunctionMar 20, 2026Inflation Problem

    IT WAS REPORTED THAT DURING A TOTALLY EXTRAPERITONEAL (TEP) HERNIOPLASTY, DURING TESTING OF THE DEVICE, THE BALLOON DID NOT INFLATE. A NEW BALLOON WAS USED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24651357 ↗
  • MalfunctionMar 19, 2026Inflation Problem

    IT WAS REPORTED THAT INTRAOPERATIVELY, THE BALLOON OF TWO DEVICES DID NOT INFLATE. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24643279 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT DURING A TOTALLY EXTRAPERITONEAL (TEP) HERNIOPLASTY, PRIOR TO PATIENT USE, THE BALLOON RUPTURED, AND COULD NOT BE INFLATED PROPERLY. A NEW BALLOON WAS USED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24638461 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT DURING USE, THE BALLOON PHYSICALLY BROKE AND RUPTURED, RESULTING IN AN INABILITY TO INFLATE THE BALLOON PROPERLY. THE OPERATOR USED A NEW BALLOON TO CONTINUE THE SURGERY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24636349 ↗
  • MalfunctionMar 18, 2026Inflation Problem

    ACCORDING TO THE REPORTER, DURING A TOTALLY EXTRAPERITONEAL HERNIOPLASTY, THE BALLOON COULD NOT BE INFLATED. A NEW DEVICE WAS USED T O RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24630876 ↗
  • MalfunctionMar 18, 2026Break; Inflation Problem

    ACCORDING TO THE REPORTER, PRE-OPERATIVELY, WHILE TESTING THE DEVICE, THERE WAS A HOLE IN THE BALLOON. THE BALLOON COULD NOT BE INFLATED BECAUSE OF THE HOLE IN IT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24628040 ↗
  • MalfunctionMar 13, 2026Break; Decrease in Pressure

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC INGUINAL HERNIA REPAIR, THE BALLOON PHYSICALLY BROKE DURING BALLOON INFLATION. THIS RESULTED IN A RAPID LOSS OF PNEUMOPERITONEUM. A NEW BALLOON WAS USED TO COMPLETE THE CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24594507 ↗

FDA clearance and classification

Cleared through 510(k) K935426 on Jan 26, 1994 (decision: Substantially Equivalent). Submitted by Origin Medsystems, Inc. as “Preperitoneal Distention Balloon System”. The FDA review took 77 days. FDA record ↗

The FDA regulates product code GCJ (laparoscope, general & plastic surgery) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Common questions about the PDB OMS-PDB1000

Has the PDB OMS-PDB1000 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10884521067929), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PDB OMS-PDB1000?

The FDA's MAUDE database lists 850 adverse event reports for this device, received between Dec 13, 2016 and Aug 5, 2026. In those reports, the most common specific device problems are “Inflation Problem” (566), “Break” (422) and “Gas/Air Leak” (34); the most common patient problem is “Foreign Body In Patient” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PDB OMS-PDB1000?

None of the reports are classified as deaths, 4 as injuries and 846 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PDB OMS-PDB1000 FDA approved?

It was cleared through 510(k) K935426 on Jan 26, 1994. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the PDB OMS-PDB1000, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand PDB, model OMS-PDB1000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page