AIRWAY MOBILESCOPE OLYMPUS MAF TYPE TMRecalls & reported problems
Olympus Medical Systems Corp. · MAF-TM AIRWAY MOBILESCOPE-TREATMENT/ ASP · Class 2
All models of AIRWAY MOBILESCOPE (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Olympus MAF-TM airway mobilescope is a flexible bronchoscope used to examine and treat the airway. The 43 adverse event reports mostly describe the scope's coating peeling or flaking off, plus cleaning problems, debris in the forceps channel, and lost images; no deaths and no injuries are reported. Two open recalls cite this device: one because the drying process for the channel was not validated, leaving water inside scopes returned after repair, and one because the scope can get lodged in an endotracheal tube connector.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 43
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 04953170288630), received between Feb 15, 2022 and Aug 11, 2026 — classified as 43 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (38 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0340-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 27, 2023Class IIOpen, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2495-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Peeled/Delaminated38
- Failure to Clean Adequately4
- Device Reprocessing Problem1
- Image Display Error/Artifact1
- No Display/Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions43
Recent adverse event reports
The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Image Display Error/Artifact
No additional information received from customer
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50214122 ↗MalfunctionApr 10, 2026No Display/Image
THE AIRWAY MOBILESCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF FAILED LEAK TEST AND CRACK AT HINGE OF SCREEN. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, NO IMAGE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24840503 ↗MalfunctionOct 7, 2024Peeled/Delaminated; Failure to Clean Adequately; Peeled/Delaminated; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE MOBILESCOPE EXHIBITED THE INSERTION SECTION OF THE FLEXIBLE TUBE COATING PEELED OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20389283 ↗MalfunctionSep 12, 2024Peeled/Delaminated; Failure to Clean Adequately; Peeled/Delaminated; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE AIRWAY MOBILESCOPE HAD FOREIGN MATERIAL IN THE FORCEPS CHANNEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20201790 ↗MalfunctionAug 21, 2024Peeled/Delaminated; Peeled/Delaminated
DURING DEVICE EVALUATION, IT WAS OBSERVED THE AIRWAY MOBILESCOPE EXHIBITED PEELING OF THE CONNECTING TUBE COATING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20015115 ↗MalfunctionAug 14, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE AIRWAY MOBILESCOPE EXHIBITED THE IMAGE GUIDE TUBE PEELED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19980642 ↗MalfunctionMay 15, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COATING WAS PEELING ON THE IMAGE GUIDE TUBE OF THE VIDEOSCOPE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19324105 ↗MalfunctionJan 31, 2024Peeled/Delaminated; Peeled/Delaminated
DURING THE DEVICE EVALUATION, IT WAS OBSERVED THE AIRWAY MOBILESCOPE EXHIBITED PEELING OF THE CONNECTING TUBE COATING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18615876 ↗MalfunctionJan 5, 2024Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE AIRWAY MOBILESCOPE HAD AN IMAGE DEFECT, WAS INOPERABLE AND SEPARATING. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE IMAGE GUIDE TUBE COAT WAS PEELING. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18451789 ↗MalfunctionDec 12, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE AIRWAY MOBILESCOPE HAD LEAKAGE FROM THE INSERTION SECTION. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE CONNECTING TUBE HAD COATING PEELING (1 TO 2 OR MORE MILLIMETERS). THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18315640 ↗MalfunctionDec 11, 2023Peeled/Delaminated; Peeled/Delaminated
AN OLYMPUS REPRESENTATIVE REPORTED TO OLYMPUS ON BEHALF OF THE CUSTOMER THAT THE AIRWAY MOBILESCOPE HAD AN ISOLATED LEAK. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE COAT OF THE IMAGE GUIDE PIPE WAS PEELING OFF (1MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THE EVENT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18307409 ↗MalfunctionNov 19, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE AIRWAY MOBILESCOPE COULD NOT ANGULATE SUFFICIENTLY. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, IT WAS OBSERVED THAT THE COATING OF THE IMAGE GUIDE PIPE WAS PEELING OFF (1 MILLIMETER (MM) TO 2MM OR MORE). THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18167323 ↗MalfunctionOct 24, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE AIRWAY MOBILESCOPE TIP OF THE BENDING SECTION COVER WAS LOOSE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE CONNECTING TUBE HAD COATING PEELING (1MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17995060 ↗MalfunctionOct 19, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER INITIALLY REPORTED TO OLYMPUS THAT THE AIRWAY MOBILESCOPE HAD PINHOLES. THERE WERE NO REPORTS OF PATIENT HARM. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS IDENTIFIED: THE INDICATION ON CONNECTING TUBE HAD A DISAPPEARED AREA (1MM SQUARED OR MORE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17966649 ↗MalfunctionSep 19, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS THAT AIRWAY MOBILESCOPE'S LIGHT GUIDE BUNDLE SHOWED BREAKAGE. THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION. DURING EVALUATION, CONNECTING TUBE SHOWED THE COATING WAS PEELING AND HAD A MISSING INDICATOR. THIS MEDICAL DEVICE REPORT (MDR)IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17777704 ↗MalfunctionSep 18, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE AIRWAY MOBILESCOPE WAS CAUGHT IN THE FORCEPS CHANNEL. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE CONNECTING TUBE COATING WAS PEELING (1MM SQUARED OR MORE). THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17765146 ↗MalfunctionSep 8, 2023Peeled/Delaminated; Peeled/Delaminated
A USER FACILITY SUBMITTED A REPAIR REQUEST TO THE OLYMPUS SERVICE CENTER, FOR AN AIRWAY MOBILESCOPE HAVING BAD OUTPUT. UPON INSPECTION AND TESTING OF THE CUSTOMER RETURNED DEVICE, THE COATING ON THE SCOPE IMAGE GUIDE SERPENTINE TUBE WAS FOUND PEELED OFF (1 TO 2 MM OR MORE). THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO REPORT OF PATIENT HARM OR USER INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17713023 ↗MalfunctionSep 7, 2023Peeled/Delaminated; Peeled/Delaminated
AN OLYMPUS REPRESENTATIVE REPORTED TO OLYMPUS ON BEHALF OF THE CUSTOMER THAT THE AIRWAY MOBILESCOPE HAD BLACK SPOTS ON THE IMAGE. THE ISSUE WAS FOUND DURING REPROCESSING AFTER THE INTENDED DIAGNOSTIC BRONCHIAL EXAMINATION PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED WITH THE SAME EQUIPMENT. THERE WAS NO DELAY, AND THE PATIENT WAS NOT UNDER ANESTHESIA. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE COATING ON THE INSERTION SECTION PEELED OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17707046 ↗MalfunctionAug 25, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE OLYMPUS REPRESENTATIVE REPORTED (ON BEHALF OF THE CUSTOMER) THAT THE AIRWAY MOBILESCOPE HAD A COMPROMISED IMAGE. THERE WERE NO REPORTS OF PATIENT HARM. DURING EVALUATION OF THE RETURNED DEVICE, IT WAS FOUND THAT A FOREIGN OBJECTS CAME OUT OF FORCEPS CHANNEL AND THEREFORE IS CONSIDERED A MEDICAL DEVICE REPORT (MDR). THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17627766 ↗MalfunctionJul 14, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS THAT THE AIRWAY MOBILESCOPE HAD SPOTS IN THE IMAGE AND BROKEN FIBERS. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE CONNECTING TUBE HAD COATING PEELING. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17325785 ↗MalfunctionJun 8, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS THE AIRWAY MOBILESCOPE EXHIBITED A LEAK DUE TO PINHOLES IN THE BENDING RUBBER (THE METAL WAS NOT STICKING OUT). THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT. DURING TESTING AND INSPECTION OF THE RETURNED DEVICE REVEALED THAT THE COATING OF THE SCOPE INSERTION GUIDE SERPENTINE WAS PEELING OFF. (1MM2 OR MORE). THIS HAD BEEN ATTRIBUTED DUE TO STRESS OF REPEATED USE, DETERIORATION, EXTERNAL FACTORS, OR HANDLING. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17094836 ↗MalfunctionApr 23, 2023Peeled/Delaminated; Peeled/Delaminated
THE DISTRIBUTOR REPORTED TO OLYMPUS MEDICAL ON BEHALF OF THE CUSTOMER THAT THE AIRWAY PORTABLE MOBILESCOPE HAD DEFECTS AT THE BENDING SECTION AND INSERTION TUBE. THE DEVICE WAS RETURNED TO OLYMPUS MEDICAL FOR EVALUATION. DURING INSPECTION, THE INSERTION SECTION WAS FOUND TO HAVE AN AREA OF COATING PEELED OFF THAT MEASURED AT LEAST 1 SQUARE MILLIMETER. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16792875 ↗MalfunctionApr 19, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THAT THE AIRWAY MOBILESCOPE EXHIBITED A LEAK. NO PATIENT HARM WAS REPORTED IN RELATION TO THIS EVENT. THE SUBJECT DEVICE WAS SUBSEQUENTLY RETURNED TO OLYMPUS AND EVALUATED. THE EVALUATION REVEALED THAT THE CORRUGATED PIPE COAT ON THE DEVICE WAS PEELING. THE IMAGE GUIDE (IG) TUBE WAS ALSO PEELING. THIS MDR (MEDICAL DEVICE REPORT) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16773797 ↗MalfunctionApr 19, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS THAT THE AIRWAY MOBILESCOPE HAD A BENDING SECTION PINHOLE. DURING INSPECTION AND EVALUATION, A FOREIGN OBJECT CAME OUT OF THE FORCEPS CHANNEL. THERE WAS NO REPORT OF PATIENT HARM OR USER INJURY ASSOCIATED WITH THIS EVENT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTIONS FOUND DURING INCOMING INSPECTION AND EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16770794 ↗MalfunctionApr 12, 2023Peeled/Delaminated; Peeled/Delaminated
THE AIRWAY MOBILESCOPE WAS SENT TO AN OLYMPUS SERVICE CENTER FOR EVALUATION AFTER BEING RETURNED FROM THE END USER WITH NO COMPLAINT. UPON INSPECTION OF THE CUSTOMER¿S RETURNED DEVICE IT WAS OBSERVED, THE IMAGE GUIDE TUBE COAT WAS PEELING. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16723872 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page