BRONCHOVIDEOSCOPE BF-P150 (UDI 04953170288876)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of BRONCHOVIDEOSCOPE (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 04953170288876), received between Oct 6, 2020 and Apr 9, 2026 — classified as 19 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (10 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 27, 2023Open, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2490-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
- Jun 8, 2023Open, Classified
There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
Z-2013-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- No Display/Image10
- Break2
- Poor Quality Image2
- Component Missing1
- Detachment of Device or Device Component1
- Failure to Clean Adequately1
- Material Deformation1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 9, 2026Poor Quality Image
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED A NOISY IMAGE OCCURS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24829495 ↗MalfunctionApr 3, 2026Poor Quality Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED AN IMAGE WITH INTERFERENCE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24780340 ↗MalfunctionMar 11, 2026No Display/Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED AN IMAGE BLACKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24568058 ↗MalfunctionNov 26, 2025No Display/Image; No Display/Image
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23657324 ↗MalfunctionOct 21, 2024No Display/Image; No Display/Image
IT WAS REPORTED, THAT THE BRONCHOVIDEOSCOPE HAD FLUCTUATIONS IN THE IMAGE. THE ISSUE OCCURRED, DURING A DIAGNOSTIC BRONCHOSCOPY PROCEDURE. A SIMILAR DEVICE WAS USED TO COMPLETE THE PROCEDURE. UPON TESTING, THE SCREEN BECAME BLACK AND SHOWED NO IMAGE. AFTER A FEW SECONDS, THE IMAGE WAS ABLE TO BE RESTORED. THERE WAS NO REPORTED PATIENT HARM OR IMPACT, DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20491350 ↗MalfunctionOct 3, 2024Component Missing; Detachment of Device or Device Component; Component Missing; Detachment of Device or Device Component
IT WAS REPORTED THE BRONCHOVIDEOSCOPE WAS MISSING A DISTAL COVER AND THE BIOPSY PORT HAD INCOMPLETE ACCESSORIES. THE ISSUES WERE DISCOVERED AT REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20362362 ↗MalfunctionSep 27, 2024No Display/Image; No Display/Image
IT WAS REPORTED THE BRONCHOVIDEOSCOPE EXPERIENCED NO IMAGE. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20326572 ↗MalfunctionSep 18, 2024No Display/Image; No Display/Image
IT WAS REPORTED, THE BRONCHOVIDEOSCOPE EXHIBITED NO ENDOSCOPY IMAGE ON THE MONITOR. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR A DIAGNOSTIC BRONCHOSCOPY PROCEDURE. THERE WERE NO REPORTS OF DELAY, PATIENT HARM, INJURY, OR DEATH. ADDITIONAL DETAILS RELATING TO THE PATIENT AND THE EVENT HAVE BEEN REQUESTED, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20250792 ↗MalfunctionMay 23, 2024Break; Break
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOVIDEOSCOPE EXHIBITED INTERMITTENT NO IMAGE DURING DOWN DIRECTION ANGULATION DUE TO DAMAGE ON THE CHARGED COUPLE DEVICE UNIT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19383214 ↗MalfunctionMay 15, 2024No Display/Image; No Display/Image
IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE HAD NO IMAGE DURING ANGULATION. IT IS UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19325145 ↗MalfunctionMay 3, 2024No Display/Image; No Display/Image
IT WAS REPORTED THE BRONCHOVIDEOSCOPE RELAYED A BLACK SCREEN INSTEAD OF AN IMAGE. THE ISSUE OCCURRED DURING CUSTOMER MAINTENANCE. THERE WAS NO PROCEDURE OR PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19239359 ↗MalfunctionMar 22, 2024No Display/Image; No Display/Image
THE CUSTOMER REPORTED THERE WAS NO PROCEDURE DELAY AND REPORTED THE PATIENT WAS NOT UNDER ANESTHESIA WHEN ISSUE WAS FOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18960140 ↗MalfunctionMar 17, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN ADHESIVE ON A-RUBBER (BENDING SECTION COVER) AND A-RUBBER (BENDING SECTION COVER). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18919128 ↗MalfunctionFeb 8, 2024Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOVIDEOSCOPE HAD A LEAK FROM BENDING SECTION COVER. THE ISSUE WAS FOUND DURING REPROCESSING. THE DEVICE WAS RETURNED FOR EVALUATION. DURING TESTING AND INSPECTION OF THE DEVICE, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: CONNECTING TUBE HAD COATING PEELED. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18670887 ↗MalfunctionJan 2, 2024No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS, WHILE USING THE BRONCHOVIDEOSCOPE, THAT A BLACK SCREEN APPEARS WHEN USING THE ANGULATION. THIS ISSUE OCCURRED DURING REPROCESSING, THE INTENDED DIAGNOSTIC BRONCHOSCOPY PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO DELAY, INJURY OR DEATH DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18427680 ↗MalfunctionOct 20, 2023No Display/Image; No Display/Image
THE THIRD PARTY REPAIR REPORTED (ON BEHALF OF THE CUSTOMER) THAT THE BRONCHOVIDEOSCOPE WAS LEAKING AND NOT DISPLAYING AN IMAGE. THIS OCCURRED DURING REPROCESSING FOR A DIAGNOSTIC BRONCHOSCOPY PROCEDURE. THE PROCEDURE WAS COMPLETED WITH SIMILAR COMPETITOR DEVICE. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17977811 ↗MalfunctionMay 30, 2023Material Deformation; Material Deformation
THE BRONCHOVIDEOSCOPE WAS RETURNED AS PART OF THE ASSET RETURN PROCESS. DURING ROUTINE INSPECTION OF THE DEVICE BY OLYMPUS, IT WAS FOUND THAT THE FORCEPS CHANNEL PORT WAS SHAVED. THERE WAS NO PROCEDURAL OR PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17024150 ↗MalfunctionApr 26, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
THE CUSTOMER REPORTED TO OLYMPUS THE ANGULATION KNOB FELT LOOSE AND THERE WAS A LEAK NEAR THE SWITCH ON THE BRONCHOVIDEOSCOPE THAT WAS FOUND DURING REPROCESSING. NO PATIENT HARM WAS REPORTED. THE DEVICE WAS RETURNED FOR EVALUATION, AND THE FORCEPS CHANNEL PORT WAS SHAVED. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16818301 ↗MalfunctionOct 6, 2020Break; Break
THE COATING OF THE INSERTION TUBE WAS PARTIALLY PEELED OFF. THE USER CANCELED THE PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT. THE USER FACILITY DID NOT PROVIDE OTHER DETAILED INFORMATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10632733 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page