RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF-T3 (UDI 04953170307188)Recalls & reported problems

Olympus Medical Systems Corp. · RHINO-LARYNGOFIBERSCOPE · Class 2

All models of RHINO-LARYNGOFIBERSCOPE (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04953170307188), received between Jul 12, 2022 and Sep 16, 2026 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0332-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately3
  • Break2
  • Corroded1
  • Detachment of Device or Device Component1
  • Device Reprocessing Problem1
  • Microbial Contamination of Device1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Corroded

    The customer returned the Rhino Laryngofiberscope to the service center for a separate issue. During the device analysis, it was indicated that a corroded distal end was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50791172 ↗
  • MalfunctionAug 26, 2026Break; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50455667 ↗
  • MalfunctionAug 10, 2026Device Reprocessing Problem

    The customer returned the rhino-laryngofiberscope to the service center for a separate issue. During the device analysis, It was indicated that dirt deposition in the bending section cover and connecting tube was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50195863 ↗
  • MalfunctionJul 1, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED THAT DUE TO CLOGGING OF THE CHANNEL TUBE (FOREIGN OBJECT), THE TUBE CLEANING BRUSH CANNOT BE INSERTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22378313 ↗
  • MalfunctionNov 7, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED FOREIGN OBJECTS ON THE OBJECTIVE LENS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20629972 ↗
  • MalfunctionApr 24, 2024Microbial Contamination of Device; Microbial Contamination of Device

    THE CUSTOMER REPORTED THE RHINO-LARYNGOFIBERSCOPE WAS STERILIZED, BUT WAS OVERLOADED WITH BACTERIA. THE ISSUE WAS FOUND WHEN PREPARING THE DEVICE FOR USE IN A DIAGNOSTIC CANNULOSCOPY PROCEDURE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19174174 ↗
  • MalfunctionJan 8, 2024Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE RHINO-LARYNGO-SCOPE HAD HOLES IN THE INSERTION SECTION AND AN ELONGATED ANGLE WIRE. THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION. DURING DEVICE EVALUATION, FOREIGN MATERIAL WAS FOUND IN THE CHANNEL. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. NO ADVERSE EFFECTS TO PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18466742 ↗
  • MalfunctionJan 20, 2023Break; Break

    ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER: THE ISSUE WAS FOUND DURING REPROCESSING AND THE INTENDED PROCEDURE WAS PERFORMED WITH A SIMILAR DEVICE REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16216355 ↗
  • InjuryJul 12, 2022Peeled/Delaminated; Peeled/Delaminated

    THE CUSTOMER REPORTED THAT DURING AN UNSPECIFIED PROCEDURE, THE COATING ON THE INSERTION TUBE OF THE SUBJECT DEVICE WAS TORN OFF, AND THE PATIENT SWALLOWED THE PEELED OFF MATERIAL. THE INTENDED PROCEDURE WAS COMPLETED USING THE SUBJECT DEVICE. THE WAS NO IMPACT ON THE PATIENT. ADDITIONAL DETAILS HAVE BEEN REQUESTED REGARDING THE REPORTED EVENT. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14988044 ↗

Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices

All NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page