RHINO-LARYNGOFIBERSCOPE OLYMPUS ENF-T3 (UDI 04953170307188)Recalls & reported problems
Olympus Medical Systems Corp. · RHINO-LARYNGOFIBERSCOPE · Class 2
All models of RHINO-LARYNGOFIBERSCOPE (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04953170307188), received between Jul 12, 2022 and Sep 16, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0332-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Clean Adequately3
- Break2
- Corroded1
- Detachment of Device or Device Component1
- Device Reprocessing Problem1
- Microbial Contamination of Device1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Corroded
The customer returned the Rhino Laryngofiberscope to the service center for a separate issue. During the device analysis, it was indicated that a corroded distal end was found. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50791172 ↗MalfunctionAug 26, 2026Break; Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50455667 ↗MalfunctionAug 10, 2026Device Reprocessing Problem
The customer returned the rhino-laryngofiberscope to the service center for a separate issue. During the device analysis, It was indicated that dirt deposition in the bending section cover and connecting tube was found. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50195863 ↗MalfunctionJul 1, 2025Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED THAT DUE TO CLOGGING OF THE CHANNEL TUBE (FOREIGN OBJECT), THE TUBE CLEANING BRUSH CANNOT BE INSERTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22378313 ↗MalfunctionNov 7, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED FOREIGN OBJECTS ON THE OBJECTIVE LENS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20629972 ↗MalfunctionApr 24, 2024Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED THE RHINO-LARYNGOFIBERSCOPE WAS STERILIZED, BUT WAS OVERLOADED WITH BACTERIA. THE ISSUE WAS FOUND WHEN PREPARING THE DEVICE FOR USE IN A DIAGNOSTIC CANNULOSCOPY PROCEDURE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19174174 ↗MalfunctionJan 8, 2024Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS THAT THE RHINO-LARYNGO-SCOPE HAD HOLES IN THE INSERTION SECTION AND AN ELONGATED ANGLE WIRE. THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION. DURING DEVICE EVALUATION, FOREIGN MATERIAL WAS FOUND IN THE CHANNEL. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18466742 ↗MalfunctionJan 20, 2023Break; Break
ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER: THE ISSUE WAS FOUND DURING REPROCESSING AND THE INTENDED PROCEDURE WAS PERFORMED WITH A SIMILAR DEVICE REPLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16216355 ↗InjuryJul 12, 2022Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED THAT DURING AN UNSPECIFIED PROCEDURE, THE COATING ON THE INSERTION TUBE OF THE SUBJECT DEVICE WAS TORN OFF, AND THE PATIENT SWALLOWED THE PEELED OFF MATERIAL. THE INTENDED PROCEDURE WAS COMPLETED USING THE SUBJECT DEVICE. THE WAS NO IMPACT ON THE PATIENT. ADDITIONAL DETAILS HAVE BEEN REQUESTED REGARDING THE REPORTED EVENT. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14988044 ↗
Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)785 reports
- NA 242268Medos International Sàrl · Nasopharyngoscope (flexible or rigid)573 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Olympus Medical Systems Corp. · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- VIVIDEO VNL9-CPHoya Corporation · Nasopharyngoscope (Flexible Or Rigid)389 reports
- PENTAX FNL-10RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)322 reports
- SINGLE USE BIOPSY VALVE MAJ-210Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)1 recall221 reports
- SINGLE USE SUCTION VALVE MAJ-209Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)228 reports
- PENTAX FNL-7RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)227 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page