RHINO-LARYNGO VIDEOSCOPE OLYMPUS ENF-V3 (REF N3828230)Recalls & reported problems

Olympus Medical Systems Corp. · ENF-V3 Flexible ultra thin video Rhino L · Class 2

All models of RHINO-LARYNGO VIDEOSCOPE (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus ENF-V3 flexible rhinolaryngoscope is a thin video scope used to look inside the nose and throat. The 62 reports mostly describe the scope coming apart: glue at the objective lens chipping or peeling away, parts detaching, the outer covering delaminating, and the scope not producing an image, along with cleaning and reprocessing difficulties. Reports involve 2 injuries and no deaths, and no recalls name this device. Reports rose sharply, from 5 in the previous 12 months to 29 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
62
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 62 adverse event reports that list this exact device (DI 04953170310416), received between Mar 16, 2021 and Aug 10, 2026 — classified as 2 injuries, 60 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (21 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 5 reportsJan 2026: 4 reportsFeb 2026: 2 reportsMar 2026: 3 reportsApr 2026: 7 reportsMay 2026: 5 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component21
  • Material Fragmentation8
  • Break6
  • Device Reprocessing Problem5
  • Peeled/Delaminated5
  • Failure to Clean Adequately4
  • Degraded3
  • Material Separation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions59
  • Insufficient Information2
  • Unspecified Tissue Injury1

Recent adverse event reports

The 25 most recent of 62 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 10, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50193502 ↗
  • MalfunctionAug 7, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50177879 ↗
  • MalfunctionJul 30, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50082733 ↗
  • MalfunctionMay 7, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED THE OBJECTIVE LENS GLUE CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25116241 ↗
  • MalfunctionMay 7, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO LARYNGO VIDEOSCOPE HAD A CHIPPED OBJECTIVE LENS GLUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25111944 ↗
  • MalfunctionMay 6, 2026Degraded; No Display/Image; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED THAT THE A-RUBBER GLUE IS PEELING (OFF), EXPOSED THREAD (DEGRADED), AND THE UNIT HAS NO IMAGE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25099430 ↗
  • MalfunctionMay 4, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO LARYNGO VIDEOSCOPE HAD A CHIPPED GLUE ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25069592 ↗
  • MalfunctionMay 1, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO LARYNGO VIDEOSCOPE EXHIBITED MISSING OBJECTIVE LENS GLUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056137 ↗
  • MalfunctionApr 28, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED CHIPPED GLUE ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023465 ↗
  • MalfunctionApr 28, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED MISSING OBJECTIVE LENS GLUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25020914 ↗
  • MalfunctionApr 24, 2026Contamination /Decontamination Problem; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED A DETACHED INSERTION TUBE AND RESIDUE ON THE BENDING SECTION COVER, BENDING SECTION COVER ADHESIVE, AND LENSES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24989036 ↗
  • MalfunctionApr 23, 2026Material Fragmentation; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED THAT THE OBJECTIVE LENS (OB LENS) HAS CHIPPED GLUE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24974460 ↗
  • MalfunctionApr 22, 2026Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEOSCOPE OBJECTIVE LENS GLUE WAS CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24954464 ↗
  • MalfunctionApr 21, 2026Material Fragmentation

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED CHIPPED ADHESIVE AROUND THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24939185 ↗
  • MalfunctionApr 16, 2026Break

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED THAT THE BENDING SECTION COVER (A-RUBBER GLUE) HAD CHIPPED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24897205 ↗
  • MalfunctionMar 18, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED CHIPPED GLUE ON DISTAL END (OBJECTIVE LENS). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24627454 ↗
  • MalfunctionMar 9, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED THE OBJECTIVE LENS GLUE CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24552009 ↗
  • MalfunctionMar 2, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED MISSING GLUE ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24486591 ↗
  • MalfunctionFeb 11, 2026Component Missing

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO VIDEOSCOPE EXHIBITED MISSING ADHESIVE AT THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24315937 ↗
  • MalfunctionFeb 3, 2026Material Separation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEOSCOPE OBJECTIVE LENS ADHESIVE WAS MISSING/DETACHED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24250244 ↗
  • MalfunctionJan 30, 2026Material Fragmentation; Detachment of Device or Device Component

    IT WAS REPORTED THAT SMALL BLACK FLEX WAS COMING OFF THE TIP OF THE RHINO-LARYNGO VIDEOSCOPE WITH EACH WIPE OF FIBER MICRO CLOTH. THE ISSUE WAS IDENTIFIED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24225680 ↗
  • MalfunctionJan 28, 2026Poor Quality Image; Failure to Clean Adequately

    DURING THE DEVICE EVALUATION, IT WAS REVEALED THAT THERE WAS A FOGGY IMAGE ON THE VIDEOSCOPE CAUSED BY A DIRTY OBJECTIVE LENS. NO PATIENTS WERE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24199016 ↗
  • MalfunctionJan 27, 2026Loss of or Failure to Bond

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOVIDEOSCOPE EXHIBITED MISSING ADHESIVE AROUND THE OBJECTIVE LENS, AND THE BENDING SECTION COVER ADHESIVE WAS MISSING A LARGE PART OF THE PLASTIC DISTAL END WITH EXPOSED THREADS AND METAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24192725 ↗
  • MalfunctionJan 9, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO LARYNGO VIDEOSCOPE HAD OBJECTIVE LENS MISSING GLUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 24020893 ↗
  • MalfunctionDec 30, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED THAT AN OILY, WAXY RESIDUE WAS FOUND ON MULTIPLE OLYMPUS ENF-V3/RHINO-LARYNGO VIDEOSCOPE AND ENF-VH/RHINO-LARYNGO VIDEOSCOPE FOLLOWING REPROCESSING IN THE MEDIVATOR AUTOMATED ENDOSCOPE REPROCESS. PRIMARILY THOSE REPROCESSED ON (B)(6) 2025. THE FACILITY NOTED THAT ALCOHOL REMOVED THE RESIDUE COMPLETELY, WHILE ENDOZYME ONLY PARTIALLY IMPROVED IT. NO PROCEDURE DELAYS WERE REPORTED. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23920778 ↗