OES URETERO-RENO FIBERSCOPE OLYMPUS URF TYPE P5 (UDI 04953170316593)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of OES URETERO-RENO FIBERSCOPE (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170316593), received between Mar 4, 2021 and Nov 14, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation2
- Break1
- Mechanical Problem1
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 14, 2023Mechanical Problem; Poor Quality Image; Material Deformation; Mechanical Problem; Poor Quality Image; Material Deformation
THE CUSTOMER REPORTED OES URETERO-RENO FIBERSCOPE WAS HAVING IMAGE ISSUE AND MULTIPLE BLACK DOTS ON THE VISION, BROKEN FIBER IMAGE PROBLEM, AND ANGULATION WAS NOT GOOD AND COULD NOT ANGULATE SUFFICIENTLY. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS IDENTIFIED: FORCEPS CHANNEL PORT WAS DEFORMED. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18130020 ↗MalfunctionMay 16, 2022Material Deformation; Material Deformation
A USER FACILITY REPORTED TO OLYMPUS, THE OES URETERO-RENO FIBERSCOPE EXHIBITED MULTIPLE BLACK DOTS IN THE IMAGE, AND ANGULATION ISSUE. UPON INSPECTION AND TESTING OF THE CUSTOMER RETURNED DEVICE, THE SCOPE BENDING SECTION REPEATEDLY ANGULATED IN THE DOWNWARD DIRECTION DUE TO DEFORMATION OF THE BENDING SECTION. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14413631 ↗MalfunctionMar 4, 2021Break; Break
OLYMPUS INSPECTED THE DEVICE AT THE SERVICE DEPARTMENT OF OLYMPUS (B)(4) AND FOUND THAT THE BENDING SECTION KEPT ANGULATED, DUE TO THE BROKEN ANGLE WIRE CAUSING THE END TO COME APART. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11409981 ↗
Other Choledochoscope and accessories, flexible/rigid devices
- SpyScope DS M00546600Boston Scientific Corporation · Choledochoscope and accessories, flexible/rigid912 reports
- SpyScope™ DS II M00546610Boston Scientific Corporation · Choledochoscope and accessories, flexible/rigid2 deaths646 reports
- Forceps/Irrigation Plug (Isolated Type) MAJ-891Olympus Medical Systems Corp. · Choledochoscope and accessories, flexible/rigid1 recall1 death257 reports
- OES CHOLEDOCHOFIBERSCOPE OLYMPUS CHF TYPE CB30LOlympus Medical Systems Corp. · Choledochoscope and accessories, flexible/rigid1 recall9 reports
- SpyGlass™ Discover M00546780Boston Scientific Corporation · Choledochoscope and accessories, flexible/rigid1 death55 reports
- Tangent Digital catheter tng-4002-indTangent Endoscopy LLC · Choledochoscope And Accessories, Flexible/Rigid1 recall0 reports
- OES TRANSDUODENAL CHOLEDOCHOFIBERSCOPE OLYMPUS CHF TYPE BP30Olympus Medical Systems Corp. · Choledochoscope and accessories, flexible/rigid1 recall0 reports
- SpyGlass DS M0054665R0Boston Scientific Corporation · Choledochoscope and accessories, flexible/rigid19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.