OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE 3C40 (UDI 04953170339219)Recalls & reported problems
Olympus Medical Systems Corp. · BF-3C40 3.6MM DIA 1.2MM CH 55CM LENGTH · Class 2
All models of OES BRONCHOFIBERSCOPE (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Olympus BF-3C40 bronchoscope is a thin flexible scope used to look inside the lungs and airways. Most of the 46 reports describe the scope's coating peeling or flaking off, along with cleaning and sterilization difficulties; 12 reports involved injuries and none involved deaths, with several mentioning fungal infection in patients. One recall, started in July 2023 and still open, concerns the scope being too wide and getting stuck in an endotracheal tube connector.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 46
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 04953170339219), received between Sep 10, 2021 and Feb 7, 2025 — classified as 12 injuries, 34 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (28 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 27, 2023Class IIOpen, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2502-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Peeled/Delaminated28
- Contamination /Decontamination Problem6
- Adverse Event Without Identified Device or Use Problem4
- Failure to Clean Adequately4
- Contamination2
- Break1
- Degraded1
- Device Reprocessing Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions34
- Fungal Infection10
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
Recent adverse event reports
The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2025Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THAT DURING REPROCESSING, THE FLEXIBLE SCOPE DID NOT PASS STERILIZATION, TESTED ON 3 DIFFERENT STERILIZERS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21337159 ↗MalfunctionOct 23, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE BRONCHOFIBERSCOPE COATING OF IG HOSE PEELED OFF (1MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20517511 ↗MalfunctionOct 23, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED, DURING DEVICE EVALUATION. THAT THE COATING ON THE IMAGE GUIDE HOSE OF THE BRONCHOFIBERSCOPE WAS PEELED OFF. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20513143 ↗MalfunctionAug 21, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE COATING OF LIGHT GUIDE HOSE ON THE OES BRONCHOFIBERSCOPE HAD PEELED OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20020495 ↗MalfunctionMar 25, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BRONCHOFIBERSCOPE EXHIBITED FOREIGN OBJECTS IN THE ANGULATION LEVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT. ADDITIONAL DETAILS RELATING TO THE PATIENT AND THE EVENT HAVE BEEN REQUESTED, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968817 ↗MalfunctionMar 25, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BRONCHOFIBERSCOPE EXHIBITED PEELING OF COATING OF THE IMAGE GUIDE PIPE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968192 ↗MalfunctionMar 23, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BRONCHOFIBERSCOPE EXHIBITED FOREIGN MATERIAL ON THE BENDING SECTION COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18965482 ↗MalfunctionFeb 5, 2024Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOFIBERSCOPE IMAGED LOOKED BLURRY. THE DEVICE WAS RETURNED FOR EVALUATION. DURING TESTING AND INSPECTION OF THE DEVICE, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: LIGHT GUIDE LENS HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18643853 ↗MalfunctionDec 21, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THE OES BRONCHOFIBERSCOPE HAS FALLEN RUBBER AT THE DISTAL END, TIGHTENED UPPER SHEATH. THIS ISSUE WAS FOUND DURING AN UNSPECIFIED PROCEDURE. THERE WAS NO DELAY AND THE PROCEDURE WAS COMPLETED BY USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18376003 ↗MalfunctionDec 12, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOFIBERSCOPE HAD SCRATCHES ON THE UNIVERSAL CORD, MEANDERING OF THE INSERTION PART, DISCOLORATION OF THE CONNECTOR AND OPERATION PARTS. THE DEVICE WAS RETURNED FOR EVALUATION. DURING TESTING AND INSPECTION OF THE DEVICE, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: COAT OF THE IMAGE GUIDE PIPE WAS PEELING OFF (1X1MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18311714 ↗MalfunctionNov 29, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THAT THE OES BRONCHOFIBERSCOPE HAD A WATER LEAK. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, IT WAS FOUND THAT INSERTION SECTION FLEXIBLE TUBE COATING WAS PEELING OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18233761 ↗MalfunctionOct 27, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER ORIGINALLY RETURNED HIS OLYMPUS OES BRONCHOVIBERSCOPE DUE TO AN ANGULATION MALFUNCTION. THERE WAS NO PATIENT HARM REPORTED AS A RESULT OF THIS EVENT. DURING THE DEVICE EVALUATION, IT WAS DISCOVERED THAT THE COATING MATERIAL WAS PEELING. THIS REPORT IS BEING SUBMITTED TO CAPTURE THE PEELING COATING MATERIAL FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18021379 ↗MalfunctionOct 19, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE OES BRONCHOFIBERSCOPE HAD PEELING OFF/STICKINESS OF THE INSERTION PART COATING AND SEPARATION. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE IMAGE GUIDE PIPE WAS PEELING OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17965972 ↗MalfunctionSep 20, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE OES BRONCHOFIBERSCOPE EPERIENCED AIR/WATER LEAKAGES. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE INSERTION PART COATING PEELING OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17781023 ↗MalfunctionSep 8, 2023Peeled/Delaminated; Peeled/Delaminated
A USER FACILITY SUBMITTED A REPAIR REQUEST TO THE OLYMPUS SERVICE CENTER, FOR AN OES BRONCHOFIBERSCOPE PROTECTOR OF UNIVERSAL CORD ON SCOPE CONNECTOR SIDE DAMAGED. UPON INSPECTION AND TESTING OF THE CUSTOMER RETURNED DEVICE, THE SCOPE INSERTION PART COATING WAS FOUND PEELED OFF. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO REPORT OF PATIENT HARM OR USER INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17715768 ↗MalfunctionSep 8, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER INITIALLY REPORTED DEFECTS IN RUBBER ON THE OES BRONCHOVIDEOSCOPE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS IDENTIFIED: THE COAT OF THE IG SERPENTINE IS PEELED OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17710868 ↗MalfunctionAug 29, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED THAT WHEN USING AN OES BRONCHOFIBERSCOPE THERE WAS A LEAKAGE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE DEVICE WAS RETURNED AND EVALUATED, AND UPON ESTIMATION, THE CONNECTING TUBE HAD COATING PEELING (1MM2 OR MORE). THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17642239 ↗MalfunctionAug 23, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED OLYMPUS THAT THE OES BRONCHOFIBERSCOPE HAD LEAK. THERE WERE NO REPORTS OF PATIENT HARM. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: PEELING OF THE COATING IN THE INSERTION PART OF THE CONNECTING TUBE (1MM SQUARED OR MORE)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17607404 ↗MalfunctionJul 7, 2023Degraded; Peeled/Delaminated; Degraded; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THAT THERE WERE DEFECTS OF THE BENDING SECTION AND INSERTION TUBE ON THE OES BRONCHOFIBERSCOPE. DURING INSPECTION AND TESTING OF THE RETURNED DEVICE IT WAS FOUND THAT THE CONNECTING TUBE HAD COATING THAT WAS PEELING. THERE WAS NO REPORT OF DELAY OR HARM TO A PATIENT OR USER. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17279346 ↗MalfunctionJun 27, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THAT THE BENDING OPERATION OF THE OES BRONCHOFIBERSCOPE WAS DISABLED. THERE WAS NO REPORT OF PATIENT HARM. THE DEVICE WAS RETURNED, EVALUATED. AND THE IT WAS FOUND, THAT THE COATING OF THE IMAGE GUIDE SERPENTINE WAS PEELING OFF (1MM2 OR MORE). THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17209517 ↗MalfunctionJun 7, 2023Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED TO OLYMPUS, THAT THE OES BRONCHOFIBERSCOPE HAD ANGULATION MALFUNCTION. INSPECTION AND TESTING OF THE RETURNED DEVICE FOUND THAT THE CONNECTING TUBE COATING WAS PEELING. THERE WERE NO REPORTS HARM ASSOCIATED WITH THE EVENT. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17080828 ↗MalfunctionMay 5, 2023Peeled/Delaminated; Peeled/Delaminated
A DISTRIBUTOR REPORTED ON BEHALF OF THE CUSTOMER, THAT THEIR OES BRONCHOFIBERSCOPE HAD AN ANGULATION MALFUNCTION. UPON INSPECTION AND TESTING OF THE RETURNED UNIT, IT WAS DISCOVERED, THAT THE COATING OF THE CONNECTING TUBE WAS PEELING. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16881222 ↗MalfunctionApr 25, 2023Peeled/Delaminated; Peeled/Delaminated
A USER FACILITY SUBMITTED A REPAIR REQUEST TO THE OLYMPUS SERVICE CENTER, FOR A CYSTO-NEPHRO VIDEOSCOPE, HAVING AIR/WATER LEAKING FROM THE SCOPE ANGLE RUBBER. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, COATING ON THE SCOPE IMAGE GUIDE SNAKE TUBE WAS FOUND PEELING. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO REPORT OF PATIENT HARM OR USER INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16811529 ↗MalfunctionMar 10, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS THAT THE OES BRONCHOFIBERSCOPE HAD A PINHOLE SIZE IN THE COIL. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE SUBJECT DEVICE AS SUBSEQUENTLY EVALUATED BY OLYMPUS AND THE COATING ON THE INSERTION SECTION OF THE TUBE WAS FOUND TO BE PEELING. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16523989 ↗MalfunctionMar 3, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, DURING AN UNSPECIFIED PROCEDURE, AN ANGULATION ISSUE OCCURRED ON THE OES BRONCHOFIBERSCOPE. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT. DURING EVALUATION OF THE RETURNED DEVICE, THE EVALUATION FOUND THAT THE COATING OF THE IG SERPENTINE IS PEELING OFF BY 1 - 2 MM OR MORE. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16479467 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page