2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus Visera tracheal intubation videoscope is a flexible bronchoscope. The 77 reports mostly describe malfunction, often its coating peeling or parts detaching or cracking, with no deaths and 1 injury. Two recalls are open: unvalidated channel drying could leave water that risks contamination, and labeling now excludes it from certain reprocessors. Reports rose to 10 in the last 12 months from 5.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
77
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 77 adverse event reports that list this exact device (DI 04953170340246), received between Jun 17, 2021 and Aug 17, 2026 — classified as 1 injury, 76 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (52 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0339-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. May 26, 2023Class IIOpen, Classified

    Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

    Z-2340-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Peeled/Delaminated52
  • Detachment of Device or Device Component8
  • Failure to Clean Adequately4
  • Scratched Material4
  • Degraded3
  • Loose or Intermittent Connection3
  • Material Split, Cut or Torn2
  • Component Missing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions77

Recent adverse event reports

The 25 most recent of 77 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2026Degraded; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50279667 ↗
  • MalfunctionJul 29, 2026Image Display Error/Artifact

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50076313 ↗
  • MalfunctionMay 1, 2026Detachment of Device or Device Component

    THE VISERA TRACHEAL INTUBATION VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF FAILED LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, THE GLUE ON THE LIGHT GUIDE LENS WAS CHIPPED WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25061930 ↗
  • MalfunctionApr 29, 2026Degraded

    THE TRACHEAL INTUBATION VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF FAILED LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, CHIPPING OFF AT DISTAL END SIDE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25031357 ↗
  • MalfunctionApr 17, 2026Contamination; Obstruction of Flow

    THE TRACHEAL INTUBATION VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF AN OBSTRUCTED ORIFICE. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, FOREIGN MATERIAL STUCK IN THE BIOPSY PORT CAUSING OBSTRUCTED BRUSH PASSAGE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24918933 ↗
  • MalfunctionApr 1, 2026Material Fragmentation

    THE VISERA TRACHEAL INTUBATION VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF "IT HAS A HOLE IN IT, THEY NOTICED IT WHILE CLEANING. DOWN TOWARDS THE END OF THE TRIP". THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, THE OBJECTIVE LENS AND THE LIGHT GUIDE WAS FOUND CHIPPED. THERE WAS NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24754310 ↗
  • MalfunctionMar 25, 2026Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE VISERA TRACHEAL INTUBATION VIDEOSCOPE HAD THE LIGHT GUIDE LENS PARTIALLY MISSING GLUE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24695169 ↗
  • MalfunctionJan 9, 2026Detachment of Device or Device Component

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE TRACHEAL INTUBATION VIDEOSCOPE HAD A DETACHED DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24020185 ↗
  • MalfunctionJan 2, 2026Corroded

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE VISERA TRACHEAL INTUBATION VIDEOSCOPE HAD CORROSION UNDER BENDING SECTION COVER AT LEAK HOLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23950339 ↗
  • InjuryNov 4, 2025Material Separation; Detachment of Device or Device Component; Material Split, Cut or Torn; Material Separation; Detachment of Device or Device Component; Material Split, Cut or Torn

    NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23471850 ↗
  • MalfunctionSep 18, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE TRACHEAL INTUBATION VIDEOSCOPE HAD DIRT ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23084762 ↗
  • MalfunctionAug 22, 2025Component Missing; Component Missing

    IT WAS REPORTED THE VIDEOSCOPE WAS MISSING A RING ON THE BIOPSY CHANNEL AND THEY WERE UNABLE TO ATTACH THE DISPOSABLE BIOPSY PORT. THE ISSUE WAS IDENTIFIED DURING SETUP/INSPECTION PRIOR TO CLINICAL USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22877232 ↗
  • MalfunctionJul 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE INTUBATION VIDEOSCOPE'S TIP COVER CAME OFF. THE EVENT OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22476070 ↗
  • MalfunctionDec 29, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED THE TRACHEAL INTUBATION VIDEOSCOPE HAD BLACK DEBRIS IN THE ENDOTRACHEAL TUBE. THE ISSUE OCCURRED DURING A THERAPEUTIC PRE-SURGICAL BRONCHOSCOPY BY ANESTHESIOLOGIST, WHILE THE DEVICE WAS INSIDE THE PATIENT. THERE WAS A PROCEDURAL DELAY OF LESS THAN 30 MINUTES UNDER SEDATION. THE PROCEDURE WAS COMPLETED USING AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21032438 ↗
  • MalfunctionDec 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20879086 ↗
  • MalfunctionOct 18, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE TRACHEAL INTUBATION VIDEOSCOPE EXHIBITED THE COATING ON THE CONNECTING TUBE HAS PEELED OFF (1MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483548 ↗
  • MalfunctionOct 16, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE VISERA TRACHEAL INTUBATION VIDEOSCOPE EXHIBITED THE COATING OF THE IMAGE GUIDE HOSE HAS PEELED OFF (1MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20460423 ↗
  • MalfunctionOct 3, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE VIDEOSCOPE COATING ON THE IMAGE GUIDE TUBE HAD PEELED OFF (1 MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20365262 ↗
  • MalfunctionSep 6, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE TRACHEAL INTUBATION VIDEOSCOPE COATING ON THE CONNECTING TUBE WAS FOUND PEELED OFF (1 MM2 OR MORE). THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR USER INJURY ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20163348 ↗
  • MalfunctionSep 5, 2024Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEOSCOPE FORCEPS PLUG NOZZLE WAS LOOSE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20145997 ↗
  • MalfunctionJun 28, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE TRACHEAL INTUBATION VIDEOSCOPE EXHIBITED A PEELED COATING ON THE FLEXIBLE TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19635763 ↗
  • MalfunctionJun 25, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE TRACHEAL INTUBATION VIDEOSCOPE EXHIBITED PEELING-OFF OF THE COATING OF THE IMAGE GUIDE HOSE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19603966 ↗
  • MalfunctionJun 11, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE TRACHEAL INTUBATION VIDEOSCOPE EXHIBITED THE COAT OF THE IMAGE GUIDE SNAKE PIPE IS PEELED OFF (MORE THAN 1MM2). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19508874 ↗
  • MalfunctionMay 21, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE INTUBATION VIDEOSCOPE HAD EXHIBITED THE INDICATION ON THE CONNECTING TUBE HAD A DISAPPEARED AREA (1MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19360742 ↗
  • MalfunctionMay 10, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SUBJECT DEVICE EXHIBITED A PEELED IMAGE GUIDE TUBE COATING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19293250 ↗