ULTRASONIC GASTROVIDEOSCOPE OLYMPUS GF TYPE UC140P-AL5 (UDI 04953170356322)Recalls & reported problems

Olympus Medical Systems Corp. · GF-UC140P-AL5 CLA EUS scope, 2.8mm chan · Class 2

All models of ULTRASONIC GASTROVIDEOSCOPE (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus ultrasonic endoscope is used for endoscopic ultrasound. Reports describe peeled or cracked parts, missing images and difficult cleaning. There are 114 reports, including 7 injuries and no deaths, and reports in the last 12 months rose to 44 from 6. There are 2 open recalls, one covering scopes returned with a wet inner channel and the other a scope no longer compatible with reprocessing, both posing infection risk.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
114
Deaths reported
0
Injuries reported
6
Recalls naming it
2

Summary

The FDA's MAUDE database contains 114 adverse event reports that list this exact device (DI 04953170356322), received between May 15, 2020 and Sep 29, 2026 — classified as 6 injuries, 108 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (22 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 5 reportsDec 2025: 11 reportsJan 2026: 7 reportsFeb 2026: 2 reportsMar 2026: 3 reportsApr 2026: 4 reportsMay 2026: 2 reportsJun 2026: 5 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0333-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Mar 18, 2022Class IIOpen, Classified

    GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients

    Z-1034-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Peeled/Delaminated22
  • Crack12
  • Failure to Clean Adequately9
  • No Display/Image9
  • Detachment of Device or Device Component7
  • Scratched Material6
  • Adverse Event Without Identified Device or Use Problem5
  • Break5

Patient problems

  • No Clinical Signs, Symptoms or Conditions100
  • No Patient Involvement7
  • Thrombocytopenia5
  • No Consequences Or Impact To Patient2
  • Biliary Cirrhosis1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 114 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939083 ↗
  • MalfunctionSep 18, 2026Detachment of Device or Device Component; Scratched Material

    The customer reported that the ultrasonic gastrovideoscope had a separate issue. During the initial evaluation of the returned device, deep scratches on the probe unit and missing thixotropic adhesive (KE45) on the light guide (LG) cover were observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50817439 ↗
  • MalfunctionAug 25, 2026No Display/Image; Scratched Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50442288 ↗
  • MalfunctionJul 14, 2026Break

    THE CUSTOMER REPORTED DISTAL TIP BROKEN INVOLVING AN OLYMPUS ULTRASONIC GASTROVIDEOSCOPE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25812462 ↗
  • MalfunctionJul 8, 2026Material Fragmentation

    THE CUSTOMER RETURNED THE ULTRASONIC GASTROVIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING, THE SERVICE REPAIR TECHNICIAN IDENTIFIED A CHIP AT THE EDGE OF THE LIGHT GUIDE AND OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25746783 ↗
  • MalfunctionJun 26, 2026Crack

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE EXHIBITED PROBE UNIT CRACK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25633795 ↗
  • MalfunctionJun 5, 2026Material Fragmentation

    THE CUSTOMER RETURNED THE ULTRASONIC GASTROVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE DEVICE EXHIBITED A CHIP IN THE ADHESIVE AROUND THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25411546 ↗
  • MalfunctionJun 5, 2026Material Fragmentation

    THE CUSTOMER RETURNED THE ULTRASONIC GASTROVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE DEVICE EXHIBITED CHIPPED ADHESIVE AROUND THE OBJECTIVE LENS AND LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25411356 ↗
  • MalfunctionJun 3, 2026Break

    THE CUSTOMER RETURNED THE ULTRASONIC GASTROVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE INVESTIGATOR INDICATES THAT THE ADHESIVE AROUND THE LENS HAD CHIPPED, THE CAUSE OF WHICH WAS TRACED TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25376389 ↗
  • MalfunctionJun 1, 2026Material Separation

    THE CUSTOMER RETURNED THE ULTRASONIC GASTROVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS MISSING ADHESIVE ON THE OBJECTIVE LENS WAS OBSERVED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25352849 ↗
  • MalfunctionMay 29, 2026No Display/Image; Material Fragmentation

    THE CUSTOMER RETURNED THE ULTRASONIC GASTROVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE DEVICE EXHIBITED CHIPPING AT THE EDGE OF THE OBJECTIVE LENS AND NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25336822 ↗
  • MalfunctionMay 5, 2026Detachment of Device or Device Component

    DURING THE DEVICE EVALUATION OF THE ULTRASONIC GASTROVIDEOSCOPE, IT WAS OBSERVED THAT THE SEALANT WAS MISSING. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25085621 ↗
  • MalfunctionApr 23, 2026Crack

    DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE PRESENTED CHIPPED GLUE AROUND ITS OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24971014 ↗
  • MalfunctionApr 23, 2026Dent in Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE EXHIBITED DEEP DENTS ON THE PROBE TIP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24968497 ↗
  • MalfunctionApr 12, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE HAD A LARGE CHIP AROUND THE OBJECTIVE (OB) LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24852971 ↗
  • MalfunctionApr 10, 2026Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE EXHIBITED PEELING OF CEMENT AROUND THE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24845100 ↗
  • MalfunctionMar 23, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE HAD A GLUE MISSING ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24664362 ↗
  • MalfunctionMar 4, 2026Crack

    IT WAS REPORTED THE ULTRASONIC GASTROVIDEOSCOPE RIGHT ABOVE THE ELEVATOR AT THE DISTAL TIP HAD A CRACK. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24505705 ↗
  • MalfunctionMar 2, 2026Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24483557 ↗
  • MalfunctionFeb 11, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASOUND GASTROVIDEOSCOPE HAD A BROKEN PROBE CABLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24312292 ↗
  • MalfunctionFeb 10, 2026No Display/Image

    IT WAS REPORTED THAT THE ULTRASONIC GASTROVIDEOSCOPE WAS LOSING THE IMAGE DURING A THERAPEUTIC PROCEDURE (ENDOSCOPIC ULTRASOUND). THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24306334 ↗
  • MalfunctionJan 29, 2026Crack

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE PROBE UNIT WAS CRACKED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24210963 ↗
  • MalfunctionJan 22, 2026Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE EXHIBITED PEELING CEMENT AT LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24144966 ↗
  • MalfunctionJan 20, 2026Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE EXHIBITED PEELED CEMENT AROUND THE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24126832 ↗
  • MalfunctionJan 8, 2026Crack

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC GASTROVIDEOSCOPE EXHIBITED THE DISTAL-END PROBE UNIT CRACKED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24004789 ↗

Other Transducer, ultrasonic, diagnostic devices