ULTRASONIC GASTROVIDEOSCOPE recallZ-1034-2022 · Olympus Corporation of the Americas · initiated Mar 18, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients

FDA root cause category: Device Design

What the company is doing

Olympus issued URGENT: MEDICAL DEVICE CORRECTIVE ACTION on 3/18/22. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory for the referenced devices and identify any device with the model number specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. The model number can be found on the device as illustrated in the following picture. 2. If you are using the OER-Pro for high level disinfection of the GF-UC140P-AL5, please transition to an OER-Elite or another AER validated for GF-UC140P-AL5 reprocessing as soon as possible. 3. When manually cleaning the GF-UC140P-AL5 carefully follow the instructions in the IFU, vigilantly cleaning all channels and the forceps elevator. Ensure that you are inspecting the GFUC140P-AL5 prior to each use and after each use according to the instructions. 4. Ensure all personnel understand that the GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and further that such personal are completely knowledgeable and thoroughly trained on the Instructions for Use applicable to the GF-UC140P-AL5 device 5. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number 0406 and provide your contact information as indicated in the portal. If you require additional information about the GF-UC140P-AL5 reprocessing, you can obtain additional information as follows: You can contact our Technical Assistance Center (TAC) at 1-800-848-9024, option 1 and the TAC department can assist you in answering questions. OMSC requests you to report any patient injuries, including infections or persistent microbial colonization associated with any OMSC endoscope. Call our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1 to report complaints.

Recall details

Recall number
Z-1034-2022
Status
Open, Classified
Initiated by the company
Mar 18, 2022
Posted by the FDA
May 5, 2022
Recalling firm
Olympus Corporation of the Americas
Quantity
731 units
Product code
ITX
FDA submissions
K011314
Distribution
Nationwide

Product description

Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging.

Code information (models, lots, serial numbers)

All serial numbers. GTIN: 04953170356322