Skip to content

Single Use 3-Lumen Extraction Balloon V B-V233P-BRecalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of Single Use 3-Lumen Extraction Balloon V (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170368714), received between Jun 19, 2024 and Jun 19, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Break” (4 reports).

3 other recalls cover devices cleared under the same FDA submission (K091495); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 4 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 19, 2024Break; Break

    IT WAS REPORTED THAT TWO BALLOONS BROKE AND FELL IN THE PATIENT'S BODY. THE FRAGMENTS WERE SWEPT OUT OF THE DUCT WITH ANOTHER EXTRACTION BALLOON AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SIMILAR DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19570693 ↗
  • InjuryJun 19, 2024Break; Break

    IT WAS REPORTED THAT TWO BALLOONS BROKE AND FELL IN THE PATIENT'S BODY. THE FRAGMENTS WERE SWEPT OUT OF THE DUCT WITH ANOTHER EXTRACTION BALLOON AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SIMILAR DEVICE. THIS REPORT IS RELATED TO LINKED PATIENT IDENTIFIER (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19570678 ↗
  • InjuryJun 19, 2024Break; Break

    IT WAS REPORTED THAT TWO BALLOONS BROKE AND FELL IN THE PATIENT'S BODY. THE FRAGMENTS WERE SWEPT OUT OF THE DUCT WITH ANOTHER EXTRACTION BALLOON AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SIMILAR DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19570508 ↗
  • InjuryJun 19, 2024Break; Break

    IT WAS REPORTED THAT TWO BALLOONS BROKE AND FELL IN THE PATIENT'S BODY. THE FRAGMENTS WERE SWEPT OUT OF THE DUCT WITH ANOTHER EXTRACTION BALLOON AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SIMILAR DEVICE. THIS REPORT IS RELATED TO LINKED PATIENT IDENTIFIER (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19570474 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K091495. One submission often covers many models, so check the product and lot details in each record.

  • Z-0968-2021 · Jan 4, 2021 · Terminated
    B-V443Q-A 3-LUMEN EXTRACTION BALLOON V, model no. B-V443Q-A - Product Usage: Used with endoscopes to endoscopically remove stones such as calculi, pancreatic and common bile duct s
  • Z-0966-2021 · Jan 4, 2021 · Terminated
    B-V233P-A 3-LUMEN EXTRACTION BALLOON V, model no. B-V233P-A - Product Usage: Used with endoscopes to endoscopically remove stones such as calculi, pancreatic and common bile duct s
  • Z-0967-2021 · Jan 4, 2021 · Terminated
    B-V243Q-A 3-LUMEN EXTRACTION BALLOON V, model no. B-V243Q-A - Product Usage: Used with endoscopes to endoscopically remove stones such as calculi, pancreatic and common bile duct s

Other DISLODGER, STONE, BILIARY devices

All DISLODGER, STONE, BILIARY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page