DISPOSABLE GRASPING FORCEPS FG-51DRecalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of DISPOSABLE GRASPING FORCEPS (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04953170375040), received between Jun 6, 2022 and Jan 15, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Close” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Open or Close4
- Deformation Due to Compressive Stress2
- Material Split, Cut or Torn2
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2025
ATTEMPTED TO PULL THE BASKET BACK INTO THE SHEATH TO INSERT DEVICE DOWN BRONCHOSCOPY SCOPE, BUT BASKET WOULD NOT MOVE DESPITE CONTROL HANDLE OPENING AND CLOSING. UNABLE TO USE THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21157836 ↗MalfunctionDec 9, 2022Difficult to Open or Close; Difficult to Open or Close
THE CUSTOMER REPORTED THE BASKET ON THE SUBJECT DEVICE DID NOT OPEN DURING USE IN A BRONCHOSCOPY PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THE SUBJECT DEVICE WAS SENT TO OLYMPUS FOR EVALUATION. DURING INSPECTION AND TESTING, THE BASKET COULD NOT BE CLOSED. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION (UNABLE TO CLOSE BASKET). THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15950507 ↗MalfunctionSep 13, 2022Material Twisted/Bent; Material Twisted/Bent
THE CUSTOMER REPORTED THAT, DURING PREPARATION FOR AN UNSPECIFIED PROCEDURE, THE CATHETER WAS BROKEN AT THE TOP OF THE HANDLE AND THE BASKET COULD NOT BE RETRACTED OR EXTENDED. TWO ADDITIONAL DEVICES WERE OPENED AND TESTED WITH THE SAME RESULT. THERE WAS NO EFFECT ON THE PATIENT DUE TO THE EVENT. THIS REPORT IS FOR THE THIRD DEVICE. THE FIRST DEVICE IS BEING REPORTED ON THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6). THE SECOND DEVICE IS BEING REPORTED ON THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15411418 ↗MalfunctionSep 9, 2022Deformation Due to Compressive Stress; Material Split, Cut or Torn; Deformation Due to Compressive Stress; Material Split, Cut or Torn
AN OLYMPUS REPRESENTATIVE REPORTED TO OLYMPUS, ON BEHALF OF THE USER, PRIOR TO AN UNKNOWN PROCEDURE, THREE DISPOSABLE GRASPING FORCEPS COULD NOT BE OPENED AND/OR CLOSED. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. IT WAS FURTHER REPORTED, THIS TYPE OF PROCEDURE TAKES 14 MINUTES, HOWEVER; THE START OF THIS PROCEDURE WAS DELAYED BY 30 MINUTES DUE TO THIS EVENT. THERE WAS NO PATIENT HARM ASSOCIATED WITH THIS EVENT. THE FIRST GRASPING FORCEPS IS BEING REPORTED ON THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6). THE SECOND GRASPING FORCEPS IS BEING REPORTED ON THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15392064 ↗MalfunctionSep 9, 2022Deformation Due to Compressive Stress; Material Split, Cut or Torn; Deformation Due to Compressive Stress; Material Split, Cut or Torn
AN OLYMPUS REPRESENTATIVE REPORTED TO OLYMPUS, ON BEHALF OF THE USER, PRIOR TO AN UNKNOWN PROCEDURE, THREE DISPOSABLE GRASPING FORCEPS COULD NOT BE OPENED AND/OR CLOSED. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. IT WAS FURTHER REPORTED, THIS TYPE OF PROCEDURE TAKES 14 MINUTES, HOWEVER; THE START OF THIS PROCEDURE WAS DELAYED BY 30 MINUTES DUE TO THIS EVENT. THERE WAS NO PATIENT HARM ASSOCIATED WITH THIS EVENT. THE FIRST GRASPING FORCEPS IS BEING REPORTED ON THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6). THE THIRD GRASPING FORCEPS IS BEING REPORTED ON THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15392025 ↗MalfunctionJul 1, 2022Difficult to Open or Close; Difficult to Open or Close
AN OLYMPUS REPRESENTATIVE REPORTED TO OLYMPUS, ON BEHALF OF THE USER, PRIOR TO AN UNKNOWN PROCEDURE, THREE DISPOSABLE GRASPING FORCEPS COULD NOT BE OPENED AND/OR CLOSED. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. IT WAS FURTHER REPORTED, THIS TYPE OF PROCEDURE TAKES 14 MINUTES, HOWEVER; THE START OF THIS PROCEDURE WAS DELAYED BY 30 MINUTES DUE TO THIS EVENT. THERE WERE NO PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14894051 ↗MalfunctionJun 6, 2022Difficult to Open or Close; Difficult to Open or Close
THE CUSTOMER REPORTED THAT WHEN THE DISPOSABLE GRASPING FORCEPS BASKET WAS OPENED AND CLOSED, THE GRIP PART FAILED TO CLOSE AFTER OPENING AND COULD NOT BE STORED IN THE SHEATH. TWO SIMILAR DEVICES WERE USED; HOWEVER, THE ISSUE PERSISTED. THE EVENT OCCURRED DURING PREPARATION FOR USE AND THE PROCEDURE WAS COMPLETED WITH A FOURTH DEVICE SUCCESSFULLY. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT. THIS MEDWATCH REPORT WITH PATIENT IDENTIFIER (B)(6) IS THE SECOND OCCURRENCE. THE MEDWATCH REPORT WITH PATIENT IDENTIFIER (B)(6) IS THE FIRST OCCURRENCE. THE MEDWATCH REPORT WITH PATIENT IDENTIFIER (B)(6) IS THE THIRD OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14601748 ↗MalfunctionJun 6, 2022Difficult to Open or Close; Difficult to Open or Close
THE CUSTOMER REPORTED THAT WHEN THE DISPOSABLE GRASPING FORCEPS BASKET WAS OPENED AND CLOSED, THE GRIP PART FAILED TO CLOSE AFTER OPENING AND COULD NOT BE STORED IN THE SHEATH. TWO SIMILAR DEVICES WERE ATTEMPTED; HOWEVER, THE ISSUE PERSISTED. THE EVENT OCCURRED DURING PREPARATION FOR USE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A FOURTH DEVICE . THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT. THIS MEDWATCH REPORT WITH PATIENT IDENTIFIER (B)(6) IS THE FIRST OCCURRENCE. THE MEDWATCH REPORT WITH PATIENT IDENTIFIER (B)(6) IS THE SECOND OCCURRENCE. THE MEDWATCH REPORT WITH PATIENT IDENTIFIER (B)(6) IS THE THIRD OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14601644 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page