DISPOSABLE BIOPSY FORCEPS FB-211D (REF N5431830)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of DISPOSABLE BIOPSY FORCEPS (23)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170389535), received between Jul 15, 2019 and Dec 27, 2023 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • Pneumothorax1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED: THE PATIENT HAS HAD AN MRI AND IT WAS REPORTED THAT WHAT THE USER HAD SEEN WAS SOMETHING FROM A BREAST SURGERY AND NOT FROM THE LUNG BRONCHOSCOPY.

    Patient problems: Foreign Body In Patient

    MAUDE report 18396977 ↗
  • MalfunctionMay 19, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE DISPOSABLE BIOPSY FORCEPS JAWS BROKE INSIDE THE PATIENT'S LUNGS. ON EXAMINATION, AN INTERNAL WIRE PROTRUDES FROM THE FORCEPS WHEN THE JAWS OPEN, POTENTIALLY LEADING TO UNINTENDED INJURY TO THE PATIENT'S AIRWAY. NO PART OF THE FORCEPS FELL INTO THE PATIENT'S LUNGS. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE AFTER A 4-MINUTE DELAY. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16968180 ↗
  • InjurySep 16, 2022Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, DURING A BRONCHOSCOPY, THE BIOPSY FORCEPS BROKE. WHEN THE FORCEPS BROKE, THE WIRES THAT MOVE THE FORCEPS BLADES WERE RELEASED AND PROTRUDED ABOUT 1.5 CM CAUSING DAMAGE TO THE PATIENT AND THE NON-OLYMPUS ENDOSCOPE, WHEN OPENING AND CLOSING AND PERFORMING MOVEMENTS FROM THE HANDLE. THE PROCEDURE WAS COMPLETED WITH SIMILAR FORCEPS. ON THE SAME DAY OF THE PROCEDURE WITH DEVICE MALFUNCTION, THE PATIENT PRESENTED WITH SYMPTOMS OF A PNEUMOTHORAX. THE PNEUMOTHORAX WAS TREATED AND THE PATIENT REMAINED IN THE HOSPITAL. THE CUSTOMER REPORTED IT IS UNKNOWN IF THE EVENT IS DIRECTLY RELATED TO DEVICE MALFUNCTION DURING THE PROCEDURE.

    Patient problems: Pneumothorax

    MAUDE report 15437634 ↗
  • MalfunctionJul 15, 2019Break; Break

    SERVICE CENTER WAS INFORMED THAT DURING A BIOPSY, THE FORCEPS WERE REMOVED, AND MALFUNCTIONED UPON DEPOSITING THE BIOPSY IN A SPECIMEN CUP. A NEW SET OF FORCEPS WAS USED FOR ADDITIONAL BIOPSIES AND THE PROCEDURE WAS CONCLUDED WITHOUT ANY COMPLICATIONS. THERE WERE NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8790952 ↗