RHINO-LARYNGO FIBERSCOPE ENF-GP2 (UDI 04953170399282)Recalls & reported problems
Olympus Medical Systems Corp. · RHINO-LARYNGO FIBERSCOPE ENF-GP2 · Class 2
All models of RHINO-LARYNGO FIBERSCOPE (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 04953170399282), received between Oct 6, 2022 and Aug 12, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K181240); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Device Reprocessing Problem2
- Physical Resistance/Sticking2
- Failure to Clean Adequately1
- Material Separation1
- Mechanical Problem1
- Microbial Contamination of Device1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 12, 2026Detachment of Device or Device Component
No new information has been provided by the customer.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50225560 ↗MalfunctionMar 23, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED MISSING ADHESIVE ON THE INSERTION TUBE SIDE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24660619 ↗MalfunctionMar 16, 2026Material Separation
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED A CHIP ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24603541 ↗MalfunctionMay 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED THE IMAGE GUIDE COVER GLASS WAS MISSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22078061 ↗MalfunctionJun 20, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED THE INSERTION SITE SOFT TUBE FOREIGN BODY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19572501 ↗MalfunctionMar 15, 2024Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED, THE FIBERSCOPE WAS BEING CLEANED WITH DETERGENT WIPES AND THE SCOPE WAS NOT BEING SOAKED IN DETERGENT AS INDICATED IN THE INSTRUCTIONS FOR USE MANUAL. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18914363 ↗MalfunctionMar 14, 2023Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO OLYMPUS, THE RHINO-LARYNGO FIBERSCOPE ANGULATION BECOMES LOCKED AND CANNOT DISENGAGE. THE ISSUE WAS FOUND DURING PREPARATION FOR USE. PER THE PHYSICIAN, THE DEVICES ARE INSPECTED PRIOR TO USE. IF DURING INSPECTION THE DEVICE DOES NOT FLEX PROPERLY, IT IS NOT USED ON THE PATIENT. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16538962 ↗MalfunctionFeb 20, 2023Physical Resistance/Sticking; Physical Resistance/Sticking
A PHYSICIAN REPORTED TO OLYMPUS, THE RHINO-LARYNGO FIBERSCOPE COULD NOT ANGULATE SUFFICIENTLY, AND THE CAMERA HEAD OR TIP WAS GETTING STUCK WHEN ARTICULATING. THE EVENT OCCURRED DURING A DIAGNOSTIC NASAL AND LARYNGEAL ENDOSCOPY. THE PROCEDURE WAS ONLY PROLONGED BY A FEW ADDITIONAL SECONDS. THE EVENT ALTHOUGH INTERMITTENT, IS OCCURRING WITH MORE FREQUENCY. AT TIMES, EITHER FLEXING OR EXTENDING THE SCOPE TIP USING THE THUMB TOGGLE, THE TIP WILL STICK IN A FULLY FLEXED OR EXTENDED POSITION. THE PHYSICIAN BELIEVES THE ISSUE IS RELATED TO THE DISTAL ASPECT OF THE SCOPE/TIP OR THE OPERATING THUMB TOGGLE. WHEN THE SCOPE IS PASSED THROUGH A PATIENT'S NOSE, FLEX THE TIP TO MOVE DOWN PAST THE NASOPHARYNX AND VISUALIZED THE GLOTTIS, THE MALFUNCTION OCCURS AT THIS PORTION OF THE EXAM. THE PHYSICIAN IS FORCED TO RETRACT THE FLEXIBLE SCOPE BACK THROUGH THE NASOPHARYNX AND NASAL PASSAGE WITH THE INABILITY TO STRAIGHT THE TIP WHICH RESULTED IN DISCOMFORT FOR THE PATIENT. THERE WAS NO PATIENT HARM OR INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16402843 ↗MalfunctionFeb 16, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THAT RHINO-LARYNGO FIBERSCOPE, PARTS FALL OFF FROM THE INSERTION TUBE, THE RUBBER IS FALLING DOWN FROM THE DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16390517 ↗MalfunctionJan 19, 2023Physical Resistance/Sticking; Physical Resistance/Sticking
AN OLYMPUS EMPLOYEE REPORTED ON BEHALF OF THE CUSTOMER, THE RHINO-LARYNGO FIBERSCOPE COULD NOT ANGULATE SUFFICIENTLY, AND THE CAMERA HEAD OR TIP WAS GETTING STUCK WHEN ARTICULATING. THE USER FACILITY STATED THE SUBJECT DEVICE WAS NOT IN USE DURING THE ACTIVE MALFUNCTION. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16211479 ↗MalfunctionDec 6, 2022Device Reprocessing Problem; Device Reprocessing Problem
AN ENDOSCOPY SUPPORT SPECIALIST (ESS) WAS PERFORMING AN IN-SERVICE AT THE CUSTOMER FACILITY AND DISCOVERED AN INAPPROPRIATE REPROCESSING ROUTINE RELATED WITH THE RHINO-LARYNGO FIBERSCOP. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15929434 ↗MalfunctionOct 6, 2022Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS, THE RHINO-LARYNGO FIBERSCOPE TESTED POSITIVE FOR 14 COLONY FORMING UNIT (CFU) OF BACILLUS SUBTILIS ON (B)(6) 2022, TESTED POSITIVE FOR 7 AND 3. COLONY FORMING UNIT (CFU) OF BACILLUS SP ON (B)(6) 2022, AND TESTED POSITIVE FOR 7 AND 3. COLONY FORMING UNIT (CFU) OF BACILLUS SP ON (B)(6) 2022. ALL CHANNELS WERE SAMPLED. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. SAMPLING WAS TAKEN AT REPROCESSING, BEFORE USE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15555794 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K181240. One submission often covers many models, so check the product and lot details in each record.
- Z-2292-2020 · May 5, 2020 · TerminatedENF-GP2 Rhino-Laryngo Fiberscope - Product Usage: is indicated for use within the nasal and nasopharyngeal lumen.
Other Nasopharyngoscope (flexible or rigid) devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)785 reports
- NA 242268Medos International Sàrl · Nasopharyngoscope (flexible or rigid)573 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Olympus Medical Systems Corp. · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- VIVIDEO VNL9-CPHoya Corporation · Nasopharyngoscope (Flexible Or Rigid)389 reports
- PENTAX FNL-10RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)322 reports
- SINGLE USE BIOPSY VALVE MAJ-210Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)1 recall221 reports
- SINGLE USE SUCTION VALVE MAJ-209Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)228 reports
- PENTAX FNL-7RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)227 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page