Single Use Loop Cutter FS-410 FS-410LRecalls & reported problems
Olympus Medical Systems Corp. · Single Use Loop Cutter FS-410 · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170401039), received between Apr 21, 2023 and May 2, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Close” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult to Open or Close1
- Failure to Cut1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 2, 2023Difficult to Open or Close; Difficult to Open or Close
AN OLYMPUS REPRESENTATIVE REPORTED TO OLYMPUS ON BEHALF OF THE CUSTOMER, THAT A PATIENT EXHIBITED BLEEDING AND HEMATOMA, DURING SUTURING AFTER AN ENDOSCOPIC SUBMUCOSAL DISSECTION (ESD). BULGING OF THE SURROUNDING MUCOSA WAS OBSERVED, SO TWO V-LOC SUTURES WERE USED. THE PHYSICIAN TRIED TO CUT THE EXCESS THREAD OF THE V-LOC WITH THE SINGLE USE LOOP CUTTER FS-410, BUT IT INITIALLY DID NOT OPEN. THE PHYSICIAN WAS EVENTUALLY ABLE TO OPEN IT. AND HE WAS ABLE TO CUT THE THREAD AND COLLECT THE LOOP CUTTER. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16857772 ↗MalfunctionApr 21, 2023Failure to Cut; Failure to Cut
THE CUSTOMER REPORTED TO OLYMPUS THAT, AFTER SUTURING IN A STOMACH CASE (EHS: ENDOSCOPIC HAND-SEWN SUTURING METHOD), WHEN TRYING TO CUT THE THREAD USING THE SINGLE USE LOOP CUTTER FS-410, THE THREAD CAUGHT, AND THE SLIDER STOPPED MOVING. THE CUSTOMER SWITCHED TO A HOOK KNIFE TO CUT THE THREAD AND THE ENDOSCOPIC SUBMUCOSAL DISSECTION (ESD) PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF PATIENT OR USER HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16791803 ↗
Other Endoscopic grasping/cutting instrument, non-powered, exempt devices
- Captura G56076Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt60 reports
- Captura G56133Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt45 reports
- Single Use Grasping Forceps FG-253SXOlympus Medical Systems Corp. · Endoscopic grasping/cutting instrument, non-powered, exempt27 reports
- Raptor 00711177US Endoscopy · Endoscopic grasping/cutting instrument, non-powered, exempt27 reports
- Captura G31576Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt19 reports
- Captura G56059Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt18 reports
- Captura Pro G50697Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt17 reports
- Captura G31577Cook Incorporated · Endoscopic grasping/cutting instrument, non-powered, exempt17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page