AIRWAY MOBILESCOPE OLYMPUS MAF-TM2Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of AIRWAY MOBILESCOPE (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Olympus MAF-TM2 Airway Mobilescope is a flexible bronchoscope used to look inside the airway. The 29 reports mostly describe the scope's outer coating peeling or flaking off, along with liquid leaking from the tip and problems cleaning or reprocessing it; 2 reports involve injuries and no deaths. Two open recalls, both from 2023, cover scopes returned after repair with a wet channel that could allow contamination, and a scope too wide that could get stuck in an endotracheal tube connector.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 29
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 04953170407147), received between Jan 6, 2022 and Dec 2, 2024 — classified as 2 injuries, 27 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (13 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0340-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 27, 2023Class IIOpen, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2499-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Peeled/Delaminated13
- Contamination3
- Device Reprocessing Problem2
- Failure to Clean Adequately2
- Leak/Splash2
- Microbial Contamination of Device2
- Break1
- Degraded1
Patient problems
- No Clinical Signs, Symptoms or Conditions27
- Drug Resistant Bacterial Infection2
Recent adverse event reports
The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2024Difficult to Remove; Difficult to Remove
NO NEW ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20821047 ↗MalfunctionSep 19, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THE MOBILESCOPE HAS LIQUID COMING OUT THROUGH THE DISTAL END. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20256350 ↗MalfunctionSep 4, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE AIRWAY MOBILE-SCOPE THE COAT OF THE IMAGE GUIDE SNAKE PIPE IS PEELED OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20140061 ↗MalfunctionJul 16, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE AIRWAY MOBILESCOPE EXHIBITED PEELING ON THE CONNECTING TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19752086 ↗MalfunctionJun 24, 2024No Display/Image; No Display/Image
IT WAS REPORTED THAT THE AIRWAY MOBILESCOPE HAD IMAGE ISSUES. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19601924 ↗MalfunctionJun 21, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE AIRWAY MOBILESCOPE ACTUATOR, SUCTION CYLINDER HAD A FOREIGN OBJECT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19589557 ↗MalfunctionJun 18, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING THE INSPECTION OF THE DEVICE, THE COATING ON THE IMAGE GUIDE SNAKE PIPE HAD PEELED OFF ON THE AIRWAY MOBILESCOPE. THERE WERE NO REPORTED INCIDENTS INVOLVING PATIENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19564159 ↗InjuryApr 15, 2024Contamination; Contamination
ADDITIONAL INFORMATION RECEIVED THAT THE PATIENT'S INFECTION WAS VERIFIED BY BLOOD CULTURES. THE REPROCESSOR WAS NOT CULTURED. ONLY THE SCOPES THAT WERE USED ON THE PATIENT WERE CULTURED, WHICH CAME BACK NEGATIVE FOR CARBAPENEM-RESISTANT ENTEROBACTERALES (CRE). AT THE TIME OF THIS REPORT, THE PATIENT WAS DISCHARGED FROM THE HOSPITAL AND THERE WERE NO OTHER PATIENTS THAT DEVELOPED CRE AFTER USING THE SCOPE. THE FACILITY WAS NOT SURE IF THE PATIENT CAME INTO THE HOSPITAL WITH CRE OR IS HE DEVELOPED IT DURING THE HOSPITAL STAY.
Patient problems: Drug Resistant Bacterial Infection
MAUDE report 19105718 ↗InjuryApr 15, 2024Contamination; Contamination
ADDITIONAL INFORMATION RECEIVED. THAT THE PATIENT'S INFECTION WAS VERIFIED, BY BLOOD CULTURES. THE REPROCESSOR WAS NOT CULTURED. ONLY THE SCOPES, THAT WERE USED ON THE PATIENT WERE CULTURED. WHICH CAME BACK NEGATIVE FOR CARBAPENEM-RESISTANT ENTEROBACTERIAL (CRE). AT THE TIME OF THIS REPORT, THE PATIENT WAS DISCHARGED FROM THE HOSPITAL AND THERE WERE NO OTHER PATIENTS, THAT DEVELOPED CRE AFTER USING THE SCOPE. THE FACILITY WAS NOT SURE IF THE PATIENT CAME INTO THE HOSPITAL WITH CRE OR IS HE DEVELOPED IT, DURING THE HOSPITAL STAY.
Patient problems: Drug Resistant Bacterial Infection
MAUDE report 19105622 ↗MalfunctionApr 15, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE AIRWAY MOBILESCOPE EXHIBITED PEELING ON THE GUIDE TUBE COATING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19104941 ↗MalfunctionApr 10, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE AIRWAY MOBILESCOPE EXHIBITED INSERTION SITE, FLEXIBLE TUBE, COATING PEELING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19080676 ↗MalfunctionMar 18, 2024Material Separation; Material Separation
IT WAS REPORTED THAT THE ADHESIVE OF THE BENDING SECTION COVER HAD SEPARATED, THE BIOPSY CHANNEL WAS BLOCKED AND THE INSERTION TUBE WAS DAMAGED ON THE AIRWAY MOBILESCOPE. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18920762 ↗MalfunctionFeb 27, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE AIRWAY MOBILESCOPE EXHIBITED THAT THE COAT OF THE IMAGE GUIDE INSERTION SECTION PIPE WAS PEELED OFF (1MM2 OR MORE). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18785948 ↗MalfunctionFeb 21, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE AIRWAY MOBILESCOPE EXHIBITED THE CONNECTING TUBE HAD COATING PEELING (1MILLIMETER TO 2 OR MORE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18753950 ↗MalfunctionFeb 7, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE COATING ON THE CONNECTING TUBE OF THE AIRWAY MOBILESCOPE WAS PEELED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18660759 ↗MalfunctionDec 1, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED THE AIRWAY MOBILESCOPE HAD A DAMAGED OPENING. THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION. EXAMINATION SHOWED THE INSERTION SECTION'S COATING WAS PEELING; THE PEELING AREA MEASURED 1 MM2 OR MORE. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18248514 ↗MalfunctionSep 20, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS THAT THE AIRWAY MOBILESCOPE HAD A CRUSHED INSERTION PART. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE DEVICE WAS RETURNED TO OLYMPUS FOR A DEVICE EVALUATION AND FOUND RESIDUAL LIQUID OR FOREIGN MATERIAL CAME OUT OF THE CHANNEL TUBE. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17786378 ↗MalfunctionSep 7, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS THAT DURING REPROCESSING, THE AIRWAY MOBILESCOPE HAD A POSITIVE CULTURE TEST. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17703348 ↗MalfunctionAug 30, 2023Contamination; Peeled/Delaminated; Contamination; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS THAT THE AIRWAY MOBILESCOPE POWER COULD NOT BE TURNED ON. THERE IS NO REPORT OF PATIENT OR USER HARM ASSOCIATED WITH THE EVENT. DURING INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS DISCOVERED THAT THE COATING ON THE CONNECTING TUBE WAS PEELING. THIS REPORT IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION WHICH WAS IDENTIFIED DURING THE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17654714 ↗MalfunctionAug 30, 2023Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE AIRWAY MOBILESCOPE WAS REPROCESSED WITH AN OER-6 ENDOSCOPE REPROCESSOR THAT COULD NOT DRAIN WATER SUFFICIENTLY WHEN REPLACING THE WATER FILTER. THE WATER FILTER WAS NOT CHANGED EVERY MONTH AND IT WAS OPERATED FOR ABOUT HALF A YEAR. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM. THIS REPORT IS RELATED TO THE FOLLOWING LINKED PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17649036 ↗MalfunctionAug 30, 2023Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE AIRWAY MOBILESCOPE WAS REPROCESSED WITH AN OER-6 ENDOSCOPE REPROCESSOR THAT COULD NOT DRAIN WATER SUFFICIENTLY WHEN REPLACING THE WATER FILTER. THE WATER FILTER WAS NOT CHANGED EVERY MONTH AND IT WAS OPERATED FOR ABOUT HALF A YEAR. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM. THIS REPORT IS RELATED TO THE FOLLOWING LINKED PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17649016 ↗MalfunctionApr 24, 2023Peeled/Delaminated; Peeled/Delaminated
THE OLYMPUS REPRESENTATIVE REPORTED (ON BEHALF OF THE CUSTOMER), THAT THE COATING ON THE AIRWAY MOBILESCOPE INSERTION SECTION WAS PEELING OFF. THE COATING DETACHMENT WAS NOTICED DURING THE ENDOTRACHEAL CHECK. THE INTENDED TRACHEAL INTUBATION FOR ANESTHESIA, DURING THE LARYNGOSCOPIC MALIGNANT LARYNGEAL TUMORECTOMY PROCEDURE, WAS COMPLETED BY CHECKING THE INSIDE OF THE TRACHEA WITH A FIBER. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT. DURING THE EVALUATION OF THE DEVICE, IT WAS NOTED, THAT THE IMAGE GUIDE SNAKE TUBE COATING WAS PEELED. THIS REPORT IS TO CAPTURE THE REPORTABLE MALFUNCTION OF THE IMAGE GUIDE COATING THAT WAS PEELED OFF AND NOTED AT ESTIMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16795454 ↗MalfunctionMar 13, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS, DURING REPROCESSING, THE AIRWAY MOBILESCOPE TESTED POSITIVE FOR RHIZOBIUM RADIOBACTER. A SECOND MICROBIOLOGICAL TEST WAS PERFORMED AND THE RESULTS ARE PENDING. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16534729 ↗MalfunctionAug 30, 2022Peeled/Delaminated; Peeled/Delaminated
THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THE COATING ON THE INSERTION TUBE WAS PEELING OFF. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION (COATING PEELING OFF). THERE WAS NO PATIENT OR USER INJURY REPORTED. ADDITIONAL DETAILS HAVE BEEN REQUESTED. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15320697 ↗MalfunctionJun 3, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THE INSERTION TUBE OF THE SUBJECT DEVICE WAS DAMAGED. THERE WAS NO PATIENT OR USER INJURY REPORTED DUE TO THE EVENT. THE DEVICE WAS RETURNED TO AN OLYMPUS SERVICE CENTER FOR EVALUATION. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THE BENDING SECTION COVER HAD FALLEN OFF FROM THE DISTAL END. THIS REPORT IS BEING SUBMITTED FOR THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION OF THE DEVICE (BENDING SECTION COVER HAD FALLEN OFF). ADDITIONAL DETAILS HAVE BEEN REQUESTED REGARDING THE REPORTED ISSUE. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14595863 ↗
Other Bronchoscope (flexible or rigid) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.