OLYMPUS OLYMPUS ENF-T3Recalls & reported problems
Olympus Medical Systems Corp. · RHINO-LARYNGOFIBERSCOPE · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04953170411076), received between Feb 16, 2023 and Apr 20, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Clean Adequately3
- Break1
- Detachment of Device or Device Component1
- Disconnection1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE FLEXIBLE SCOPE EXHIBITED A FOREIGN BODY ON THE ADHESIVE OF THE BENDING SECTION COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19149921 ↗MalfunctionMar 26, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED FOREIGN OBJECT ON THE ACTUATOR BODY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18980011 ↗MalfunctionJan 16, 2024Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS, THE RHINO-LARYNGOFIBERSCOPE HAD A CLOUDY IMAGE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE OBJECTIVE LENS AND LIGHT GUIDE LENS HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18522213 ↗MalfunctionDec 4, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THE RHINO-LARYNGOFIBERSCOPE HAD COATING PEELING OFF AT THE TIP OF THE SCOPE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE RESIN PART OF THE IMAGE GUIDE FLEXIBLE PIPE HAS FALLEN OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18263740 ↗MalfunctionOct 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
AN OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE RHINO-LARYNGOFIBERSCOPE HAD PEELING OF THE OUTER SURFACE OF THE INSERTION PART. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE RESIN PART OF THE IMAGE GUIDE FLEXIBLE PIPE HAS FALLEN OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17862050 ↗MalfunctionFeb 16, 2023Disconnection; Disconnection
THE CUSTOMER REPORTED TO OLYMPUS, THE RHINO-LARYNGOFIBERSCOPE HAD DEFECTS ON THE INSERTION TUBE-NON BENDING SECTION. THE INTENDED PROCEDURE WAS A BIOPSY OF A LARYNGEAL TUMOR. THE PROCEDURE WAS COMPLETED. THE CUSTOMER USES THE SUBJECT DEVICE ONE TIME PER MONTH. THE CUSTOMER USES PHTHALAL FOR CLEANING, DISINFECTION, AND STERILIZATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16386570 ↗
Other Nasopharyngoscope (flexible or rigid) devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)785 reports
- NA 242268Medos International Sàrl · Nasopharyngoscope (flexible or rigid)573 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Olympus Medical Systems Corp. · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- VIVIDEO VNL9-CPHoya Corporation · Nasopharyngoscope (Flexible Or Rigid)389 reports
- PENTAX FNL-10RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)322 reports
- SINGLE USE BIOPSY VALVE MAJ-210Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)1 recall221 reports
- SINGLE USE SUCTION VALVE MAJ-209Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)228 reports
- PENTAX FNL-7RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)227 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page