OLYMPUS OLYMPUS ENF-T3Recalls & reported problems

Olympus Medical Systems Corp. · RHINO-LARYNGOFIBERSCOPE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04953170411076), received between Feb 16, 2023 and Apr 20, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately3
  • Break1
  • Detachment of Device or Device Component1
  • Disconnection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 20, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE FLEXIBLE SCOPE EXHIBITED A FOREIGN BODY ON THE ADHESIVE OF THE BENDING SECTION COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19149921 ↗
  • MalfunctionMar 26, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED FOREIGN OBJECT ON THE ACTUATOR BODY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18980011 ↗
  • MalfunctionJan 16, 2024Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED TO OLYMPUS, THE RHINO-LARYNGOFIBERSCOPE HAD A CLOUDY IMAGE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE OBJECTIVE LENS AND LIGHT GUIDE LENS HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18522213 ↗
  • MalfunctionDec 4, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, THE RHINO-LARYNGOFIBERSCOPE HAD COATING PEELING OFF AT THE TIP OF THE SCOPE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE RESIN PART OF THE IMAGE GUIDE FLEXIBLE PIPE HAS FALLEN OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18263740 ↗
  • MalfunctionOct 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    AN OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE RHINO-LARYNGOFIBERSCOPE HAD PEELING OF THE OUTER SURFACE OF THE INSERTION PART. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE RESIN PART OF THE IMAGE GUIDE FLEXIBLE PIPE HAS FALLEN OFF. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17862050 ↗
  • MalfunctionFeb 16, 2023Disconnection; Disconnection

    THE CUSTOMER REPORTED TO OLYMPUS, THE RHINO-LARYNGOFIBERSCOPE HAD DEFECTS ON THE INSERTION TUBE-NON BENDING SECTION. THE INTENDED PROCEDURE WAS A BIOPSY OF A LARYNGEAL TUMOR. THE PROCEDURE WAS COMPLETED. THE CUSTOMER USES THE SUBJECT DEVICE ONE TIME PER MONTH. THE CUSTOMER USES PHTHALAL FOR CLEANING, DISINFECTION, AND STERILIZATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16386570 ↗

Other Nasopharyngoscope (flexible or rigid) devices

All Nasopharyngoscope (flexible or rigid) devices →