RHINO-LARYNGO FIBERSCOPE ENF-GP2 (UDI 04953170411328)Recalls & reported problems

Olympus Medical Systems Corp. · RHINO-LARYNGO FIBERSCOPE · Class 2

All models of RHINO-LARYNGO FIBERSCOPE (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus Rhino-Laryngo Fiberscope ENF-GP2 is a flexible nasopharyngoscope used to look at the nose and throat. The 31 reports mostly describe the scope not cleaning properly, parts detaching, plastic or fiber pieces chipping or breaking off, a chipped objective lens, corrosion, and poor image quality; none involved a patient. There are no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 04953170411328), received between Sep 6, 2022 and Sep 17, 2026 — classified as 33 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (5 reports).

1 other recall cover devices cleared under the same FDA submission (K181240); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 10 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 5 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately5
  • Detachment of Device or Device Component4
  • Material Separation4
  • Corroded3
  • Device Reprocessing Problem2
  • Material Fragmentation2
  • Mechanical Problem2
  • Poor Quality Image2

Patient problems

  • No Clinical Signs, Symptoms or Conditions32
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Material Separation

    The customer reported the Rhino-Laryngofiberscope had a separate issue. During the initial evaluation of the device it was observed, that light guide lenses and objective lens (charged coupled device) are missing adhesive. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50803826 ↗
  • MalfunctionSep 9, 2026Corroded; Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50698455 ↗
  • MalfunctionAug 28, 2026Contamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50506749 ↗
  • MalfunctionAug 28, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50502256 ↗
  • MalfunctionAug 26, 2026Environmental Particulates

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456633 ↗
  • MalfunctionAug 17, 2026Material Fragmentation

    The customer returned the Rhino-Laryngofiberscope for a separate issue. During the initial evaluation of the returned device, a chip on the objective lens was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50272568 ↗
  • MalfunctionAug 5, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50143489 ↗
  • MalfunctionMay 6, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE RHINO-LARYNGO FIBERSCOPE EXHIBITED THE UNIT PLASTIC DISTAL END COVER CHIPPED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096368 ↗
  • MalfunctionApr 15, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO LARYNGO FIBERSCOPE EXHIBITED MISSING OBJECTIVE LENS SEALING GLUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24884718 ↗
  • MalfunctionFeb 9, 2026Material Integrity Problem

    IT WAS REPORTED THE RHINO-LARYNGO FIBERSCOPE WAS DROPPED AND FOUND TO HAVE FIBERGLASS INSIDE OF THE DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24293880 ↗
  • MalfunctionJan 21, 2026Component Missing

    THE CUSTOMER REPORTED THAT THE END OF THEIR RHINO-LARYNGO FIBERSCOPE DEVICE HAD A PIECE MISSING; THE FIBEROPTICS COULD BE OBSERVED COMING THROUGH WHERE THE SHEATH SHOULD BE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24136088 ↗
  • MalfunctionDec 4, 2025Device Reprocessing Problem; Fluid/Blood Leak; Device Reprocessing Problem; Fluid/Blood Leak

    IT WAS REPORTED THE RHINO-LARYNGO FIBERSCOPE HAD GRAY DRAINAGE FROM THE SCOPE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23718178 ↗
  • MalfunctionNov 5, 2025Peeled/Delaminated; Peeled/Delaminated

    DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE DISTAL END COVER WAS MISSING GLUE ON THE FIBERSCOPE. NO PATIENTS WERE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23477168 ↗
  • MalfunctionNov 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SUBJECT DEVICE EXHIBITED MISSING GLUE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23460463 ↗
  • MalfunctionOct 31, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE SHEATH ON THE TIP OF THE FIBERSCOPE. WHICH IS INSERTED INTO THE PATIENT¿S NOSE, WAS LOOSE. THE LOCATION WHERE THE EVENT OCCURRED WAS NOT SPECIFIED BY THE REPORTER. NO PATIENT HARM WAS REPORTED IN CONNECTION WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23439595 ↗
  • MalfunctionOct 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO FIBERSCOPE EXHIBITED THAT THERE WAS A MISSING PART OF THE INSERTION TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23414798 ↗
  • MalfunctionOct 29, 2025Mechanical Problem; Mechanical Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO FIBERSCOPE EXHIBITED THE ANGULATION WAS LOCKED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23413769 ↗
  • MalfunctionOct 29, 2025Material Separation; Material Separation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED THE LIGHT GUIDE LENS SEPARATED BETWEEN DISTAL END METAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23410645 ↗
  • MalfunctionOct 28, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGO FIBERSCOPE EXHIBITED A DIRTY OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23408267 ↗
  • MalfunctionOct 28, 2025Corroded; Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED CORROSION AROUND THE LENS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23400870 ↗
  • MalfunctionOct 27, 2025Corroded; Corroded

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23398098 ↗
  • MalfunctionOct 23, 2025Material Separation; Material Separation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RHINO-LARYNGOFIBERSCOPE EXHIBITED A MISSING COVER GLASS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368657 ↗
  • MalfunctionOct 22, 2025Poor Quality Image; Poor Quality Image

    DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE IMAGE WAS CLOUDY ON THE FIBERSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23358307 ↗
  • MalfunctionOct 16, 2025Degraded; Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED MISSING ADHESIVE FROM THE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23315754 ↗
  • MalfunctionSep 18, 2025Material Separation; Material Separation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SCOPE EXHIBITED LOOSE LIGHT GUIDE COVER GLASS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 23082265 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K181240. One submission often covers many models, so check the product and lot details in each record.

  • Z-2292-2020 · May 5, 2020 · Terminated
    ENF-GP2 Rhino-Laryngo Fiberscope - Product Usage: is indicated for use within the nasal and nasopharyngeal lumen.

Other Nasopharyngoscope (flexible or rigid) devices

All Nasopharyngoscope (flexible or rigid) devices →