12G-SDI cable 8.5M MAJ-2429Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus SDI cable is an 8.5-meter video cable used with a bronchoscope imaging stack to send the picture to a monitor. The 30 reports mostly describe the monitor showing no image at all, or a flickering, noisy, poor-quality picture, plus some connection problems where the cable worked loose. All30 describe malfunctions with no patient involved, and no deaths or injuries were reported. There are no recalls naming this cable.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 04953170415630), received between Sep 29, 2023 and Sep 30, 2026 — classified as 32 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (20 reports).

1 other recall cover devices cleared under the same FDA submission (K222584, K222861); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 4 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 6 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image20
  • Poor Quality Image11
  • Erratic or Intermittent Display5
  • Disconnection2
  • Signal Artifact/Noise2
  • Break1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions32

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026No Display/Image

    The customer reported that the image output stopped during therapeutic procedure while the patient was under sedation and the device was inside the patient, involving an Olympus serial digital Interface cable. The procedure was completed using an olympus backup device. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944628 ↗
  • MalfunctionSep 8, 2026Erratic or Intermittent Display

    The customer reported an image blinking during inspection for use involving an Olympus Serial Digital Interface (SDI) cable. There was no patient harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50667024 ↗
  • MalfunctionMay 28, 2026No Display/Image

    THE CUSTOMER REPORTED NO IMAGE ON MAIN MONITOR DURING INSPECTION FOR USE INVOLVING AN OLYMPUS SERIAL DIGITAL INTERFACE CABLE. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25319164 ↗
  • MalfunctionMar 20, 2026Disconnection

    NO ADDITIONAL INFORMATION FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24650517 ↗
  • MalfunctionMar 4, 2026No Display/Image; Poor Quality Image

    IT WAS REPORTED THAT WHEN THE SDI CABLE WAS CONNECTED TO THE STACK, THE IMAGE INITIALLY BEGAN FLICKERING WITH GREEN LINES AND THEN WENT BLACK. THE ISSUE OCCURRED DURING PREPARATION FOR USE, AND THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24510235 ↗
  • MalfunctionMar 3, 2026No Display/Image; Poor Quality Image

    IT WAS REPORTED THAT THE MONITOR INITIALLY STARTED FLICKERING WITH GREEN LINES AND THEN WENT BLANK WHEN USED WITH THE SDI CABLE. THIS ISSUE WAS FOUND DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24502984 ↗
  • MalfunctionFeb 20, 2026Erratic or Intermittent Display

    IT WAS REPORTED THE SERIAL DIGITAL INTERFACE CABLE PRESENTED A MONITOR WITH A BLINKING IMAGE. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24405248 ↗
  • MalfunctionFeb 18, 2026No Display/Image

    IT WAS REPORTED THE SERIAL DIGITAL INTERFACE CABLE WAS UNABLE TO GET CONSISTENT IMAGE. THIS ISSUE OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24378864 ↗
  • MalfunctionFeb 6, 2026Disconnection; Poor Quality Image

    IT WAS REPORTED THE SERIAL DIGITAL INTERFACE HAD IMAGE NOISE AND DISCONNECTION UNDER THE CONNECTOR DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24275578 ↗
  • MalfunctionFeb 5, 2026Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24272163 ↗
  • MalfunctionFeb 5, 2026Erratic or Intermittent Display; Poor Quality Image

    IT WAS REPORTED THE SERIAL DIGITAL INTERFACE CABLE CAUSED IMAGE NOISE. THE ISSUE WAS FOUND DURING A THERAPEUTIC VIDEO-ASSISTED THORACOSCOPIC SURGERY, AND IT WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24271899 ↗
  • MalfunctionFeb 5, 2026Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24268348 ↗
  • MalfunctionNov 12, 2025No Display/Image; Poor Quality Image; No Display/Image; Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CABLE CONNECTED TO THE MONITOR EXHIBITED NOISE AND MOMENTARY BLACKOUTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23537935 ↗
  • MalfunctionSep 19, 2025Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23095095 ↗
  • MalfunctionSep 3, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE WAS UNABLE TO SEND AN SDI (SERIAL DIGITAL INTERFACE) SIGNAL DURING SET UP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22952051 ↗
  • MalfunctionAug 27, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE EXPERIENCED MOMENTARY BLACKOUTS DURING A UNSPECIFIED THERAPEUTIC PROCEDURE. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22907530 ↗
  • MalfunctionAug 20, 2025No Display/Image; Loose or Intermittent Connection; No Display/Image; Loose or Intermittent Connection

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22848664 ↗
  • MalfunctionMay 2, 2025No Display/Image; Poor Quality Image; No Display/Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD GREEN LINES AND INTERMITTENT BLACK SCREEN ON DISPLAY. THE ISSUE OCCURRED DURING SERVICE / MAINTENANCE (NOT IN A CLINICAL SETTING). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21954113 ↗
  • MalfunctionMar 28, 2025No Display/Image; Poor Quality Image; No Display/Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD NOISE AND NO VIDEO IMAGE. THE ISSUE WAS FOUND DURING SEDATION OF THERAPEUTIC PROCEDURE. IT WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21719118 ↗
  • MalfunctionMar 12, 2025Signal Artifact/Noise; Signal Artifact/Noise

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21585584 ↗
  • MalfunctionJan 16, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE'S CABLE WAS DAMAGED, RESULTING IN GREEN NOISE IN THE IMAGES. THE ISSUE OCCURRED DURING THE PRE-OPERATION INSPECTION, AND THE INTENDED PROCEDURE WAS COMPLETED USING THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21167630 ↗
  • MalfunctionJan 16, 2025Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE EXPERIENCED A MALFUNCTION DURING SET UP / INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21167569 ↗
  • MalfunctionDec 2, 2024Poor Quality Image; Poor Quality Image

    A FIELD SERVICE ENGINEER REPORTED ON BEHALF OF THE CUSTOMER, THAT THE SERIAL DIGITAL INTERFACE (SDI) CABLE WAS PRODUCING IMAGE NOISE. THE ISSUE WAS FOUND DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20818927 ↗
  • MalfunctionOct 24, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SERIAL DIGITAL INTERFACE CABLE EXHIBITED AN IMAGE BLACKOUT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20524384 ↗
  • MalfunctionAug 27, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED, THE SERIAL DIGITAL INTERFACE CABLE WAS USED IN A CASE WHERE THERE WAS INTERMITTENT IMAGE. THE ISSUE OCCURRED DURING A THERAPEUTIC LIVER RESECTION PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM AND THE PROCEDURE WAS ABLE TO BE COMPLETED USING THE SAME SET OF EQUIPMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20075274 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K222584, K222861. One submission often covers many models, so check the product and lot details in each record.

  • Z-0181-2025 · Aug 21, 2024 · Open, Classified
    Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100

Other Bronchoscope (Flexible Or Rigid) devices

All Bronchoscope (Flexible Or Rigid) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page