Skip to content

BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of BRONCHOVIDEOSCOPE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus BF-1TH1100 bronchoscope is a flexible scope used to look inside the lungs and airways. The 54 reports mostly describe the scope's tip cover burning or melting when used with laser, argon plasma coagulation or high-frequency therapy equipment, along with loss of the video image and connection errors between the scope and its processor. No deaths and no injuries are described. One open recall, started in September 2025, adds instructions clarifying correct use of bronchoscopes alongside laser, argon plasma coagulation and high-frequency therapy equipment.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 04953170424199), received between Jul 10, 2023 and Dec 6, 2024 — classified as 54 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (18 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K250862); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 3 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 5 reportsJun 2024: 6 reportsJul 2024: 3 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 11 reportsNov 2024: 5 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 11, 2025Class IOpen, Classified

    Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.

    Z-0048-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Display/Image18
  • Communication or Transmission Problem14
  • Thermal Decomposition of Device6
  • Microbial Contamination of Device4
  • Detachment of Device or Device Component2
  • Melted2
  • Output Problem2
  • Peeled/Delaminated2

Patient problems

  • No Clinical Signs, Symptoms or Conditions54

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 6, 2024Break; Device Damaged by Another Device; Break; Device Damaged by Another Device

    IT WAS REPORTED THE DISTAL END OF THE SUBJECT DEVICE BURNT AFTER USING ARGON PROBE. "ARGON PROBE IS SAID TO HAVE HAD A LATERAL CRACK". PER THE REPORT, THE ISSUE WAS FOUND "AFTER PROCEDURE (UPON COMPLETION OF PROCEDURE, DEVICE NO LONGER USED ON PATIENT). THERE WAS NO PATIENT HARM, NO INJURY REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20870104 ↗
  • MalfunctionNov 12, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOVIDEOSCOPE¿S PLASTIC DISTAL END COVER WAS BURNED OR MELTED AROUND THE INSTRUMENT CHANNEL OUTLET AT THE DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20667253 ↗
  • MalfunctionNov 7, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOVIDEOSCOPE¿S PLASTIC DISTAL END COVER WAS BURNED OR MELTED AROUND THE INSTRUMENT CHANNEL OUTLET AT THE DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20628798 ↗
  • MalfunctionNov 6, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOVIDEOSCOPE¿S PLASTIC DISTAL END COVER WAS BURNED OR MELTED AROUND THE INSTRUMENT CHANNEL OUTLET AT THE DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20621582 ↗
  • MalfunctionNov 4, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE'S PLASTIC DISTAL END COVER WAS BURNED/MELTED AROUND THE INSTRUMENT CHANNEL OUTLET AT THE DISTAL END. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20601828 ↗
  • MalfunctionNov 4, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE'S PLASTIC DISTAL END COVER HAD BURNED OR MELTED AROUND THE INSTRUMENT CHANNEL OUTLET AT THE DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20600777 ↗
  • MalfunctionOct 31, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THERE WAS NO IMAGE ON THE BRONCHOVIDEOSCOPE. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20577425 ↗
  • MalfunctionOct 30, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE DISPLAYED ERROR E216 (SCOPE COMMUNICATION ERROR). THE ISSUE OCCURRED DURING AN UNKNOWN EVENT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20569492 ↗
  • MalfunctionOct 23, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, THE BRONCHOVIDEOSCOPE DID NOT TRANSFER THE IMAGE TO THE MONITOR. AND THERE WAS RANDOM IMAGE LOSS. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20513711 ↗
  • MalfunctionOct 15, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE BRONCHOVIDEOSCOPE HAD NO IMAGE DURING ANGULATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20450959 ↗
  • MalfunctionOct 9, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20411316 ↗
  • MalfunctionOct 9, 2024Melted; Melted

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BRONCHOVIDEOSCOPE EXHIBITED THE PROBE COVER WITH A BURN. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20410248 ↗
  • MalfunctionOct 9, 2024No Display/Image; Output Problem; No Display/Image; Output Problem

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOVIDEOSCOPE EXHIBITED NO IMAGE AND AN E217 OUTPUT ERROR CODE WAS DISPLAYED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20408972 ↗
  • MalfunctionOct 8, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE IMAGE DISAPPEARED DURING THE UP ANGULATION ON THE BRONCHOVIDEOSCOPE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20404486 ↗
  • MalfunctionOct 4, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE ENDOSCOPE CONSTANTLY REPORTS A COMMUNICATION ERROR (ERROR CODE B30). THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20371818 ↗
  • MalfunctionOct 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE FRONT RING AT THE DISTAL END OF THE BRONCHOVIDEOSCOPE CAME LOOSE DURING AN UNSPECIFIED PROCEDURE. THE RING REMAINED ATTACHED TO THE DISTAL END UNTIL THE END OF THE PROCEDURE AND THEN FELL OFF DURING THE FIRST WIPE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20351663 ↗
  • MalfunctionOct 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE BRONCHOVIDEOSCOPE'S BIOPSY CHANNEL BECAME LOOSE DURING AN UNSPECIFIED PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20350781 ↗
  • MalfunctionSep 6, 2024Output Problem; Output Problem

    IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE HAD A COMMUNICATION ERROR WITH THE PROCESSOR. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20162691 ↗
  • MalfunctionAug 22, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THE BRONCHOVIDEOSCOPE HAD NO IMAGE. THE ISSUE OCCURRED DURING PREPARATION FOR USE IN AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20024700 ↗
  • MalfunctionAug 8, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE HAD NO IMAGE. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19940756 ↗
  • MalfunctionJul 18, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THE BRONCHOVIDEOSCOPE HAD AN E216 SCOPE COMMUNICATION ERROR DURING A BRONCHOSCOPY PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE WITH A FEW MINUTES OF DELAY. THE PATIENT WAS IN SUPPRESSION BEFORE THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19775041 ↗
  • MalfunctionJul 11, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, THE BRONCHOVIDEOSCOPE IMAGE DOES NOT WORK WITH STRONG ANGULATION. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19719815 ↗
  • MalfunctionJul 2, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED, THE BRONCHOVIDEOSCOPE HAD AN ERROR MESSAGE COME UP STATING CONTACT OLYMPUS "SCOPE COMMUNICATION ERROR" WHEN PLUGGING SCOPE INTO PROCESSOR. THE ISSUE OCCURRED DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19656656 ↗
  • MalfunctionJun 28, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THE BRONCHOVIDEOSCOPE HAD SPORADIC IMAGE FAILURES. THE ISSUE OCCURRED DURING THE PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19632739 ↗
  • MalfunctionJun 21, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THE BRONCHOVIDEOSCOPE HAD THE OCCASIONAL LOSS OF THE IMAGE. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR A BRONCHOSCOPY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19583314 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K250862. One submission often covers many models, so check the product and lot details in each record.

  • Z-0066-2026 · Sep 11, 2025 · Open, Classified
    EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS BF-XT190
  • Z-0057-2026 · Sep 11, 2025 · Open, Classified
    EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190
  • Z-0052-2026 · Sep 11, 2025 · Open, Classified
    BRONCHOVIDEOSCOPE OLYMPUS BF-H1100

Other Bronchoscope (Flexible Or Rigid) devices

All Bronchoscope (Flexible Or Rigid) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page