Skip to content

BALLOON MAJ-1351 (REF N5369631)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of BALLOON (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
25
Deaths reported
0
Injuries reported
13
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 04953170463938), received between Apr 25, 2025 and Sep 16, 2026 — classified as 13 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 8 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 4 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component17
  • Incomplete or Inadequate Connection3
  • Material Puncture/Hole3
  • Break1
  • Device Slipped1
  • Loss of or Failure to Bond1
  • Material Rupture1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2026Detachment of Device or Device Component; Material Split, Cut or Torn

    It was reported that during a diagnostic EBUS (endobronchial ultrasound), the balloon was partially torn off when the scope was being withdrawn through the endotracheal tube. The torn piece could not be located or recovered.

    Patient problems: Foreign Body In Patient

    MAUDE report 50793201 ↗
  • MalfunctionSep 16, 2026Detachment of Device or Device Component

    It was reported by the customer that the balloon detached during the withdrawal of the scope. The detached balloon subsequently recovered after being coughed out by the patient. The issue occurred during an Endobronchial Ultrasound (EBUS) procedure. No harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50790956 ↗
  • MalfunctionAug 31, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50521706 ↗
  • InjuryJul 28, 2026Detachment of Device or Device Component

    No additional information was received from the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50059582 ↗
  • InjuryJul 28, 2026Detachment of Device or Device Component

    It was reported that during an unknown procedure, the balloon that was used with the bronchofibervideoscope fell inside the patient. The balloon was successfully retrieved. The procedure was then completed with a similar device. There were no further reports of patient harm.This report is 1 of 2 devices, for the balloon.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50058672 ↗
  • MalfunctionApr 9, 2026Detachment of Device or Device Component

    DURING AN ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE, THE MAJ-1351 LATEX BALLOON DETACHED FROM THE TIP OF THE SCOPE WHEN PULLING BACK THE ENDOSCOPE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24832609 ↗
  • InjuryMar 10, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BALLOON FELL OFF FROM THE ULTRASONIC BRONCHOFIBERVIDEOSCOPE AND WAS LEFT BEHIND IN THE PATIENT'S AIRWAY DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND. THE PROCEDURE WAS STOPPED AND THE PROVIDER SWITCHED TO A REGULAR BRONCHOSCOPE TO RETRIEVE THE BALLOON FROM THE LEFT MAIN STEM BRONCHUS USING SUCTION. THE PROCEDURE WAS COMPLETED WITH A BACK-UP DEVICE WITH A DELAY OF LESS THAN 30 MINUTES. THERE WERE NO REPORTS OF FURTHER PATIENT HARM. THIS REPORT IS 1 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557196 ↗
  • MalfunctionMar 3, 2026Break

    NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24497699 ↗
  • InjuryFeb 20, 2026Detachment of Device or Device Component

    REPORT 1 OF 2 IT WAS REPORTED THAT DURING AN EBUS (ENDOBRONCHIAL ULTRASOUND) THE BALLOON FELL OFF THE ULTRASONIC BRONCHOFIBERVIDEOSCOPE INTO THE PATIENT¿S LUNG. THE PHYSICIAN NOTICED IT WAS NOT ON THE SCOPE WHEN HE EXITED THE ET TUBE (ENDOTRACHEAL TUBE). THE BALLOON WAS SUCTIONED TO THE TIP OF THE SCOPE, RETRIEVED THROUGH THE ET TUBE AND THE PROCEDURE WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24402387 ↗
  • MalfunctionFeb 10, 2026Incomplete or Inadequate Connection

    NO NEW ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24307833 ↗
  • MalfunctionFeb 2, 2026Device Slipped; Incomplete or Inadequate Connection

    IT WAS REPORTED THE BALLOON CAME LOOSE AND REMAINED HANGING AT THE END OF THE SCOPE DURING A DIAGNOSTIC, ENDOBRONCHIAL ULTRASOUND PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT BALLOON. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24236063 ↗
  • MalfunctionFeb 2, 2026Incomplete or Inadequate Connection

    IT WAS REPORTED THAT DURING A DIAGNOSTIC EBUS PROCEDURE, THE BALLOON ON THE EBUS SCOPE BECAME LOOSE AND "REMAINED HANGING AT THE END OF THE SCOPE." THERE WERE NO INJURIES OR HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24231544 ↗
  • MalfunctionDec 30, 2025Material Rupture; Material Rupture

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23921951 ↗
  • MalfunctionDec 4, 2025Loss of or Failure to Bond; Detachment of Device or Device Component; Loss of or Failure to Bond; Detachment of Device or Device Component

    THE CUSTOMER CONFIRMED THERE WAS NO PATIENT INFECTION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23716588 ↗
  • InjuryJun 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE, UNDER GENERAL ANESTHESIA, THE MAJ-1351 BALLOON FELL OFF INTO THE PATIENT'S AIRWAY. THE BALLOON WAS RETRIEVED WITH DISPOSABLE BIOPSY FORCEPS (MODEL# FB ¿ 211D OLYMPUS) RIGHT AFTER THE SCOPE WAS REMOVED AND THE MISSING PIECE WAS DISCOVERED. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITHOUT A DELAY AND NO UNPLANNED DIAGNOSTIC TEST WAS REQUIRED. NO REPORTS OF PATIENT HARM. REPORTEDLY, THERE WAS NO PROBLEM WITH PATIENT'S EXTUBATING PROCESS OR RECOVERY. THIS EVENT IS FOR THE BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332424 ↗
  • InjuryJun 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE, UNDER GENERAL ANESTHESIA, THE MAJ-1351 BALLOON FELL OFF INTO THE PATIENT'S AIRWAY. THE BALLOON WAS RETRIEVED WITH DISPOSABLE BIOPSY FORCEPS (MODEL# FB ¿ 211D OLYMPUS) RIGHT AFTER THE SCOPE WAS REMOVED AND THE MISSING PIECE WAS DISCOVERED. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITHOUT A DELAY AND NO UNPLANNED DIAGNOSTIC TEST WAS REQUIRED. NO REPORTS OF PATIENT HARM. REPORTEDLY, THERE WAS NO PROBLEM WITH PATIENT'S EXTUBATING PROCESS OR RECOVERY. THIS EVENT IS FOR THE BALLOON. THE COMPLAINT IS RELATED TO THE PATIENT IDENTIFIER (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332411 ↗
  • InjuryJun 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION WAS RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22206813 ↗
  • InjuryJun 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BALLOON FELL OFF THE SCOPE AND INTO THE PATIENT'S LUNG DURING AN ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE. THE BALLOON WAS RETRIEVED USING FORCEPS. THE PATIENT REPORTEDLY RECOVERED WITHOUT ISSUE. THE FOLLOWING PATIENT IDENTIFIER IS LINKED: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22206673 ↗
  • InjuryJun 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22195210 ↗
  • InjuryJun 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE, A SECOND BALLOON FELL INTO THE PATIENT'S LUNGS THAT WAS RETRIEVED BY SUCTIONING THE BALLOON AND PULLED IT OUT OF THE PATIENT WITHOUT ANY COMPLICATION. AFTER THE PROCEDURE, THE PATIENT WAS STABLE AND WENT HOME. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22195167 ↗
  • InjuryJun 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE, THE BALLOON FELL INTO THE PATIENT'S LUNGS THAT WAS RETRIEVED BY SUCTIONING THE BALLOON AND PULLED IT OUT OF THE PATIENT WITHOUT ANY COMPLICATION. AFTER THE PROCEDURE, THE PATIENT WAS STABLE AND WENT HOME. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22188077 ↗
  • InjuryJun 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE, THE BALLOON FELL INTO THE PATIENT'S LUNGS THAT WAS RETRIEVED BY SUCTIONING THE BALLOON AND PULLED IT OUT OF THE PATIENT WITHOUT ANY COMPLICATION. AFTER THE PROCEDURE, THE PATIENT WAS STABLE AND WENT HOME. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22188066 ↗
  • MalfunctionMay 29, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT WHEN THE CUSTOMER OPENED THE PACKAGE, THE BALLOONS HAD HOLES IN THEM. THIS OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22104045 ↗
  • MalfunctionMay 29, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT WHEN THE CUSTOMER OPENED THE PACKAGE, THE BALLOONS HAD HOLES IN THEM. THIS OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22103953 ↗
  • MalfunctionApr 25, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THE BALLOON HAD HOLE IN IT. THE ISSUE OCCURRED DURING UNPACKING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21913091 ↗

Other Bronchoscope (Flexible Or Rigid) devices

All Bronchoscope (Flexible Or Rigid) devices →

Sources and updates