PENTAX FNL-10RBSRecalls & reported problems

Hoya Corporation · Perform cleaning, high-level disinfection, and/or sterilization according to the procedure specified in the Instructions for Use (Reprocessing) of this instrument. · Class 2

All models of PENTAX (72)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Pentax FNL-10RBS nasopharyngoscope is a flexible endoscope used to view the nose and throat. Of its 171 reports, most describe image failures from broken fibers or fluid getting inside the scope. The reports mention no deaths and no injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
171
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 171 adverse event reports that list this exact device (DI 04961333224036), received between Jun 3, 2022 and Jul 22, 2024 — classified as 171 malfunctions.

The most frequently reported specific device problem is “Break” (88 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 7 reportsJan 2024: 6 reportsFeb 2024: 4 reportsMar 2024: 6 reportsApr 2024: 3 reportsMay 2024: 7 reportsJun 2024: 3 reportsJul 2024: 7 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break88
  • Fluid/Blood Leak67
  • Optical Obstruction7
  • Optical Distortion4
  • Detachment of Device or Device Component2
  • Component Missing1
  • Degraded1
  • Misfocusing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions171

Recent adverse event reports

The 25 most recent of 171 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2024Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19789952 ↗
  • MalfunctionJul 17, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19758148 ↗
  • MalfunctionJul 17, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19758117 ↗
  • MalfunctionJul 12, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19727588 ↗
  • MalfunctionJul 9, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19698745 ↗
  • MalfunctionJul 8, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19686990 ↗
  • MalfunctionJul 2, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19653109 ↗
  • MalfunctionJun 25, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19602985 ↗
  • MalfunctionJun 13, 2024Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19526155 ↗
  • MalfunctionJun 13, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19526119 ↗
  • MalfunctionMay 27, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19399360 ↗
  • MalfunctionMay 24, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19389288 ↗
  • MalfunctionMay 13, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19296058 ↗
  • MalfunctionMay 13, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19296053 ↗
  • MalfunctionMay 9, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19275862 ↗
  • MalfunctionMay 9, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19275090 ↗
  • MalfunctionMay 9, 2024Optical Distortion; Optical Distortion

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19275084 ↗
  • MalfunctionApr 26, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19188968 ↗
  • MalfunctionApr 25, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19178052 ↗
  • MalfunctionApr 24, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19170050 ↗
  • MalfunctionMar 25, 2024Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18966694 ↗
  • MalfunctionMar 25, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18966672 ↗
  • MalfunctionMar 19, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18927730 ↗
  • MalfunctionMar 14, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18900334 ↗
  • MalfunctionMar 11, 2024Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18872748 ↗

Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices

All NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices →