Skip to content

PENTAX FB-15VRecalls & reported problems

Hoya Corporation · Perform cleaning, high-level disinfection, and/or sterilization according to the procedure specified in the Instructions for Use (Reprocessing) of this instrument. · Class 2

Quick summary

This PENTAX FB-15V bronchoscope is a flexible or rigid scope used to look inside the airways. The 66 reports mostly describe the device breaking, leaking fluid or blood, and losing its fiber image, sometimes from fluid damage; none mention patient harm. There are 66 malfunction reports, no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
66
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 66 adverse event reports that list this exact device (DI 04961333224166), received between Jul 16, 2021 and Jul 19, 2024 — classified as 66 malfunctions.

The most frequently reported specific device problem is “Break” (40 reports).

2 other recalls cover devices cleared under the same FDA submission (K951199); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break40
  • Fluid/Blood Leak21
  • Component Missing1
  • Failure to Fold1
  • Optical Distortion1
  • Overcorrection1
  • Poor Quality Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions66

Recent adverse event reports

The 25 most recent of 66 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 19, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19785883 ↗
  • MalfunctionJun 25, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19602977 ↗
  • MalfunctionJun 18, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19555996 ↗
  • MalfunctionJun 12, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19517229 ↗
  • MalfunctionMar 11, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18872469 ↗
  • MalfunctionFeb 6, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18648001 ↗
  • MalfunctionJan 16, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18515363 ↗
  • MalfunctionJan 12, 2024Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18500974 ↗
  • MalfunctionDec 7, 2023Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18280416 ↗
  • MalfunctionOct 3, 2023Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17858450 ↗
  • MalfunctionOct 1, 2023Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17849585 ↗
  • MalfunctionJun 7, 2023Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17076472 ↗
  • MalfunctionMay 18, 2023Break; Break

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16951023 ↗
  • MalfunctionApr 26, 2023Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16812709 ↗
  • MalfunctionApr 14, 2023Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16741549 ↗
  • MalfunctionFeb 9, 2023Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16345167 ↗
  • MalfunctionFeb 7, 2023Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16309753 ↗
  • MalfunctionJan 31, 2023Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16274920 ↗
  • MalfunctionJan 18, 2023Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16193508 ↗
  • MalfunctionJan 5, 2023Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(OBJECTIVE LENS BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16100348 ↗
  • MalfunctionDec 21, 2022Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16020214 ↗
  • MalfunctionDec 18, 2022Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16001747 ↗
  • MalfunctionDec 12, 2022Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15957575 ↗
  • MalfunctionDec 6, 2022Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15921984 ↗
  • MalfunctionDec 1, 2022Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15891777 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K951199. One submission often covers many models, so check the product and lot details in each record.

  • Z-2065-2016 · Apr 6, 2016 · Terminated
    Bronchoscope Intended to provide optical visualization of and therapeutic access to the Airway and Bronchial Tree. This anatomy includes, but is not restricted to, the organs, tiss
  • Z-2076-2016 · Apr 6, 2016 · Terminated
    Ultrasound Bronchoscope Intended to provide optical visualization of, ultrasonic visualization of and therapeutic access to, the Pulmonary Tract including but not restricted to org

Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices

All BRONCHOSCOPE (FLEXIBLE OR RIGID) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.