Skip to content

PENTAX EB-1975KRecalls & reported problems

Hoya Corporation · PENTAX VIDEO BRONCHOSCOPE EB-1975K · Class 2

All models of PENTAX (72)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04961333226863), received between Feb 26, 2020 and Jan 17, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

6 other recalls cover devices cleared under the same FDA submission (K131028); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fluid/Blood Leak2
  • Insufficient Device Problem Information2
  • No Display/Image1
  • Optical Obstruction1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Viral Infection2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 17, 2025Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21175916 ↗
  • MalfunctionOct 28, 2024No Display/Image; No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20542145 ↗
  • MalfunctionOct 15, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20448308 ↗
  • MalfunctionMay 15, 2023Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16928497 ↗
  • MalfunctionOct 5, 2022Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15541744 ↗
  • MalfunctionSep 9, 2022Break; Break

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15388242 ↗
  • MalfunctionSep 5, 2022Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15354207 ↗
  • InjuryFeb 26, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    PENTAX MEDICAL BECAME AWARE OF A REPORT ON (B)(6) 2020 STATING, "POSSIBLE CONTAMINATION-PATIENT TESTED POSITIVE FOR VIRUS. HARBORING INFECTION." INVOLVING VIDEO BRONCHOSCOPE MODEL EB-1975K/SERIAL (B)(4). THE CUSTOMER OWNED BRONCHOSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON (B)(6) 2020. THE BRONCHOSCOPE WAS INSPECTED BY PENTAX MEDICAL SERVICE ON (B)(6) 2020 AND THE FOLLOWING INSPECTION FINDINGS WERE DOCUMENTED: INSERTION TUBE MILD CRUSH AT STAGE 2, INSERTION TUBE MILD CRUSH AT STAGE 1, PASSED WET LEAK TEST, PASSED DRY LEAK TEST. THE BRONCHOSCOPE UNDERWENT REPAIRS INCLUDING THE FOLLOWING COMPONENTS: DISTAL END ASSY WITH TUBE, DISTAL JOINT SCREW M1, BENDING RUBBER, BIOPSY INLET T-PIECE, O-RINGS AND SEALS, INSERTION/S-NIPPLE ATTACHING SCREW, O-RING(1.25X3.5), SUCTION CONNECTION TUBE, O-RING(1.2X3.5). THE VIDEO BRONCHOSCOPE IS CURRENTLY AWAITING ORGANIC SAMPLING AND QC FINAL APPROVAL AS OF (B)(6) 2020. PENTAX MEDICAL VIDEO BRONCHOSCOPE, MODEL EB-1975K, SERIAL NUMBER (B)(4), HAS BEEN PREVIOUSLY SERVICED AT A PENTAX FACILITY ONCE PRIOR SINCE THE DEVICE WAS PUT INTO SERVICE ON (B)(6) 2017.

    Patient problems: Viral Infection

    MAUDE report 9753436 ↗
  • InjuryFeb 26, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    PENTAX MEDICAL BECAME AWARE OF A REPORT ON (B)(6) 2020 STATING, "POSSIBLE CONTAMINATION-PATIENT TESTED POSITIVE FOR VIRUS. HARBORING INFECTION." INVOLVING VIDEO BRONCHOSCOPE MODEL EB-1975K/SERIAL (B)(4). THE CUSTOMER OWNED BRONCHOSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON 05-FEB-2020. THE BRONCHOSCOPE WAS INSPECTED BY PENTAX MEDICAL SERVICE ON 14-FEB-2020 AND THE FOLLOWING INSPECTION FINDINGS WERE DOCUMENTED: INSERTION TUBE MILD CRUSH AT STAGE 2, INSERTION TUBE MILD CRUSH AT STAGE 1, PASSED WET LEAK TEST, PASSED DRY LEAK TEST. THE BRONCHOSCOPE UNDERWENT REPAIRS INCLUDING THE FOLLOWING COMPONENTS: DISTAL END ASSY WITH TUBE, DISTAL JOINT SCREW M1, BENDING RUBBER, BIOPSY INLET T-PIECE, O-RINGS AND SEALS, INSERTION/S-NIPPLE ATTACHING SCREW, O-RING(1.25X3.5), SUCTION CONNECTION TUBE, O-RING(1.2X3.5). THE VIDEO BRONCHOSCOPE IS CURRENTLY AWAITING ORGANIC SAMPLING AND QC FINAL APPROVAL AS OF 25-FEB-2020. PENTAX MEDICAL VIDEO BRONCHOSCOPE, MODEL EB-1975K, SERIAL NUMBER (B)(4), HAS BEEN PREVIOUSLY SERVICED AT A PENTAX FACILITY ONCE PRIOR SINCE THE DEVICE WAS PUT INTO SERVICE ON 23-MAR-2017.

    Patient problems: Viral Infection

    MAUDE report 9753427 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131028. One submission often covers many models, so check the product and lot details in each record.

  • Z-2875-2016 · May 31, 2016 · Terminated
    Pentax Video Bronchoscope Designed to be used with a Pentax Video Processor (including light source), documentation equipment, video monitor, endo-therapy accessories (such as bio
  • Z-2065-2016 · Apr 6, 2016 · Terminated
    Bronchoscope Intended to provide optical visualization of and therapeutic access to the Airway and Bronchial Tree. This anatomy includes, but is not restricted to, the organs, tiss
  • Z-2076-2016 · Apr 6, 2016 · Terminated
    Ultrasound Bronchoscope Intended to provide optical visualization of, ultrasonic visualization of and therapeutic access to, the Pulmonary Tract including but not restricted to org
  • Z-2324-2017 · Jan 11, 2011 · Terminated
    Video Bronchoscope
  • Z-2326-2017 · Jan 11, 2011 · Terminated
    Video Naso pharyngo Laryngoscopes
  • Z-2325-2017 · Jan 11, 2011 · Terminated
    Video Cytoscopes

Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices

All BRONCHOSCOPE (FLEXIBLE OR RIGID) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page