The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04961333226863), received between Feb 26, 2020 and Jan 17, 2025 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
6 other recalls cover devices cleared under the same FDA submission (K131028); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Fluid/Blood Leak2
- Insufficient Device Problem Information2
- No Display/Image1
- Optical Obstruction1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Viral Infection2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 17, 2025Optical Obstruction; Optical Obstruction
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21175916 ↗MalfunctionOct 28, 2024No Display/Image; No Display/Image
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20542145 ↗MalfunctionOct 15, 2024Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20448308 ↗MalfunctionMay 15, 2023Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16928497 ↗MalfunctionOct 5, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15541744 ↗MalfunctionSep 9, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15388242 ↗MalfunctionSep 5, 2022Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15354207 ↗InjuryFeb 26, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
PENTAX MEDICAL BECAME AWARE OF A REPORT ON (B)(6) 2020 STATING, "POSSIBLE CONTAMINATION-PATIENT TESTED POSITIVE FOR VIRUS. HARBORING INFECTION." INVOLVING VIDEO BRONCHOSCOPE MODEL EB-1975K/SERIAL (B)(4). THE CUSTOMER OWNED BRONCHOSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON (B)(6) 2020. THE BRONCHOSCOPE WAS INSPECTED BY PENTAX MEDICAL SERVICE ON (B)(6) 2020 AND THE FOLLOWING INSPECTION FINDINGS WERE DOCUMENTED: INSERTION TUBE MILD CRUSH AT STAGE 2, INSERTION TUBE MILD CRUSH AT STAGE 1, PASSED WET LEAK TEST, PASSED DRY LEAK TEST. THE BRONCHOSCOPE UNDERWENT REPAIRS INCLUDING THE FOLLOWING COMPONENTS: DISTAL END ASSY WITH TUBE, DISTAL JOINT SCREW M1, BENDING RUBBER, BIOPSY INLET T-PIECE, O-RINGS AND SEALS, INSERTION/S-NIPPLE ATTACHING SCREW, O-RING(1.25X3.5), SUCTION CONNECTION TUBE, O-RING(1.2X3.5). THE VIDEO BRONCHOSCOPE IS CURRENTLY AWAITING ORGANIC SAMPLING AND QC FINAL APPROVAL AS OF (B)(6) 2020. PENTAX MEDICAL VIDEO BRONCHOSCOPE, MODEL EB-1975K, SERIAL NUMBER (B)(4), HAS BEEN PREVIOUSLY SERVICED AT A PENTAX FACILITY ONCE PRIOR SINCE THE DEVICE WAS PUT INTO SERVICE ON (B)(6) 2017.
Patient problems: Viral Infection
MAUDE report 9753436 ↗InjuryFeb 26, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
PENTAX MEDICAL BECAME AWARE OF A REPORT ON (B)(6) 2020 STATING, "POSSIBLE CONTAMINATION-PATIENT TESTED POSITIVE FOR VIRUS. HARBORING INFECTION." INVOLVING VIDEO BRONCHOSCOPE MODEL EB-1975K/SERIAL (B)(4). THE CUSTOMER OWNED BRONCHOSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON 05-FEB-2020. THE BRONCHOSCOPE WAS INSPECTED BY PENTAX MEDICAL SERVICE ON 14-FEB-2020 AND THE FOLLOWING INSPECTION FINDINGS WERE DOCUMENTED: INSERTION TUBE MILD CRUSH AT STAGE 2, INSERTION TUBE MILD CRUSH AT STAGE 1, PASSED WET LEAK TEST, PASSED DRY LEAK TEST. THE BRONCHOSCOPE UNDERWENT REPAIRS INCLUDING THE FOLLOWING COMPONENTS: DISTAL END ASSY WITH TUBE, DISTAL JOINT SCREW M1, BENDING RUBBER, BIOPSY INLET T-PIECE, O-RINGS AND SEALS, INSERTION/S-NIPPLE ATTACHING SCREW, O-RING(1.25X3.5), SUCTION CONNECTION TUBE, O-RING(1.2X3.5). THE VIDEO BRONCHOSCOPE IS CURRENTLY AWAITING ORGANIC SAMPLING AND QC FINAL APPROVAL AS OF 25-FEB-2020. PENTAX MEDICAL VIDEO BRONCHOSCOPE, MODEL EB-1975K, SERIAL NUMBER (B)(4), HAS BEEN PREVIOUSLY SERVICED AT A PENTAX FACILITY ONCE PRIOR SINCE THE DEVICE WAS PUT INTO SERVICE ON 23-MAR-2017.
Patient problems: Viral Infection
MAUDE report 9753427 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131028. One submission often covers many models, so check the product and lot details in each record.
- Z-2875-2016 · May 31, 2016 · TerminatedPentax Video Bronchoscope Designed to be used with a Pentax Video Processor (including light source), documentation equipment, video monitor, endo-therapy accessories (such as bio
- Z-2065-2016 · Apr 6, 2016 · TerminatedBronchoscope Intended to provide optical visualization of and therapeutic access to the Airway and Bronchial Tree. This anatomy includes, but is not restricted to, the organs, tiss
- Z-2076-2016 · Apr 6, 2016 · TerminatedUltrasound Bronchoscope Intended to provide optical visualization of, ultrasonic visualization of and therapeutic access to, the Pulmonary Tract including but not restricted to org
- Z-2324-2017 · Jan 11, 2011 · TerminatedVideo Bronchoscope
- Z-2326-2017 · Jan 11, 2011 · TerminatedVideo Naso pharyngo Laryngoscopes
- Z-2325-2017 · Jan 11, 2011 · TerminatedVideo Cytoscopes
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page