PENTAX FB-18VRecalls & reported problems
Hoya Corporation · Perform cleaning, high-level disinfection, and/or sterilization according to the procedure specified in the Instructions for Use (Reprocessing) of this instrument. · Class 2
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 04961333227624), received between Nov 8, 2020 and Jun 28, 2024 — classified as 22 malfunctions.
The most frequently reported specific device problem is “Break” (15 reports).
2 other recalls cover devices cleared under the same FDA submission (K951199); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break15
- Fluid/Blood Leak6
- Device Reprocessing Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions22
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 28, 2024Break; Break
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19630897 ↗MalfunctionJun 28, 2024Break; Break
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19630890 ↗MalfunctionJun 26, 2024Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19612231 ↗MalfunctionMar 1, 2024Break; Break
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18813827 ↗MalfunctionDec 20, 2023Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18361558 ↗MalfunctionDec 11, 2023Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18298863 ↗MalfunctionOct 3, 2023Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17858452 ↗MalfunctionAug 29, 2023Break; Break
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17639740 ↗MalfunctionJul 21, 2023Break; Break
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17369381 ↗MalfunctionMay 16, 2023Break; Break
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16936711 ↗MalfunctionApr 16, 2023Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16748898 ↗MalfunctionApr 11, 2023Break; Break
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16711350 ↗MalfunctionFeb 22, 2023Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16415363 ↗MalfunctionFeb 6, 2023Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16302237 ↗MalfunctionJan 19, 2023Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16200714 ↗MalfunctionDec 13, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15966587 ↗MalfunctionDec 12, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15957563 ↗MalfunctionNov 10, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. DISTAL BODY DAMAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15768195 ↗MalfunctionNov 6, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15741462 ↗MalfunctionSep 7, 2022Fluid/Blood Leak; Fluid/Blood Leak
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15367204 ↗MalfunctionAug 31, 2022Break; Break
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15332688 ↗MalfunctionNov 8, 2020Device Reprocessing Problem; Device Reprocessing Problem
PENTAX MEDICAL WAS MADE AWARE OF AN EVENT THAT OCCURRED DURING PRE-PROCEDURAL INSPECTION IN THE UNITED STATES. THE REPORTED COMPLAINT RECEIVED UNDER NOTIFICATION (B)(4) FOR ACCESSORY STUCK IN SCOPE "WHITE PIECES STUCK IN DOES NOT KNOW WHAT EXACTLY IT IS", INVOLVING THE PENTAX MEDICAL FIBEROPTIC BRONCHOSCOPE, MODEL FB-18V, SERIAL NUMBER (B)(4). THE USER FACILITY RESPONDED TO A GFE REQUEST VIA EMAIL ON 09-OCT-2020, AND STATED THE WHITE PIECES STUCK IN THE ENDOSCOPE WERE FOUND DURING THE PRE-INSPECTION CHECK AND REMOVED FROM CIRCULATION IMMEDIATELY AFTER THE FAILURE/EVENT OCCURRED AND SUBSEQUENTLY CALLED IN FOR SERVICE/REPLACEMENT. NO SERIOUS INJURY OR DEATH OF A PATIENT OR USER, OR DELAY IN A PROCEDURE WHICH WOULD REQUIRE MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER OWNED FIBEROPTIC BRONCHOSCOPE WAS RETURNED FOR EVALUATION ON 05-OCT-2020. DURING EVALUATION OF THE ENDOSCOPE UNDER SERVICE ORDER (B)(4), THE PENTAX MEDICAL SERVICE REPAIR TECHNICIAN WAS UNABLE TO CONFIRM THE CUSTOMERS EXPERIENCE AND DOCUMENTED THE FOLLOWING FINDINGS ON 06-OCT-2020: IMAGE ALIGNMENT OUT, PASSED WET LEAK TEST, PASSED DRY LEAK TEST. THE DEVICE UNDERWENT REPAIRS INCLUDING THE FOLLOWING COMPONENTS: O-RINGS A
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10806329 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K951199. One submission often covers many models, so check the product and lot details in each record.
- Z-2076-2016 · Apr 6, 2016 · TerminatedUltrasound Bronchoscope Intended to provide optical visualization of, ultrasonic visualization of and therapeutic access to, the Pulmonary Tract including but not restricted to org
- Z-2065-2016 · Apr 6, 2016 · TerminatedBronchoscope Intended to provide optical visualization of and therapeutic access to the Airway and Bronchial Tree. This anatomy includes, but is not restricted to, the organs, tiss
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.