Skip to content

OPTIVISTA EPK-i7010 (UDI 04961333233618)Recalls & reported problems

Hoya Corporation · PENTAX Medical Video Processor EPK-i7010 · Class 2

All models of OPTIVISTA (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04961333233618), received between Apr 16, 2025 and Apr 13, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Erratic or Intermittent Display” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Erratic or Intermittent Display2
  • No Display/Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 13, 2026Erratic or Intermittent Display

    ON 18-MAR-2026, PENTAX MEDICAL BECAME AWARE OF AN EVENT INVOLVING A PENTAX MEDICAL VIDEO PROCESSOR MODEL EPK-I7010, SERIAL NUMBER (B)(6) IN CANADA WITHIN THE (B)(6). PENTAX MEDICAL RECEIVED, VIA EMAIL FROM HEALTH CANADA, A COPY OF A REPORT SUBMITTED BY THE USER FACILITY TO HEALTH CANADA (REPORT NUMBER: (B)(4) / LRN-20260311-432 / RCT-20260311-371). THE REPORT DESCRIBED THAT DURING AN ENDOSCOPIC PROCEDURE, A MALFUNCTION OCCURRED IN THE LIGHT SOURCE OF THE PROCESSOR, CAUSING THE ENDOSCOPIC IMAGE TO FLICKER FOR SEVERAL MINUTES. THE PROCEDURE WAS TEMPORARILY INTERRUPTED WHILE THE BIOMEDICAL ENGINEERING DEPARTMENT INTERVENED. THE ENDOSCOPIC IMAGE WAS NOT ADEQUATE, AND IT WAS REPORTED THAT THE PATIENT WAS EXPOSED TO POTENTIAL RISKS SUCH AS PERFORATION. THERE WAS NO REPORT OF PATIENT HARM. THIS EVENT OCCURRED DURING USE. THIS EVENT MEETS THE REQUIREMENTS FOR FDA REPORTABILITY. HOWEVER, SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24854811 ↗
  • MalfunctionDec 22, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23861444 ↗
  • MalfunctionApr 16, 2025No Display/Image; No Display/Image

    ON (B)(6) 2025, PENTAX MEDICAL BECAME AWARE OF AN EVENT INVOLVING A MODEL EPK-I7010, SERIAL NUMBER (B)(6) VIDEO PROCESSOR IN ARGENTINA WITHIN THE PAI REGION. THE LAMP OF THE PROCESSOR DID NOT TURN ON, AND AN ERROR MESSAGE (19-06/59-06) WAS DISPLAYED. A PENTAX MEDICAL ENGINEER CHECKED THE CABLE CONNECTION AND THE OE SLIDE MOTOR BUT FOUND NO ABNORMALITIES, AND THE ISSUE WAS NOT RESOLVED. WHEN THE PROCESSOR WAS INSPECTED USING THE PERIPHERAL DEVICE CIRCUIT BOARD, IT OPERATED NORMALLY. THIS EVENT OCCURRED AT THE TIME OF DURING USE. THERE WAS NO REPORT OF PATIENT HARM. THIS EVENT MEETS THE REQUIREMENTS FOR FDA REPORTABILITY; HOWEVER, SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, MANUFACTURER OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21847479 ↗

Other Endoscope, accessories, image post-processing for color enhancement devices

All Endoscope, accessories, image post-processing for color enhancement devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page