PROGREAT MC*PV2415YRecalls & reported problems
Terumo Corporation · PROGREAT Microcatheter System · Class 2
All models of PROGREAT (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04987350718440), received between Mar 5, 2024 and Mar 5, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Split, Cut or Torn2
- Physical Resistance/Sticking2
- Stretched2
Patient problems
- Insufficient Information2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 5, 2024Stretched; Material Split, Cut or Torn; Physical Resistance/Sticking; Stretched; Material Split, Cut or Torn; Physical Resistance/Sticking
THE USER FACILITY REPORTED THAT A 6 FRENCH 90CM DESTINATION SHEATH WAS PLACED INTO THE LEFT BRACHIAL ARTERY AND EXTENDED INTO THE DISTAL AORTA. AN 035 RUBICON CATHETER WAS PLACED INTO A SMALL BRANCH OFF THE RIGHT INTERNAL ILIAC ARTERY AND A 018 THRUWAY WIRE WAS PLACED. THE 2.4 FRENCH PROGREAT WAS PLACED OVER THE WIRE WITH GREAT RESISTANCE FELT BY THE TECHNOLOGIST LOADING IT. IT WAS NOTED BY THE PHYSICIAN THAT THE TECHNOLOGIST WAS ACCIDENTALLY PINNING THE CATHETER TIP AND PUSHING THE BACK END OF THE CATHETER. AFTER THE TECHNOLOGIST REPOSITIONED THEIR HANDS, THE CATHETER WAS ABLE TO BE DELIVERED TO THE APPROPRIATE LOCATION OVER THE WIRE. THE WIRE WAS REMOVED, AND THE 018 10CM HYDOPACK COIL WAS INTRODUCED. ONE COIL WAS SUCCESSFULLY DELIVERED. THE SECOND COIL WAS DELIVERED; HOWEVER, INTENDED TO BE REPOSITIONED AND UPON REMOVAL, THE COIL BECAME DETACHED DUE TO RESISTANCE AND THE TAIL OF THE COIL WAS HANGING INTO THE COMMON ILIAC ARTERY IN AN UNWANTED LOCATION. THE PROGREAT WAS REMOVED; HOWEVER, IT WAS ELONGATED AND PARTIALLY TORN. THE PHYSICIAN DECIDED THAT NO ADDITIONAL INTERVENTION WAS NECESSARY AT THIS TIME. THE PROCEDURE PERFORMED WAS A TYPE 2 ENDOLEAK EMBOLIZATION OF RIGHT INTERNAL
Patient problems: Insufficient Information
MAUDE report 18838280 ↗InjuryMar 5, 2024Stretched; Material Split, Cut or Torn; Physical Resistance/Sticking; Stretched; Material Split, Cut or Torn; Physical Resistance/Sticking
THE USER FACILITY REPORTED THAT A 6 FRENCH 90CM DESTINATION SHEATH WAS PLACED INTO THE LEFT BRACHIAL ARTERY AND EXTENDED INTO THE DISTAL AORTA. AN 035 RUBICON CATHETER WAS PLACED INTO A SMALL BRANCH OFF THE RIGHT INTERNAL ILIAC ARTERY AND A 018 THRUWAY WIRE WAS PLACED. THE 2.4 FRENCH PROGREAT WAS PLACED OVER THE WIRE WITH GREAT RESISTANCE FELT BY THE TECHNOLOGIST LOADING IT. IT WAS NOTED BY THE PHYSICIAN THAT THE TECHNOLOGIST WAS ACCIDENTALLY PINNING THE CATHETER TIP AND PUSHING THE BACK END OF THE CATHETER. AFTER THE TECHNOLOGIST REPOSITIONED THEIR HANDS, THE CATHETER WAS ABLE TO BE DELIVERED TO THE APPROPRIATE LOCATION OVER THE WIRE. THE WIRE WAS REMOVED, AND THE 018 10CM HYDOPACK COIL WAS INTRODUCED. ONE COIL WAS SUCCESSFULLY DELIVERED. THE SECOND COIL WAS DELIVERED; HOWEVER, INTENDED TO BE REPOSITIONED AND UPON REMOVAL, THE COIL BECAME DETACHED DUE TO RESISTANCE AND THE TAIL OF THE COIL WAS HANGING INTO THE COMMON ILIAC ARTERY IN AN UNWANTED LOCATION. THE PROGREAT WAS REMOVED; HOWEVER, IT WAS ELONGATED AND PARTIALLY TORN. THE PHYSICIAN DECIDED THAT NO ADDITIONAL INTERVENTION WAS NECESSARY AT THIS TIME. THE PROCEDURE PERFORMED WAS A TYPE 2 ENDOLEAK EMBOLIZATION OF RIGHT INTERNAL
Patient problems: Insufficient Information
MAUDE report 18838156 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page