Skip to content

GLIDEWIRE RG*GA1818SPRecalls & reported problems

Terumo Corporation · GLIDEWIRE GT Guide Wire · Class 2

All models of GLIDEWIRE (30)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04987350734808), received between Feb 2, 2023 and Feb 2, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Separation2

Patient problems

  • Foreign Body In Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 2, 2023Material Separation; Material Separation

    THE USER FACILITY REPORTED THAT THE WHEN THE GLIDEWIRE INVOLVED WAS USED, IT WAS SPUN, AND THE WIRE BROKE AT THAT MOMENT. THE PATIENT WAS IN STABLE CONDITION. THE SHEATH WAS USED INSIDE A TRAILBLAZER. THERE WAS NO PATIENT INJURY/MEDICAL OR SURGICAL INTERVENTION REQUIRED. THE PROCEDURE WAS STOPPED. THE ESTIMATED BLOOD LOSS WAS LESS THAN 250CC'S. ADDITIONAL INFORMATION WAS RECEIVED ON 18JAN2023: THE PROCEDURE PERFORMED WHEN THE ISSUE OCCURRED WAS AN ANGIOGRAM. THE ANGIOGRAM WAS PERFORMED PRIOR TO THE PROCEDURE. THERE WERE NO DIFFICULTIES WITH ARTERIAL ACCESS, SHEATH, GUIDEWIRE, OR CATHETER INSERTION. THERE WAS NO TORTUOSITY NOTED IN THE PATIENT'S ANATOMY. APPROXIMATELY 300 MM OF THE DEVICE BROKE OFF. THE ISSUE RESOLVED FOR THE PATIENT BY BEING ASPIRATED ON THE TRAILBLAZER CATHETER AND REMOVED FROM THE SHEATH. THE BROKEN WIRE WAS REMOVED FROM THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON 27JAN2023: THE WIRE BROKE WHEN IT WAS INSERTED AND MANIPULATED INSIDE THE PATIENT'S BODY.

    Patient problems: Foreign Body In Patient

    MAUDE report 16286923 ↗
  • InjuryFeb 2, 2023Material Separation; Material Separation

    ADDITIONAL INFORMATION WAS RECEIVED ON 14 FEB 2023: WHEN THE DOCTOR WAS MANIPULATING THE WIRE, THE WIRE FRACTURED AND BROKE OFF INSIDE OF THE PATIENT, WITHIN THE SHEATH AND TRAILBLAZER CATHETER. THIS ALLOWED THE DOCTOR TO IMMEDIATELY RETRIEVE THE WIRE BY ASPIRATING THE CATHETER. NO WIRE FRAGMENT REMAINED IN THE PATIENT POST PROCEDURE. THE GLIDEWIRE FRACTURED AND DETACHED WITHIN THE PATIENT'S SHEATH AND GUIDE CATHETER. THERE WERE NO COMPLICATIONS OCCURRED AS RESULT OF THE DEVICE. THE PATIENT'S OUTCOME WAS NO INJURY/UNAFFECTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 16286139 ↗

Other Wire, guide, catheter devices

All Wire, guide, catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page