PROGREAT MC*PE24151YVRecalls & reported problems
Terumo Corporation · PROGREAT Microcatheter Coaxial System · Class 2
All models of PROGREAT (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04987350735287), received between Feb 10, 2026 and Feb 10, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Material Deformation2
- Material Separation2
Patient problems
- Foreign Body In Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 10, 2026Break; Material Separation; Material Deformation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: DURING THE SPLENIC ARTERY EMBOLIZATION, THE PROGREAT WAS INSERTED THROUGH GLIDECATH (INSIDE DESTINATION SHEATH) AND ADVANCED TO THE SPLENIC ARTERY. PER THE MEDICAL DOCTOR (MD), THE COIL 5X16 CX BECAME STUCK AND COULD NOT PUSH OR PULL COIL; THEREFORE, A COIL WAS DEPLOYED AND ATTEMPTED TO PUSH WITH 018 COMMAND WIRE. THE WIRE COULD NOT PUSH THE COIL, PER MD THE "PROGREAT BROKE". THE MD WAS ABLE TO RECOVER THE PROGREAT AND WAS ABLE TO RELEASE COIL PACK. COIL PACK LANDED IN THE INTENDED LOCATION, AND PROGREAT WAS REMOVED AND REPLACED. EMBOLIZATION WAS THEN COMPLETED, SUCCESSFUL, AND THE PATIENT WAS OK. LATER, THE MD STATED HE BELIEVED THE PROGREAT BECAME STUCK IN GLIDECATH DUE TO INADEQUATE FLUSHING, AND THE COIL BECAME STUCK FOR SIMILAR REASONS. ADDITIONAL INFORMATION WAS RECEIVED ON 31DEC2025: THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE. THEY WERE TAKEN BACK TO THE OPERATING ROOM (OR) FOR ANOTHER PROCEDURE THE FOLLOWING DAY, THOUGH IT WAS NOT THOUGHT THE PREVIOUS PROCEDURE WAS AT FAULT. THE TREATMENT WAS SUCCESSFUL; THE SPLENIC ARTERY WAS EMBOLIZED RESULTING IN REDUCTION OF BLEEDING FROM THE SPLEEN. NO ADDITION
Patient problems: Foreign Body In Patient
MAUDE report 24305898 ↗InjuryFeb 10, 2026Break; Material Separation; Material Deformation
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: DURING THE SPLENIC ARERTY EMBOLIZATION, THE PROGREAT WAS INSERTED THROUGH GLIDECATH (INSIDE DESTINATION SHEATH) AND ADVANCED TO THE SPLENIC ARTERY. PER THE MEDICAL DOCTOR (MD), THE COIL 5X16 CX BECAME STUCK AND COULD NOT PUSH OR PULL COIL; THEREFORE, A COIL WAS DEPLOYED AND ATTEMPTED TO PUSH WITH 018 COMMAND WIRE. THE WIRE COULD NOT PUSH THE COIL, PER MD THE "PROGREAT BROKE". THE MD WAS ABLE TO RECOVER THE PROGREAT AND WAS ABLE TO RELEASE COIL PACK. COIL PACK LANDED IN THE INTENDED LOCATION, AND PROGREAT WAS REMOVED AND REPLACED. EMBOLIZATION WAS THEN COMPLETED, SUCCESSFUL, AND THE PATIENT WAS OK. LATER, THE MD STATED HE BELIEVED THE PROGREAT BECAME STUCK IN GLIDECATH DUE TO INADEQUATE FLUSHING, AND THE COIL BECAME STUCK FOR SIMILAR REASONS. ADDITIONAL INFORMATION WAS RECEIVED ON 31DEC2025: THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE. THEY WERE TAKEN BACK TO THE OPERATING ROOM (OR) FOR ANOTHER PROCEDURE THE FOLLOWING DAY, THOUGH IT WAS NOT THOUGHT THE PREVIOUS PROCEDURE WAS AT FAULT. THE TREATMENT WAS SUCCESSFUL; THE SPLENIC ARTERY WAS EMBOLIZED RESULTING IN REDUCTION OF BLEEDING FROM THE SPLEEN. NO ADDITION
Patient problems: Foreign Body In Patient
MAUDE report 24305896 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H749085263001Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.