PROGREAT MC*PE24151YVRecalls & reported problems

Terumo Corporation · PROGREAT Microcatheter Coaxial System · Class 2

All models of PROGREAT (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04987350735287), received between Feb 10, 2026 and Feb 10, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Deformation2
  • Material Separation2

Patient problems

  • Foreign Body In Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 10, 2026Break; Material Separation; Material Deformation

    TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: DURING THE SPLENIC ARTERY EMBOLIZATION, THE PROGREAT WAS INSERTED THROUGH GLIDECATH (INSIDE DESTINATION SHEATH) AND ADVANCED TO THE SPLENIC ARTERY. PER THE MEDICAL DOCTOR (MD), THE COIL 5X16 CX BECAME STUCK AND COULD NOT PUSH OR PULL COIL; THEREFORE, A COIL WAS DEPLOYED AND ATTEMPTED TO PUSH WITH 018 COMMAND WIRE. THE WIRE COULD NOT PUSH THE COIL, PER MD THE "PROGREAT BROKE". THE MD WAS ABLE TO RECOVER THE PROGREAT AND WAS ABLE TO RELEASE COIL PACK. COIL PACK LANDED IN THE INTENDED LOCATION, AND PROGREAT WAS REMOVED AND REPLACED. EMBOLIZATION WAS THEN COMPLETED, SUCCESSFUL, AND THE PATIENT WAS OK. LATER, THE MD STATED HE BELIEVED THE PROGREAT BECAME STUCK IN GLIDECATH DUE TO INADEQUATE FLUSHING, AND THE COIL BECAME STUCK FOR SIMILAR REASONS. ADDITIONAL INFORMATION WAS RECEIVED ON 31DEC2025: THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE. THEY WERE TAKEN BACK TO THE OPERATING ROOM (OR) FOR ANOTHER PROCEDURE THE FOLLOWING DAY, THOUGH IT WAS NOT THOUGHT THE PREVIOUS PROCEDURE WAS AT FAULT. THE TREATMENT WAS SUCCESSFUL; THE SPLENIC ARTERY WAS EMBOLIZED RESULTING IN REDUCTION OF BLEEDING FROM THE SPLEEN. NO ADDITION

    Patient problems: Foreign Body In Patient

    MAUDE report 24305898 ↗
  • InjuryFeb 10, 2026Break; Material Separation; Material Deformation

    TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: DURING THE SPLENIC ARERTY EMBOLIZATION, THE PROGREAT WAS INSERTED THROUGH GLIDECATH (INSIDE DESTINATION SHEATH) AND ADVANCED TO THE SPLENIC ARTERY. PER THE MEDICAL DOCTOR (MD), THE COIL 5X16 CX BECAME STUCK AND COULD NOT PUSH OR PULL COIL; THEREFORE, A COIL WAS DEPLOYED AND ATTEMPTED TO PUSH WITH 018 COMMAND WIRE. THE WIRE COULD NOT PUSH THE COIL, PER MD THE "PROGREAT BROKE". THE MD WAS ABLE TO RECOVER THE PROGREAT AND WAS ABLE TO RELEASE COIL PACK. COIL PACK LANDED IN THE INTENDED LOCATION, AND PROGREAT WAS REMOVED AND REPLACED. EMBOLIZATION WAS THEN COMPLETED, SUCCESSFUL, AND THE PATIENT WAS OK. LATER, THE MD STATED HE BELIEVED THE PROGREAT BECAME STUCK IN GLIDECATH DUE TO INADEQUATE FLUSHING, AND THE COIL BECAME STUCK FOR SIMILAR REASONS. ADDITIONAL INFORMATION WAS RECEIVED ON 31DEC2025: THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE. THEY WERE TAKEN BACK TO THE OPERATING ROOM (OR) FOR ANOTHER PROCEDURE THE FOLLOWING DAY, THOUGH IT WAS NOT THOUGHT THE PREVIOUS PROCEDURE WAS AT FAULT. THE TREATMENT WAS SUCCESSFUL; THE SPLENIC ARTERY WAS EMBOLIZED RESULTING IN REDUCTION OF BLEEDING FROM THE SPLEEN. NO ADDITION

    Patient problems: Foreign Body In Patient

    MAUDE report 24305896 ↗