Skip to content

GLIDECATH CG412Recalls & reported problems

Terumo Corporation · RADIFOCUS GLIDECATH · Class 2

Adverse event reports
7
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04987350772138), received between Jun 7, 2017 and May 15, 2025 — classified as 5 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation3
  • Fracture2
  • Break1

Patient problems

  • Foreign Body In Patient3
  • Calcium Deposits/Calcification2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2025Fracture; Fracture

    THE USER FACILITY REPORTED THAT THE 4FR GLIDECATH WAS PLACED UP AND OVER THE AORTIC BIFURCATION CONTRALATERALLY FROM THE LEFT SIDE. WHILE PULLING THE CATHETER BACK ACROSS THE BIFURCATION THE CATHETER TORE WITH APPROXIMATELY 30% OF THE CATHETER LODGING ITSELF IN THE RIGHT COMMON ILIAC. THE RIGHT GROIN WAS ACCESSED, AND THE BROKEN PIECE OF CATHETER WAS SNARED AND SAFELY REMOVED. THE PHYSICIAN STATED THAT THE PATIENTS' VESSELS WERE CALCIFIED. THE WERE NO BLOOD LOSS. THE PATIENT CONDITION WAS STABLE. THE PROCEDURE WAS AN AORTA BILATERAL FEMORAL ANGIOGRAM. ADDITIONAL INFORMATION WAS RECEIVED ON 18APR2025: THE DOCTOR BELIEVED THE PATIENTS CALCIFICATION HAD SOMETHING TO DO WITH THE ISSUE.

    Patient problems: Foreign Body In Patient

    MAUDE report 22024886 ↗
  • InjuryMay 15, 2025Fracture; Fracture

    THE USER FACILITY REPORTED THAT THE 4FR GLIDECATH WAS PLACED UP AND OVER THE AORTIC BIFURCATION CONTRALATERALLY FROM THE LEFT SIDE. WHILE PULLING THE CATHETER BACK ACROSS THE BIFURCATION THE CATHETER TORE WITH APPROXIMATELY 30% OF THE CATHETER LODGING ITSELF IN THE RIGHT COMMON ILIAC. THE RIGHT GROIN WAS ACCESSED, AND THE BROKEN PIECE OF CATHETER WAS SNARED AND SAFELY REMOVED. THE PHYSICIAN STATED THAT THE PATIENTS' VESSELS WERE CALCIFIED. THE WERE NO BLOOD LOSS. THE PATIENT CONDITION WAS STABLE. THE PROCEDURE WAS AN AORTA BILATERAL FEMORAL ANGIOGRAM. ADDITIONAL INFORMATION WAS RECEIVED ON 18APR2025: THE DOCTOR BELIEVED THE PATIENTS CALCIFICATION HAD SOMETHING TO DO WITH THE ISSUE.

    Patient problems: Foreign Body In Patient

    MAUDE report 22023932 ↗
  • MalfunctionApr 29, 2024

    ELDERLY MALE WITH HISTORY OF PERIPHERAL ARTERIAL DISEASE AND RECENT GANGRENE OF BILATERAL LOWER EXTREMITIES. PROCEDURE: AORTOGRAM AND ANGIOPLASTY OF THE LEFT COMMON ILIAC ARTERY AND EXTERNAL ILIAC ARTERY WITH A 6 MM X 100 MM DRUG-COATED BALLOON, AND ANGIOPLASTY OF THE COMMON FEMORAL ARTERY AND SUPERFICIAL FEMORAL ARTERY WITH A 6 MM X 6 CM DRUG-COATED BALLOON AND MYNX. CATHETER BROKE INTO 2 PIECES INSIDE THE PATIENT. SURGEON SUCCESSFULLY RETRIEVED BOTH PIECES WITHOUT KNOWN INJURY. PATIENT DISCHARGED THE SAME DAY.

    Patient problems: Foreign Body In Patient

    MAUDE report 19201354 ↗
  • InjuryApr 4, 2022Material Separation; Material Separation

    THE USER FACILITY REPORTED THAT THE GLIDECATH DEVICE INVOLVED SHEARED OFF INSIDE THE PATIENT DUE TO SEVERE CALCIUM DEPOSITS. THE PIECE WAS RETRIEVED WITH A SNARE AND THE PATIENT WAS UNAFFECTED. THERE PATIENT CONDITION WAS FINE. THE PROCEDURE OUTCOME WAS FINE. THERE WAS NO PATIENT INJURY/SURGICAL INTERVENTION REQUIRED. ADDITIONAL INFORMATION WAS RECEIVED ON 29 MAR 2022: THE PROCEDURE PERFORMED WAS A THROMBOENDOARTERECTOMY FEMORAL ARTERY AND ABD AORTOGRAPHY.

    Patient problems: Calcium Deposits/Calcification

    MAUDE report 13998719 ↗
  • InjuryApr 4, 2022Material Separation; Material Separation

    THE USER FACILITY REPORTED THAT THE GLIDECATH DEVICE INVOLVED SHEARED OFF INSIDE THE PATIENT DUE TO SEVERE CALCIUM DEPOSITS. THE PIECE WAS RETRIEVED WITH A SNARE AND THE PATIENT WAS UNAFFECTED. THERE PATIENT CONDITION WAS FINE. THE PROCEDURE OUTCOME WAS FINE. THERE WAS NO PATIENT INJURY/SURGICAL INTERVENTION REQUIRED. ADDITIONAL INFORMATION WAS RECEIVED ON 29 MAR 2022: THE PROCEDURE PERFORMED WAS A THROMBOENDOARTERECTOMY FEMORAL ARTERY AND ABD AORTOGRAPHY.

    Patient problems: Calcium Deposits/Calcification

    MAUDE report 13998495 ↗
  • InjuryJul 18, 2018Material Separation; Material Separation

    THE USER FACILITY REPORTED THE GLIDECATH WAS INSERTED THROUGH 6 FR PINNACLE R/O OVER AMPLATZ WIRE. WHEN IT WAS REMOVED, THE TECH NOTICED THAT THE END OF THE CATHETER LOOKED DIFFERENT THAN NORMAL. IT WAS QUICKLY REALIZED THAT A DISTAL PART OF THE CATHETER HAD BROKEN OFF. THE END OF THE CATHETER WAS EASILY RETRIEVED WITH AN ENSNARE DEVICE. THE PATIENT DID FINE AND THE REST OF THE PROCEDURE CONTINUED AS NORMAL. THE PATIENT WAS UNAFFECTED AND DID FINE. THE PROCEDURE OUTCOME WAS REPORTED TO BE A SUCCESS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7698183 ↗
  • MalfunctionJun 7, 2017Break; Break

    THE USER FACILITY REPORTED BREAKAGE WITH THE INVOLVED GLIDECATH DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE USER FACILITY: THEY USED THE CATHETER TO CROSS THE BIFURCATION AND WHEN THEY WERE TRYING TO REMOVE IT OVER THE WIRE, IT GOT STUCK ON THE WIRE AND SNAPPED IN HALF; THE PROCEDURE WAS COMPLETED; AND THE PATIENT WAS REPORTED TO BE IN NORMAL CONDITION. ADDITIONAL INFORMATION WAS RECEIVED ON JUNE 1, 2017. THE CATHETER WAS ABLE TO BE REMOVED COMPLETELY. THE PATIENT WAS REPORTED TO BE STABLE. THE PROCEDURE WAS SUCCESSFUL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6618624 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →