GLIDEWIRE GR1805Recalls & reported problems
Terumo Corporation · RADIFOCUS GLIDEWIRE · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04987350772855), received between Jun 27, 2018 and Sep 3, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Break1
- Difficult to Remove1
- Material Separation1
- Physical Resistance/Sticking1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Code Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 3, 2026Break; Difficult to Remove; Material Separation; Physical Resistance/Sticking
Terumo received the following information: It was reported that A 72-year-old male was undergoing a cardiac device implantation procedure when a 0.018-inch guidewire became trapped beneath the left clavicle and could not be removed through the sheath. A CT scan indicated the wire was located superficially in the subcutaneous tissue. The patient was brought back to the electrophysiology lab, prepped and draped in standard fashion. Fluoroscopy was used to mark the location of the wire. This wire was not attached to anything in the heart and was essentially entrapped behind the clavicle, likely in the fascia. An incision was made over the sternoclavicular joint, and the dissection was carried down into the subcutaneous tissue over the head of the clavicle. Without entering the left pleural space, we followed the clavicle and located the wire. lt was grasped with the hemostat and retrieved without difficulty. We then irrigated, making sure there was no bleeding. Satisfied, the wound was closed in a layered fashion with absorbable suture. The patient tolerated the procedure well. There were no unforeseen complications. During advancement of the glidewire through the patient's left subcl
Patient problems: Foreign Body In Patient
MAUDE report 50590719 ↗InjuryJan 30, 2019Fracture; Fracture
THE USER FACILITY REPORTED THAT DURING THE RADIOLOGY INSERTION OF CCIP LINE, IT WAS DISCOVERED THAT THE OUTER MANTLE OF THE GLIDEWIRE WAS REMOVED FROM THE GUIDE IN THE PATIENT'S ARM. ADDITIONAL INFORMATION WAS RECEIVED ON JANUARY 28, 2019. THE PROCEDURE WAS A PICC LINE. THE MANTLE IS THE BLACK COVER PART. THE MANTLE OF THE GUIDEWIRE WAS STUCK IN THE PATIENTS ARM. THE OUTER MANTLE OF THE GUIDEWIRE WAS ABLE TO BE PULLED OUT. THE PATIENT WAS REPORTED TO BE DOING WELL. THE PROCEDURE WAS STOPPED.
Patient problems: No Code Available
MAUDE report 8290365 ↗InjuryJun 27, 2018Fracture; Fracture
THE USER FACILITY REPORTED BREAKAGE ON THE INVOLVED RF GUIDEWIRE DEVICE. THE PROCEDURE WAS A PERMA CATH INSERTION, UTILIZING THE RIGHT INTERNAL JUGULAR, WHICH WAS VERY TORTUOUS. ACCESS WAS OBTAINED WITH A MICROPUNCTURE KIT AND INTRODUCER 5FR, GLIDEWIRE UTILIZED VIA THE MICROPUNCTURE SHEATH INTRODUCER, NOT THE ACCESS NEEDLE. IT WAS NOTED AND APPEARED THE GLIDEWIRE TIP FRACTURED, WITH THE TIP REMAINING IN THE EXTERNAL JUGULAR VEIN, COILED, VERY TORTUOUS TINY VEIN, SUSPECTING PROBABLY INTIMA (THE INNERMOST COATING OF A VEIN OR ARTERY). IT WAS REPORTED THAT THE DOCTOR CHOOSE NOT TO INTERVENE AT THIS TIME FOR RETRIEVAL OF TIP. THE PATIENT WAS REPORTED TO BE "OK". IT WAS REPORTED THAT THE PERMACHATH INSERTION WAS SUCCESSFUL.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7642642 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page