GLIDEWIRE GR3806Recalls & reported problems
Terumo Corporation · RADIFOCUS GLIDEWIRE · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04987350773357), received between Aug 2, 2019 and Aug 3, 2020 — classified as 4 injuries.
The most frequently reported specific device problem is “Material Separation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Separation4
Patient problems
- Device Embedded In Tissue or Plaque2
- No Consequences Or Impact To Patient2
- No Code Available1
- No Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 3, 2020Material Separation; Material Separation
THE USER FACILITY REPORTED THAT THE DOCTOR INSERTED THE GLIDEWIRE THROUGH HIS ACCESS NEEDLE AND IT SHEARED OFF ABOUT 30-40CM OF THE JACKET OF THE WIRE. THE DOCTOR HAD TO ACCESS THE OPPOSITE FEMORAL ARTERY TO RETRIEVE THE DETACHED SLEEVE WITH A SNARE. THE PATIENT'S CONDITION WAS STABLE. THERE WAS NO PATIENT INJURY, MEDICAL/SURGICAL INTERVENTION REQUIRED. ADDITIONAL INFORMATION WAS RECEIVED ON 23JUL2020. AN ANGIO PROCEDURE WAS BEING PERFORMED. THEY ARE NOT AWARE THAT ANY OF THE GLIDEWIRE WAS LEFT IN THE PATIENT. FLUOROSCOPY WAS USED TO SNARE THE JACKET. THERE WAS NO HARM TO THE PATIENT. ONCE JACKET PORTION WAS SNARED, THEY FINISHED GETTING THEIR IMAGING.
Patient problems: No Consequences Or Impact To Patient; No Information
MAUDE report 10355831 ↗InjuryAug 3, 2020Material Separation; Material Separation
THE USER FACILITY REPORTED THAT THE DOCTOR INSERTED THE GLIDEWIRE THROUGH HIS ACCESS NEEDLE AND IT SHEARED OFF ABOUT 30-40CM OF THE JACKET OF THE WIRE. THE DOCTOR HAD TO ACCESS THE OPPOSITE FEMORAL ARTERY TO RETRIEVE THE DETACHED SLEEVE WITH A SNARE. THE PATIENT'S CONDITION WAS STABLE. THERE WAS NO PATIENT INJURY, MEDICAL/SURGICAL INTERVENTION REQUIRED. ADDITIONAL INFORMATION WAS RECEIVED ON 23JUL2020. AN ANGIO PROCEDURE WAS BEING PERFORMED. THEY ARE NOT AWARE THAT ANY OF THE GLIDEWIRE WAS LEFT IN THE PATIENT. FLUOROSCOPY WAS USED TO SNARE THE JACKET. THERE WAS NO HARM TO THE PATIENT. ONCE JACKET PORTION WAS SNARED, THEY FINISHED GETTING THEIR IMAGING.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10355829 ↗InjuryAug 2, 2019Material Separation; Material Separation
THE USER FACILITY REPORTED THAT THE DOCTOR WAS DOING A PERCUTANEOUS NEPHROLITHOTOMY FOR A VERY EXTENSIVE CAST STONE IN THE RIGHT KIDNEY, THERE WAS A GREAT DEAL OF DIFFICULTY IN GETTING GUIDEWIRE ACCESS PAST THE STONES INTO THE URETER (CASE PREVIOUSLY FAILED), HE HAD ACCESS WITH AN 18GA X 15CM COOK TROCAR NEEDLE AND EVENTUALLY ATTEMPTED TO CROSS USING A GR3806 GLIDEWIRE. HE WAS UNABLE TO GET THE WIRE THROUGH THE CALYX TO THE URETER, CONTRAST WAS INJECTED TO HELP WITH NAVIGATION, UPON REMOVING THE WIRE IT WAS NOTICED THAT THE POLYURETHANE JACKET WAS SHEARED OFF LEAVING JUST THE NITINOL CORE WIRE FOR ABOUT THE DISTAL 5CM. HE ATTEMPTED ACCESS WITH ANOTHER GR3806 WIRE AND THIS TIME WAS ABLE TO GET DOWN THE URETER; HOWEVER, WAS UNABLE TO PASS A 4FR KUMPE CATHETER OR DILATOR TO BE ABLE TO INSERT THE SCOPE TO RETRIEVE THE JACKET FRAGMENT AND PERFORM THE LITHOTOMY. THE CASE WAS EVENTUALLY ABORTED WITH THE APPROXIMATELY 5CM PORTION OF THE JACKET STILL IN THE PATIENT. THE DOCTOR AGREED THAT IT WAS VERY LIKELY THE JACKET OF THE WIRE WAS SHEARED OFF AGAINST THE BLUNT METAL CANNULA OF THE COOK TROCAR NEEDLE DUE TO PERHAPS BECOMING ENTRAPPED AGAINST A STONE OR POSSIBLY STICKY FROM THE CONTRAST. T
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8855836 ↗InjuryAug 2, 2019Material Separation; Material Separation
THE USER FACILITY REPORTED THAT THE DOCTOR WAS DOING A PERCUTANEOUS NEPHROLITHOTOMY FOR A VERY EXTENSIVE CAST STONE IN THE RIGHT KIDNEY, THERE WAS A GREAT DEAL OF DIFFICULTY IN GETTING GUIDEWIRE ACCESS PAST THE STONES INTO THE URETER (CASE PREVIOUSLY FAILED), HE HAD ACCESS WITH AN 18GA X 15CM COOK TROCAR NEEDLE AND EVENTUALLY ATTEMPTED TO CROSS USING A GR3806 GLIDEWIRE. HE WAS UNABLE TO GET THE WIRE THROUGH THE CALYX TO THE URETER, CONTRAST WAS INJECTED TO HELP WITH NAVIGATION, UPON REMOVING THE WIRE IT WAS NOTICED THAT THE POLYURETHANE JACKET WAS SHEARED OFF LEAVING JUST THE NITINOL CORE WIRE FOR ABOUT THE DISTAL 5CM. HE ATTEMPTED ACCESS WITH ANOTHER GR3806 WIRE AND THIS TIME WAS ABLE TO GET DOWN THE URETER; HOWEVER, WAS UNABLE TO PASS A 4FR KUMPE CATHETER OR DILATOR TO BE ABLE TO INSERT THE SCOPE TO RETRIEVE THE JACKET FRAGMENT AND PERFORM THE LITHOTOMY. THE CASE WAS EVENTUALLY ABORTED WITH THE APPROXIMATELY 5CM PORTION OF THE JACKET STILL IN THE PATIENT. THE DOCTOR AGREED THAT IT WAS VERY LIKELY THE JACKET OF THE WIRE WAS SHEARED OFF AGAINST THE BLUNT METAL CANNULA OF THE COOK TROCAR NEEDLE DUE TO PERHAPS BECOMING ENTRAPPED AGAINST A STONE OR POSSIBLY STICKY FROM THE CONTRAST. T
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8855834 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page