GLIDEWIRE GS3506Recalls & reported problems
Terumo Corporation · RADIFOCUS GLIDEWIRE · Class 2
All models of GLIDEWIRE (30)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04987350773593), received between May 1, 2024 and Nov 25, 2025 — classified as 8 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Material Fragmentation2
- Material Separation2
- Peeled/Delaminated2
Patient problems
- Foreign Body In Patient6
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryNov 25, 2025Peeled/Delaminated; Peeled/Delaminated
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: DURING A NEPHROURETERECTOMY DRAIN INSERTION, THE COATING ON THE STIFF GLIDEWIRE PEELED OFF INSIDE THE METAL CANNULA STIFFENER FOR THE DRAIN. WE WERE ABLE TO RETRIEVE THE COATING THAT PEELED OFF. THERE WAS NO BLOOD LOSS. THERE WAS NO SERIOUS HARM REPORTED. THE DEVICE DEFECT WAS OBSERVED PRIOR TO USE. THERE WAS NO PROLONGED CARE. ADDITIONAL INFORMATION AS RECEIVED ON 04NOV2025: THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23646447 ↗InjuryNov 25, 2025Peeled/Delaminated; Peeled/Delaminated
TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: DURING A NEPHROURETERECTOMY DRAIN INSERTION, THE COATING ON THE STIFF GLIDEWIRE PEELED OFF INSIDE THE METAL CANNULA STIFFENER FOR THE DRAIN. WE WERE ABLE TO RETRIEVE THE COATING THAT PEELED OFF. THERE WAS NO BLOOD LOSS. THERE WAS NO SERIOUS HARM REPORTED. THE DEVICE DEFECT WAS OBSERVED PRIOR TO USE. THERE WAS NO PROLONGED CARE. ADDITIONAL INFORMATION AS RECEIVED ON 04NOV2025: THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23646446 ↗InjuryJul 23, 2025Material Fragmentation; Material Fragmentation
THE USER FACILITY REPORTED THAT THE PROCEDURE WAS COMPLICATED BY A FRACTURED GUIDEWIRE WITH THE GUIDEWIRE FRAGMENT LODGED IN A SUBINTIMAL LOCATION ALONG THE LEFT ILIAC ARTERY. THE ATTEMPTS TO SNARE AND REMOVE THE FRAGMENT WERE NOT SUCCESSFUL. EVENTUALLY, THEY DECIDED THAT THE SAFEST THING FOR THE PATIENT WAS TO LEAVE THE WIRE FRAGMENT IN PLACE. THE FRAGMENT WOULD HAVE BEEN SECURED BY ITS SUBINTIMAL LOCATION. THE WIRE FRAGMENT WAS FURTHER SECURED WITH THE DEPLOYMENT OF ADJACENT BALLOON-EXPANDABLE STENTS. THE PROCEDURE WAS ARTERIAL ANGIOPLASTY AND ATHERECTOMY. NO BLOOD LOSS WAS REPORTED. THE REPORTED EVENT DID NOT RESULT IN PATIENT INJURY AND/OR REQUIRE MEDICAL OR SURGICAL INTERVENTION. ADDITIONAL INFORMATION WAS RECEIVED ON 14JUL2025: THE SIZE OF FRAGMENT LEFT IN PATIENT WAS 13 CM. WITHIN A SUBINTIMAL TRACT WITH TORTUOSITY AND RESISTANCE WITH TORQUE APPLIED TO THE WIRE USING A TORQUE DEVICE. THE WIRE HAD BEEN ADVANCED THROUGH 5F KUMPE AND OMNI SELECT CATHETERS. THE PATIENT WAS IN STABLE CONDITION. THERE WERE NO PLANS TO REMOVE WIRE FRAGMENTS SINCE THEY PLACED STENTS TO SECURE.
Patient problems: Foreign Body In Patient
MAUDE report 22578816 ↗InjuryJul 23, 2025Material Fragmentation; Material Fragmentation
THE USER FACILITY REPORTED THAT THE PROCEDURE WAS COMPLICATED BY A FRACTURED GUIDEWIRE WITH THE GUIDEWIRE FRAGMENT LODGED IN A SUBINTIMAL LOCATION ALONG THE LEFT ILIAC ARTERY. THE ATTEMPTS TO SNARE AND REMOVE THE FRAGMENT WERE NOT SUCCESSFUL. EVENTUALLY, THEY DECIDED THAT THE SAFEST THING FOR THE PATIENT WAS TO LEAVE THE WIRE FRAGMENT IN PLACE. THE FRAGMENT WOULD HAVE BEEN SECURED BY ITS SUBINTIMAL LOCATION. THE WIRE FRAGMENT WAS FURTHER SECURED WITH THE DEPLOYMENT OF ADJACENT BALLOON-EXPANDABLE STENTS. THE PROCEDURE WAS ARTERIAL ANGIOPLASTY AND ATHERECTOMY. NO BLOOD LOSS WAS REPORTED. THE REPORTED EVENT DID NOT RESULT IN PATIENT INJURY AND/OR REQUIRE MEDICAL OR SURGICAL INTERVENTION. ADDITIONAL INFORMATION WAS RECEIVED ON 14JUL2025: THE SIZE OF FRAGMENT LEFT IN PATIENT WAS 13 CM. WITHIN A SUBINTIMAL TRACT WITH TORTUOSITY AND RESISTANCE WITH TORQUE APPLIED TO THE WIRE USING A TORQUE DEVICE. THE WIRE HAD BEEN ADVANCED THROUGH 5F KUMPE AND OMNI SELECT CATHETERS. THE PATIENT WAS IN STABLE CONDITION. THERE WERE NO PLANS TO REMOVE WIRE FRAGMENTS SINCE THEY PLACED STENTS TO SECURE.
Patient problems: Foreign Body In Patient
MAUDE report 22578595 ↗InjurySep 27, 2024Break; Break
THE USER FACILITY REPORTED THAT DURING A VASCULAR PROCEDURE IN IR, THE TIP OF THE STIFF ANGLED GLIDEWIRE BROKE OFF IN A COLLATERAL VESSEL. THE TIP COULD NOT BE RETRIEVED BEFORE THE END OF THE PROCEDURE. THERE WAS NO BLOOD LOSS REPORTED. THE INJURY INVOLVED THE TIP OF THE GLIDEWIRE BREAKING OFF AND BEING LEFT INSIDE A COLLATERAL ARTERY. NO DIRECT ALLEGATION HAS BEEN MADE. PRE-PROCEDURE CONDITION: THE PATIENT PRESENTED WITH PROLONGED BLEEDING DURING DIALYSIS TREATMENTS. AN ANGIOGRAM CONFIRMED A TOTAL CEPHALIC ARCH OCCLUSION. ADDITIONAL INFORMATION WAS RECEIVED ON 3 SEPT 2024: THERE WAS NO HARM TO THE PATIENT'S VESSEL. THE PHYSICIAN ATTEMPTED TO CROSS A TOTAL OCCLUSION IN THE PATIENT'S CEPHALIC ARCH. ALTHOUGH THE OCCLUSION WAS NOT CROSSED, NO HARM WAS CAUSED. THE TIP WAS LEFT IN THE STENOSIS AS IT IS IN A STABLE POSITION WITH LITTLE RISK OF MIGRATION. THE PATIENT IS STABLE AND EXPERIENCED NO ADVERSE EFFECTS FROM THE EVENT. THE SAMPLE WAS IN THE BODY AND IN CONTACT WITH BLOOD BUT WAS WIPED OFF BEFORE COLLECTION. IT IS NOT CONTAMINATED BY INFECTIOUS AGENTS OR ANTI-CANCER AGENTS.
Patient problems: Foreign Body In Patient
MAUDE report 20317344 ↗InjurySep 26, 2024Break; Break
THE USER FACILITY REPORTED THAT DURING A VASCULAR PROCEDURE IN IR, THE TIP OF THE STIFF ANGLED GLIDEWIRE BROKE OFF IN A COLLATERAL VESSEL. THE TIP COULD NOT BE RETRIEVED BEFORE THE END OF THE PROCEDURE. THERE WAS NO BLOOD LOSS REPORTED. THE INJURY INVOLVED THE TIP OF THE GLIDEWIRE BREAKING OFF AND BEING LEFT INSIDE A COLLATERAL ARTERY. NO DIRECT ALLEGATION HAS BEEN MADE. PRE-PROCEDURE CONDITION: THE PATIENT PRESENTED WITH PROLONGED BLEEDING DURING DIALYSIS TREATMENTS. AN ANGIOGRAM CONFIRMED A TOTAL CEPHALIC ARCH OCCLUSION. ADDITIONAL INFORMATION WAS RECEIVED ON 3 SEPT 2024: THERE WAS NO HARM TO THE PATIENT'S VESSEL. THE PHYSICIAN ATTEMPTED TO CROSS A TOTAL OCCLUSION IN THE PATIENT'S CEPHALIC ARCH. ALTHOUGH THE OCCLUSION WAS NOT CROSSED, NO HARM WAS CAUSED. THE TIP WAS LEFT IN THE STENOSIS AS IT IS IN A STABLE POSITION WITH LITTLE RISK OF MIGRATION. THE PATIENT IS STABLE AND EXPERIENCED NO ADVERSE EFFECTS FROM THE EVENT. THE SAMPLE WAS IN THE BODY AND IN CONTACT WITH BLOOD BUT WAS WIPED OFF BEFORE COLLECTION. IT IS NOT CONTAMINATED BY INFECTIOUS AGENTS OR ANTI-CANCER AGENTS.
Patient problems: Foreign Body In Patient
MAUDE report 20312966 ↗InjuryMay 1, 2024Material Separation; Material Separation
THE USER FACILITY REPORTED THAT WHEN TORQUING THE INVOLVED GUIDEWIRE (GW), THE DISTAL ANGLED PORTION OF THE GW BROKE OFF AND REMAINED IN THE PATIENT. THERE WAS NO BLOOD LOSS. THE REPORTED EVENT RESULTED IN PATIENT INJURY AND/OR REQUIRE MEDICAL OR SURGICAL INTERVENTION. THERE IS NOT A DIRECT ALLEGATION THAT THE REPORTED DEVICE CAUSED OR CONTRIBUTED TO PATIENT INJURY AND/OR NEED FOR MEDICAL INTERVENTION. ADDITIONAL INFORMATION WAS RECEIVED ON 16APR2024: THE PROCEDURE PERFORMED WAS A BILATERAL PELVIC ANGIOGRAM. THERE WAS NO INTERVENTION PERFORMED WHEN THE WIRE BROKE OFF, THE PHYSICIAN WAS ATTEMPTING TO CROSS A LESION. THERE WAS NO MORE RESISTANCE THAN WHAT WAS TO BE EXPECTED. THE BROKEN PIECE WAS NOT REMOVED FROM THE PATIENT, IT WAS IN THE RETROPERITONEAL SOFT TISSUE. THE PIECE REMAINED IN THE PATIENT. THE PATIENT WAS STABLE. THE PROCEDURE WAS NOT COMPLETED AT THAT TIME DUE TO PATIENT TOLERANCE. THE PATIENT NEEDED ANESTHESIA AND THE PROCEDURE WAS COMPLETED A DIFFERENT DAY WITH GENERAL ANESTHESIA. THE DEVICE DID NOT APPEAR TO BE DAMAGED PRIOR TO USE.
Patient problems: Foreign Body In Patient
MAUDE report 19220780 ↗InjuryMay 1, 2024Material Separation; Material Separation
THE USER FACILITY REPORTED THAT WHEN TORQUING THE INVOLVED GUIDEWIRE (GW), THE DISTAL ANGLED PORTION OF THE GW BROKE OFF AND REMAINED IN THE PATIENT. THERE WAS NO BLOOD LOSS. THE REPORTED EVENT RESULTED IN PATIENT INJURY AND/OR REQUIRE MEDICAL OR SURGICAL INTERVENTION. THERE IS NOT A DIRECT ALLEGATION THAT THE REPORTED DEVICE CAUSED OR CONTRIBUTED TO PATIENT INJURY AND/OR NEED FOR MEDICAL INTERVENTION. ADDITIONAL INFORMATION WAS RECEIVED ON 16APR2024: THE PROCEDURE PERFORMED WAS A BILATERAL PELVIC ANGIOGRAM. THERE WAS NO INTERVENTION PERFORMED WHEN THE WIRE BROKE OFF, THE PHYSICIAN WAS ATTEMPTING TO CROSS A LESION. THERE WAS NO MORE RESISTANCE THAN WHAT WAS TO BE EXPECTED. THE BROKEN PIECE WAS NOT REMOVED FROM THE PATIENT, IT WAS IN THE RETROPERITONEAL SOFT TISSUE. THE PIECE REMAINED IN THE PATIENT. THE PATIENT WAS STABLE. THE PROCEDURE WAS NOT COMPLETED AT THAT TIME DUE TO PATIENT TOLERANCE. THE PATIENT NEEDED ANESTHESIA AND THE PROCEDURE WAS COMPLETED A DIFFERENT DAY WITH GENERAL ANESTHESIA. THE DEVICE DID NOT APPEAR TO BE DAMAGED PRIOR TO USE.
Patient problems: Foreign Body In Patient
MAUDE report 19220610 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page