Skip to content

DIAGNOSTIC ULTRASOUND SYSTEM CUS-AA550Recalls & reported problems

Canon Inc. · DIAGNOSTIC ULTRASOUND SYSTEM · Class 2

All models of DIAGNOSTIC ULTRASOUND SYSTEM (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Sep 29, 2022Class IIOpen, Classified

    After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.

    Z-0198-2023 · Root cause: Software Design Change · Recall details · FDA record ↗

Other Transducer, ultrasonic, diagnostic devices

All Transducer, ultrasonic, diagnostic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page