DIAGNOSTIC ULTRASOUND SYSTEM recallZ-0198-2023 · Canon Medical System, USA, INC. · initiated Sep 29, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Canon Medical System, USA, INC. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.
FDA root cause category: Software Design Change
What the company is doing
On 9/29/22, correction notices were mailed to customers who were asked to do the following until the software corrective action had been implemented: Terminate Stress Echo by pressing "X" or "Quit" on the touch panel when completing the exam. If the exam is completed without terminating Stress Echo, and the confirmation message shown in section 2 is displayed when saving the first image in the next Stress Echo, terminate Stress Echo by pressing "X" or "Quit", and then start Stress Echo again. However, if the confirmation message is not displayed when saving the first image in Stress Echo, the function can be used normally. Share the contents of this letter with all users and reviewing radiologist as well as clinical engineering or biomedical group at your facility. Complete and return the response form. For software scheduling questions, please contact your respective Area Service Manager or our InTouch team at [email protected].
Recall details
- Recall number
- Z-0198-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 29, 2022
- Posted by the FDA
- Nov 8, 2022
- Recalling firm
- Canon Medical System, USA, INC.
- Quantity
- 65
- Product code
- IYN
- FDA submissions
- K212333, K212960
- Distribution
- US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL
Product description
Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450
Code information (models, lots, serial numbers)
Model/UDI-DI, with software: Aplio i900 TUS-AI900/ 04987670102523, Aplio i800 TUS-AI800/ 04987670102516, Aplio i700 TUS-AI700/ 04987670102509, With System Version:V4.6/V5.0/V5.1, SP0005 to SP1014, or V6.0/V6.5 SP0000 to SP1009; Aplio a550 CUS-AA550/ 04987670103353, Aplio a450 CUS-AA450/ 04987670103346, With System Version: V3.0, SP0005 to SP1014, or V4.0/V4.5, SP0000 to SP1009
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- DIAGNOSTIC ULTRASOUND SYSTEM TUS-AI700Canon Inc. · System, imaging, pulsed echo, ultrasonic1 recall1 report
- DIAGNOSTIC ULTRASOUND SYSTEM TUS-AI800Canon Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- DIAGNOSTIC ULTRASOUND SYSTEM TUS-AI900Canon Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- DIAGNOSTIC ULTRASOUND SYSTEM CUS-AA450Canon Inc. · Transducer, ultrasonic, diagnostic1 recall0 reports
- DIAGNOSTIC ULTRASOUND SYSTEM CUS-AA550Canon Inc. · Transducer, ultrasonic, diagnostic1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.