Skip to content

Thoraflex™ Hybrid THP2628X150ARecalls & reported problems

Vascutek Ltd · Thoraflex Hybrid Plexus 4 Hybrid Stent Device · Class 3 · Implant

All models of Thoraflex Hybrid (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05037881026800), received between May 5, 2025 and Jan 26, 2026 — classified as 1 death, 1 injury.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • Insufficient Device Problem Information1
  • Partial Blockage1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Neuropathy1
  • Stroke/CVA1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • DeathJan 26, 2026Partial Blockage; Fluid/Blood Leak

    EVENT REPORTED FROM CLINICAL STUDY AS ACUTE MESENTERIC ISCHEMIA (AMI) CLINICAL TRIAL (B)(4), ON POST OPERATIVE DAY 12 ((B)(6) 2025), PATIENT (B)(6) WAS NOTED TO HAVE INCREASED ABDOMINAL DISTENSION, COLONIC DILATATION, AND INCREASED PRESSOR REQUIREMENTS. THE PATIENT HAD A COMPUTED TOMOGRAPHY (CT) SCAN ON (B)(6) 2025 FOR CONCERN OF BOWEL ISCHEMIA. CRITICAL FINDINGS OF RETROPERITONEAL HEMATOMA AND SMA (SUPERIOR MESENTERIC ARTERY) OCCLUSION. ON (B)(6) 2025 THE PATIENT HAD AN EXPLORATORY LAPAROTOMY; VASCULAR SURGERY WAS CONSULTED FOR SMA EMBOLECTOMY. DURING THE OPERATION CHRONIC APPEARING THROMBUS WAS RETRIEVED FROM THE DISTAL SMA WITH A SMALL AMOUNT OF THROMBUS EMANATING INTO ONE OF THE BRANCHES. A THROMBECTOMY WAS PERFORMED ON EACH OF THE 5 PRIMARY BRANCHES THAT HAD BEEN ISOLATED. A SMALL AMOUNT OF THROMBUS WAS RETRIEVED FROM ONLY ONE OF THE BRANCHES, THE REMAINDER OF THE BRANCHES HAD ALL BACK BLEEDING AND NO THROMBUS. THE INVESTIGATOR CONSIDERED THE EVENT OF ACUTE MESENTERIC ISCHEMIA AS SERIOUS, WITH SEVERE SEVERITY AND NOT RELATED TO THE DEVICE, THERE WAS NO DEVICE DEFICIENCY DURING OR AFTER THE PROCEDURE. HOWEVER, THE CLINICAL EVENTS COMMITTEE (CEC) ON ADJUDICATION HAVE DOCUMENTED

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Thrombosis/Thrombus

    MAUDE report 24176122 ↗
  • InjuryMay 5, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    EVENT REPORTED FROM EXTEND STUDY - (B)(6) (SITE (B)(4) FOR PATIENT (B)(6) EVENT WAS INITIALLY COMMUNICATED WITH NO DEVICE DETAILS ON 03 APR 25, DEVICE DETAILS PROVIDED ON 07 APR 25. EVENT OF RIGHT UPPER EXTREMITY AND RIGHT LOWER EXTREMITY WEAKNESS EVENT DATE (B)(6) 2025 FOLLOWED BY LEFT ARM WEAKNESS EVENT DATE (B)(6) 2025 AND ALSO EVENT OF MODERATE ACUTE MICROVASCULAR ISCHEMIC INFARCTS EVENT DATE (B)(6) 2025. THERE WAS NO DEVICE FAILURE/DEFICIENCY REPORTED. THIS REPORT IS BEING SUBMITTED AS FOLLOW UP # 2 FOR MANUFACTURING REPORT NUMBER 9612515-2025-00041 TO PROVIDE EVENT CLOSURE INFORMATION FOR (B)(4).

    Patient problems: Neuropathy; Stroke/CVA

    MAUDE report 21959729 ↗

Other Hybrid stent graft, thoracic aortic lesion treatment devices

All Hybrid stent graft, thoracic aortic lesion treatment devices →

Sources and updates