Skip to content

Thoraflex™ Hybrid THA3034X100ARecalls & reported problems

Vascutek Ltd · Thoraflex Hybrid Ante-Flo Hybrid Stent Device · Class 3 · Implant

All models of Thoraflex Hybrid (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05037881026947), received between Feb 20, 2024 and Jun 24, 2025 — classified as 1 death, 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Insufficient Device Problem Information1

Patient problems

  • Thrombosis/Thrombus2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    EVENT WAS REPORTED BY CLINICAL STUDY: EXTEND - (B)(4) ON POST OPERATIVE DAY 30 (B)(6) 2024), A ROUTINE FOLLOW UP CTA C/A/P (THE AUTHOR DEEMS THIS TO BE COMPUTED TOMOGRAPHY ANGIOGRAPHY CHEST/ABDOMEN/PELVIS) REVEALS THERE WAS A FILLING OF THE FALSE LUMEN VIA RE-ENTRY TEARS IN THE DISTAL DESCENDING THORACIC AORTA SHOWING A PARTIAL THROMBOSIS OF FALSE LUMEN IN THE TAA. TREATMENT OF THE EVENT INCLUDED COMPLETION ENDOVASCULAR STENTING OF THE DISTAL DESCENDING THORACIC EVENT. PATIENT UNDERWENT ON (B)(6) 2025, AN ENDOVASCULAR REPAIR OF CHRONIC RESIDUAL DESCENDING DISSECTION DUE TO CONTINUED FILLING OF FALSE LUMEN VIA DISTAL RE-ENTRY TEARS USING A RELAYPRO NBS STENT GRAFT - CATALOGUE NUMBER 28- N4-34-154-3OU, LOT NUMBER: 2405300500, UDI NUMBER: (B)(4). THE PATIENT RECOVERED WITH TREATMENT. THE EVENT WAS CONSIDERED RESOLVED ON 26 MAR 25. THE PATIENT CONTINUED IN THE STUDY. CONCOMITANT MEDICATIONS (TAKEN AT THE TIME OF THE EVENT) INCLUDED: ENALAPRIL (ONGOING) METOPROLOL (ONGOING), WARFARIN WAS STARTED ON (B)(6) 2024 AND IS ONGOING.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 22317726 ↗
  • DeathFeb 20, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    AS REPORTED EXTEND STUDY: AUTOMATED E-MAIL NOTIFICATION RECEIVED FROM STUDY DATABASE ON 23 JAN 24 REPORTING THE EVENT AORTIC THROMBUS FROM (B)(6) 2023. NO ADDITIONAL DESCRIPTION WAS PROVIDED. THE SITE RECORDED THAT THIS WAS RELATED TO THE THORAFLEX HYBRID DEVICE AND EMAIL CORRESPONDENCE INDICATES THE DEVICE ITSELF HAD NO DEFICIENCIES, AND THERE WAS NO MISPLACEMENT OR MALDEPLOYMENT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Thrombosis/Thrombus

    MAUDE report 18736770 ↗

Other Hybrid stent graft, thoracic aortic lesion treatment devices

All Hybrid stent graft, thoracic aortic lesion treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page