HEALON DUET PRO 10240016Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · HEALON DUET PRO US · Class 3

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 05050474651425), received between Oct 8, 2019 and Feb 19, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Contamination” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination4
  • Adverse Event Without Identified Device or Use Problem2
  • Mechanical Problem2
  • Operating System Becomes Nonfunctional2
  • Delivered as Unsterile Product1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Eye Infections2
  • Loss of Vision2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2025Delivered as Unsterile Product; Manufacturing, Packaging or Shipping Problem; Delivered as Unsterile Product; Manufacturing, Packaging or Shipping Problem

    THREE (03) HEALON DUET PRO BOXES IN WHICH THE SYRINGES WERE PUT WERE SIGNIFICANTLY DAMAGED. THE EXTERNAL SHIPPING BOX WAS COMPLETELY INTACT WITH NO SIGNS OF DAMAGE AND THE ACTUAL DUET SYRINGES ARE FINE. IN ADDITION, THEY WERE SHORTED 5 UNITS (35 RECEIVED) ALTHOUGH THE PACKING SLIP STATES 40 WERE SHIPPED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21413230 ↗
  • MalfunctionOct 9, 2024Contamination; Contamination

    IT WAS REPORTED THAT THE HEALON PRO OVD CONTAINED A THREAD-LIKE PARTICLE AFTER BEING INJECTED INTO A PATIENT¿S EYE. THE SURGEON SUCCESSFULLY REMOVED IT USING THE PHACO HANDPIECE AND NO COMPLICATIONS WERE REPORTED. ADDITIONAL INFORMATION SUGGESTS THAT THE FOREIGN PARTICLE MAY HAVE BEEN PRESENT IN THE HEALON OR COULD HAVE RESULTED FROM CONTAMINATION PRIOR TO DELIVERY INTO THE EYE. IT IS POSSIBLE THAT THE HAIR ORIGINATED FROM SOMETHING IN THE STERILE TRAY SETUP. PATIENT IS DOING WELL. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20409304 ↗
  • MalfunctionOct 11, 2022Contamination; Contamination

    IT WAS REPORTED DURING SURGERY THE DOCTOR INJECTED THE HEALON BEFORE INTRAOCULAR LENS (IOL) INSERTION AND NOTICED A SMALL GRAY PARTICLE APPEAR IN THE PATIENT'S OPERATIVE EYE AFTER DISCHARGING THE HEALON. THE ACCOUNT INDICATED IT IS POSSIBLE THAT THE PARTICLE CAME OUT OF THE HEALON SYRINGE. THE GRAY PARTICLE WAS REMOVED DURING IRRIGATION/ASPIRATION (I/A) AND THE SURGERY WAS SUCCESSFUL. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS RECEIVED CONFIRMING THERE WAS NO MEDICAL INTERVENTION AND O SURGICAL INTERVENTION. PATIENT OUTCOME POST-PROCEDURE WAS REPORTED AS STABLE, NO COMPLICATIONS. BOTH THE GRAY PARTICLE AND THE HEALON ARE NOT AVAILABLE FOR RETURN AS BOTH WERE DISCARDED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15578203 ↗
  • MalfunctionMay 23, 2022Mechanical Problem; Operating System Becomes Nonfunctional; Mechanical Problem; Operating System Becomes Nonfunctional

    IT WAS REPORTED THAT THE THREADS AT THE END OF THE SYRINGE OF THE OPHTHALMIC VISCOELASTIC DEVICE (OVD) DID NOT LOCK AND WAS JUST SHOOTING OUT AT THE END OF THE SYRINGE. A BRIEF DESCRIPTION OF THE ISSUE INDICATED THAT THE CANNULA WAS NOT GETTING DISCONNECTED BUT WOULD NOT CONNECT AT ALL. IT WOULD SPIN, THERE WAS SOMETHING WRONG WITH THE THREADS ON THE SYRINGE. THE CANNULA COULD NOT BE TIGHTENED ONTO THE SYRINGE. THERE WAS PATIENT CONTACT; HOWEVER, THERE WAS NO PATIENT INJURY. NO FURTHER INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 FOR THE SECOND DEVICE REPORTED. ANOTHER REPORT IS BEING FILED FOR THE FIRST DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14478982 ↗
  • MalfunctionMay 23, 2022Mechanical Problem; Operating System Becomes Nonfunctional; Mechanical Problem; Operating System Becomes Nonfunctional

    IT WAS REPORTED THAT THE THREADS AT THE END OF THE SYRINGE OF THE OPHTHALMIC VISCOELASTIC DEVICE (OVD) DID NOT LOCK AND WAS JUST SHOOTING OUT AT THE END OF THE SYRINGE. A BRIEF DESCRIPTION OF THE ISSUE INDICATED THAT THE CANNULA WAS NOT GETTING DISCONNECTED BUT WOULD NOT CONNECT AT ALL. IT WOULD SPIN, THERE WAS SOMETHING WRONG WITH THE THREADS ON THE SYRINGE. THE CANNULA COULD NOT BE TIGHTENED ONTO THE SYRINGE. THERE WAS PATIENT CONTACT. HOWEVER; THERE WAS NO PATIENT INJURY. NO FURTHER INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 2 FOR THE FIRST DEVICE REPORTED. ANOTHER REPORT IS BEING FILED FOR THE SECOND DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14478815 ↗
  • MalfunctionApr 15, 2022Contamination; Contamination

    IT IS REPORTED THAT AFTER INJECTION OF THE HEALON5 PRO A SMALL PARTICLE OF PLASTIC WAS OBSERVED IN THE PATIENT'S LEFT EYE. ISSUE WAS NOTIFIED DURING IMPLANTATION OR APPLICATION. IT WAS REMOVED AND THERE WAS NO DELAY IN PROCEDURE REPORTED. NO MEDICAL OR SURGICAL INTERVENTIONS REPORTED. IT WAS NOTED THAT PATIENT'S DAILY ACTIVITIES WERE NOT SIGNIFICANTLY AFFECTED. PATIENT IS FULLY RECOVERED POST-OP. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14123818 ↗
  • InjuryApr 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A SURGERY CENTER REPORTED THE PRODUCT DID NOT BEHAVE AS USUAL (PERFORMANCE WAS NOT SPECIFIED) AND BECAUSE OF THAT THEY SUSPECT THE PRODUCT CAME FROM AN OUTSIDE VENDOR AND HAD BEEN STORED OUTSIDE THROUGHOUT THE NIGHT. IT WAS INDICATED THAT THE PRODUCT LOOKED NORMAL, HOWEVER FOLLOWING CATARACT EXTRACTION PROCEDURES TWO PATIENTS SUFFERED EYE INFECTIONS RESULTING IN BLINDNESS. THE CUSTOMER NOTED THAT THEY ARE NOT BLAMING THE PRODUCT, OR JOHNSON & JOHNSON AND THAT THEY ARE ONLY TRYING TO RULE OUT THE VARIABLES FOR THESE CASES WHERE THE VISCOELASTIC PRODUCT USED MAY HAVE PREVIOUSLY BEEN FROZEN AND BROUGHT BACK TO ROOM TEMPERATURE PRIOR TO USE. THE CUSTOMER STATED HE COULD NOT CONFIRM KNOWLEDGE IF THE ITEM HAD IN FACT BEEN FROZEN BUT DID INDICATE THE SURGEON PERFORMING THE SURGERIES STATED HE FELT THE ITEM WAS NOT PERFORMING AS EXPECTED (PERFORMANCE WAS NOT SPECIFIED). NO PATIENT INFORMATION COULD BE PROVIDED. IT WAS NOTED THAT THE SURGERIES TOOK PLACE IN THE MONTH OF (B)(6). NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Eye Infections; Loss of Vision

    MAUDE report 11611509 ↗
  • InjuryApr 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A SURGERY CENTER REPORTED THE PRODUCT DID NOT BEHAVE AS USUAL (PERFORMANCE WAS NOT SPECIFIED) AND BECAUSE OF THAT THEY SUSPECT THE PRODUCT CAME FROM AN OUTSIDE VENDOR AND HAD BEEN STORED OUTSIDE THROUGHOUT THE NIGHT. IT WAS INDICATED THAT THE PRODUCT LOOKED NORMAL, HOWEVER FOLLOWING CATARACT EXTRACTION PROCEDURES TWO PATIENTS SUFFERED EYE INFECTIONS RESULTING IN BLINDNESS. THE CUSTOMER NOTED THAT THEY ARE NOT BLAMING THE PRODUCT, OR JOHNSON & JOHNSON AND THAT THEY ARE ONLY TRYING TO RULE OUT THE VARIABLES FOR THESE CASES WHERE THE VISCOELASTIC PRODUCT USED MAY HAVE PREVIOUSLY BEEN FROZEN AND BROUGHT BACK TO ROOM TEMPERATURE PRIOR TO USE. THE CUSTOMER STATED HE COULD NOT CONFIRM KNOWLEDGE IF THE ITEM HAD IN FACT BEEN FROZEN BUT DID INDICATE THE SURGEON PERFORMING THE SURGERIES STATED HE FELT THE ITEM WAS NOT PERFORMING AS EXPECTED (PERFORMANCE WAS NOT SPECIFIED). NO PATIENT INFORMATION COULD BE PROVIDED. IT WAS NOTED THAT THE SURGERIES TOOK PLACE IN THE MONTH OF (B)(6). NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Eye Infections; Loss of Vision

    MAUDE report 11611428 ↗
  • MalfunctionOct 8, 2019Contamination; Contamination

    IT WAS REPORTED THAT A SURGEON OBSERVED ''WHITE FIBER'' WHEN INJECTING HEALON DUET PRO OPHTHALMIC VISCOSURGICAL DEVICE (OVD). THE WHITE FIBER WAS REMOVED BY IRRIGATION AND ASPIRATION. THERE WAS NO PATIENT ADVERSE EVENT REPORTED BY THE SURGEON. THROUGH FOLLOW-UP IT WAS CLARIFIED THAT THE DOCTOR DID NOT OBSERVE THE WHITE FIBER INSIDE OF THE OVD. HOWEVER, WHEN THE DOCTOR INJECTED THE VISCO DEVICE, HE OBSERVED THE WHITE FIBER IN THE PATIENT'S EYE. HENCE, THE DOCTOR CONCLUDED THAT THE WHITE FIBER MAY HAVE COME FROM THE HEALON DUET PRO. DOCTOR WAS ABLE TO REMOVE THE WHITE FIBER BY IRRIGATION AND ASPIRATION. THE PATIENT WAS FINE POST-OP, SURGERY WENT WELL. IT WAS NOTED THERE WILL BE NO PRODUCT RETURNED, AND NO DEBRIS AVAILABLE FOR EVALUATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9169893 ↗

Other AID, SURGICAL, VISCOELASTIC devices