HEALON ENDOCOAT PRO 10300012Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · HEALON ENDOCOAT PRO 0,85 ML USA · Class 3
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05050474755093), received between Jun 5, 2026 and Jun 5, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Contamination” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 31, 2026Class IIOpen, Classified
Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.
Z-3064-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Contamination1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2026Contamination
IT WAS REPORTED THAT A PHYSICIAN OBSERVED WHITE DEBRIS BEING RELEASED FROM A NEW HEALON ENDOCOAT PRO DURING USE. THE PHYSICIAN WAS ABLE TO REMOVE THE DEBRIS FROM THE PATIENT¿S EYE. NO FURTHER INFORMATION WAS PROVIDED. THE PHYSICIAN REPORTED HAVING OBSERVED A SIMILAR OCCURRENCE PREVIOUSLY; HOWEVER, THAT EVENT WAS NOT REPORTED AT THE TIME. THIS MEDICAL DEVICE REPORT PERTAINS SPECIFICALLY TO THE INCIDENT REPORTED ON MAY 12, 2026. A SEPARATE REPORT WILL BE SUBMITTED FOR THE PRIOR SIMILAR EVENT EXPERIENCED BY THE PHYSICIAN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25411174 ↗
FDA clearance and classification
The FDA regulates product code LZP (aid, surgical, viscoelastic) under 21 CFR 886.4275 as a Class III device. 21 CFR 886.4275 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the HEALON ENDOCOAT PRO 10300012
Has the HEALON ENDOCOAT PRO 10300012 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3064-2026 (Class II), began on Jul 31, 2026: Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the HEALON ENDOCOAT PRO 10300012?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 5, 2026. That report lists device problem “Contamination” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the HEALON ENDOCOAT PRO 10300012?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the HEALON ENDOCOAT PRO 10300012, and is it still sold?
The FDA's device database lists Johnson & Johnson Surgical Vision, Inc. as the company responsible for it (brand HEALON ENDOCOAT PRO, model 10300012). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other AID, SURGICAL, VISCOELASTIC devices
- HEALON DUET PRO II 10300016Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC1 recall0 reports
- HEALON ENDOCOAT VT585UJohnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC22 reports
- HEALON 5 10290051Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC20 reports
- HEALON GV PRO 10240014Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC15 reports
- Amvisc 59081LBausch & Lomb Incorporated · Aid, Surgical, Viscoelastic10 reports
- DUOVISC 8065183150Alcon Laboratories, Inc. · Aid, Surgical, Viscoelastic9 reports
- HEALON DUET PRO 10240016Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC9 reports
- HEALON5 PRO 10240015Johnson & Johnson Surgical Vision, Inc. · AID, SURGICAL, VISCOELASTIC7 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page