Freelite® Human Kappa Free Kit for use on the Roche cobas® systems LK016.CBRecalls & reported problems

The Binding Site Group Limited · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05051700012096), received between Mar 22, 2024 and Mar 28, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 28, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    A CUSTOMER CONTACTED THE BINDING SITE (PDX) ON THE 08TH OF FEBRUARY 2024 TO REPORT THAT THEIR PDX FREELITE® HUMAN KAPPA FREE KIT FOR USE ON THE ROCHE COBAS® C SYSTEMS ASSAY (PRODUCT CODE: LK016.CB, KIT LOT: 490912) WAS PRODUCING UNUSUALLY HIGH KAPPA RESULTS ON ONE OF THEIR INSTRUMENTS. ON THE 27TH OF FEBRUARY 2024, THE CUSTOMER INFORMED PDX THAT 20 RESULTS RELEASED FROM THE LABORATORY FOR KAPPA FREE WERE FALSELY ELEVATED. 15 OF THESE RESULTS WERE ABOVE THE TOTAL ALLOWABLE ERROR (TEA) OF 40% OF THE ASSAY. THE TESTING TOOK PLACE ON THE 24TH JANUARY 2024, 02ND FEBRUARY 2024 AND 04TH FEBRUARY 2024 TO THE 07TH FEBRUARY 2024 AND THE DISCREPANCY WAS IDENTIFIED BY THE CLINICIAN ON 08TH FEBRUARY 2024. UPON REPEAT TESTING ON AN ADDITIONAL ANALYSER, THE CUSTOMER OBTAINED RESULTS WHICH THEY REPORTED TO BE MORE IN LINE WITH THE CLINICAL PICTURE. THE CUSTOMER REPORTED THAT FOR THE INITIAL INSTRUMENT THE ISSUE WAS RESOLVED FOLLOWING REPLACEMENT OF A SAMPLE PROBE AND A NEW KIT LOT. ALTHOUGH NO REPORTS HAVE BEEN RECEIVED OF HARM TO PATIENTS, CHANGE IN PATIENT MANAGEMENT OR INAPPROPRIATE THERAPY, THE DECISION WAS TAKEN TO REPORT AS A MATTER OF CAUTION. IF THE ERROR WERE TO RECUR FOR 1 OF THE 15 RELE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18993740 ↗
  • MalfunctionMar 22, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    A CUSTOMER CONTACTED THE BINDING SITE (PDX) ON THE 08TH OF FEBRUARY 2024 TO REPORT THAT THEIR PDX FREELITE® HUMAN KAPPA FREE KIT FOR USE ON THE ROCHE COBAS® C SYSTEMS ASSAY (PRODUCT CODE: LK016.CB, KIT LOT: 490912) WAS PRODUCING UNUSUALLY HIGH KAPPA RESULTS ON ONE OF THEIR INSTRUMENTS. ON THE (B)(6) 2024, THE CUSTOMER INFORMED PDX THAT 20 RESULTS RELEASED FROM THE LABORATORY FOR KAPPA FREE WERE FALSELY ELEVATED. 15 OF THESE RESULTS WERE ABOVE THE TOTAL ALLOWABLE ERROR (TEA) OF 40% OF THE ASSAY. THE TESTING TOOK PLACE ON THE (B)(6) 2024, (B)(6) 2024 AND (B)(6) 2024 TO THE (B)(6) 2024 AND THE DISCREPANCY WAS IDENTIFIED BY THE CLINICIAN ON (B)(6) 2024. UPON REPEAT TESTING ON AN ADDITIONAL ANALYSER, THE CUSTOMER OBTAINED RESULTS WHICH THEY REPORTED TO BE MORE IN LINE WITH THE CLINICAL PICTURE. THE CUSTOMER REPORTED THAT FOR THE INITIAL INSTRUMENT THE ISSUE WAS RESOLVED FOLLOWING REPLACEMENT OF A SAMPLE PROBE AND A NEW KIT LOT. ALTHOUGH NO REPORTS HAVE BEEN RECEIVED OF HARM TO PATIENTS, CHANGE IN PATIENT MANAGEMENT OR INAPPROPRIATE THERAPY, THE DECISION WAS TAKEN TO REPORT AS A MATTER OF CAUTION. IF THE ERROR WERE TO RECUR FOR 1 OF THE 15 RELEASED DISCREPANT RESULTS ABOVE THE TEA THERE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18959893 ↗

Other Kappa, Antigen, Antiserum, Control devices