Freelite® Human Kappa Free Kit for use on the Roche cobas® systems LK016.CBRecalls & reported problems
The Binding Site Group Limited · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05051700012096), received between Mar 22, 2024 and Mar 28, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 28, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
A CUSTOMER CONTACTED THE BINDING SITE (PDX) ON THE 08TH OF FEBRUARY 2024 TO REPORT THAT THEIR PDX FREELITE® HUMAN KAPPA FREE KIT FOR USE ON THE ROCHE COBAS® C SYSTEMS ASSAY (PRODUCT CODE: LK016.CB, KIT LOT: 490912) WAS PRODUCING UNUSUALLY HIGH KAPPA RESULTS ON ONE OF THEIR INSTRUMENTS. ON THE 27TH OF FEBRUARY 2024, THE CUSTOMER INFORMED PDX THAT 20 RESULTS RELEASED FROM THE LABORATORY FOR KAPPA FREE WERE FALSELY ELEVATED. 15 OF THESE RESULTS WERE ABOVE THE TOTAL ALLOWABLE ERROR (TEA) OF 40% OF THE ASSAY. THE TESTING TOOK PLACE ON THE 24TH JANUARY 2024, 02ND FEBRUARY 2024 AND 04TH FEBRUARY 2024 TO THE 07TH FEBRUARY 2024 AND THE DISCREPANCY WAS IDENTIFIED BY THE CLINICIAN ON 08TH FEBRUARY 2024. UPON REPEAT TESTING ON AN ADDITIONAL ANALYSER, THE CUSTOMER OBTAINED RESULTS WHICH THEY REPORTED TO BE MORE IN LINE WITH THE CLINICAL PICTURE. THE CUSTOMER REPORTED THAT FOR THE INITIAL INSTRUMENT THE ISSUE WAS RESOLVED FOLLOWING REPLACEMENT OF A SAMPLE PROBE AND A NEW KIT LOT. ALTHOUGH NO REPORTS HAVE BEEN RECEIVED OF HARM TO PATIENTS, CHANGE IN PATIENT MANAGEMENT OR INAPPROPRIATE THERAPY, THE DECISION WAS TAKEN TO REPORT AS A MATTER OF CAUTION. IF THE ERROR WERE TO RECUR FOR 1 OF THE 15 RELE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18993740 ↗MalfunctionMar 22, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
A CUSTOMER CONTACTED THE BINDING SITE (PDX) ON THE 08TH OF FEBRUARY 2024 TO REPORT THAT THEIR PDX FREELITE® HUMAN KAPPA FREE KIT FOR USE ON THE ROCHE COBAS® C SYSTEMS ASSAY (PRODUCT CODE: LK016.CB, KIT LOT: 490912) WAS PRODUCING UNUSUALLY HIGH KAPPA RESULTS ON ONE OF THEIR INSTRUMENTS. ON THE (B)(6) 2024, THE CUSTOMER INFORMED PDX THAT 20 RESULTS RELEASED FROM THE LABORATORY FOR KAPPA FREE WERE FALSELY ELEVATED. 15 OF THESE RESULTS WERE ABOVE THE TOTAL ALLOWABLE ERROR (TEA) OF 40% OF THE ASSAY. THE TESTING TOOK PLACE ON THE (B)(6) 2024, (B)(6) 2024 AND (B)(6) 2024 TO THE (B)(6) 2024 AND THE DISCREPANCY WAS IDENTIFIED BY THE CLINICIAN ON (B)(6) 2024. UPON REPEAT TESTING ON AN ADDITIONAL ANALYSER, THE CUSTOMER OBTAINED RESULTS WHICH THEY REPORTED TO BE MORE IN LINE WITH THE CLINICAL PICTURE. THE CUSTOMER REPORTED THAT FOR THE INITIAL INSTRUMENT THE ISSUE WAS RESOLVED FOLLOWING REPLACEMENT OF A SAMPLE PROBE AND A NEW KIT LOT. ALTHOUGH NO REPORTS HAVE BEEN RECEIVED OF HARM TO PATIENTS, CHANGE IN PATIENT MANAGEMENT OR INAPPROPRIATE THERAPY, THE DECISION WAS TAKEN TO REPORT AS A MATTER OF CAUTION. IF THE ERROR WERE TO RECUR FOR 1 OF THE 15 RELEASED DISCREPANT RESULTS ABOVE THE TEA THERE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18959893 ↗
Other Kappa, Antigen, Antiserum, Control devices
- Optilite Freelite Mx Kappa Free Kit LK016.M.OPT.AThe Binding Site Group Limited · Kappa, Antigen, Antiserum, Control1 recall0 reports
- Optilite Freelite Kappa Kit LK016.OPT.AThe Binding Site Group Limited · Kappa, Antigen, Antiserum, Control4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page