Optilite Freelite Kappa Kit LK016.OPT.ARecalls & reported problems
The Binding Site Group Limited · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05051700017794), received between Nov 18, 2025 and Mar 11, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K150658, K231290); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Low Test Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 11, 2026High Test Results
THE CUSTOMER RELEASED A DISCREPANT, FALSELY ELEVATED KAPPA FREE LIGHT CHAIN (FLC) RESULT (1262.39 MG/L; REF RANGE: 3.30-19.40 MG/L) OBTAINED USING THE FREELITE® ASSAY. THE PATIENT HAD A PRIOR DIAGNOSIS OF AL AMYLOIDOSIS AND WAS UNDER ROUTINE MONITORING. THE ELEVATED RESULT (TAKEN (B)(6) 2026) WAS INCONSISTENT WITH THE PATIENT'S CLINICAL HISTORY AND FOLLOW UP TESTING. A NEW SAMPLE, TAKEN ON (B)(6) 2026, GAVE A KAPPA RESULT OF 11.47 MG/L, CONSISTENT WITH HISTORICAL PATIENT RESULTS. THE CUSTOMER CONFIRMED THAT PATIENT TREATMENT WAS NOT AFFECTED. THE INVESTIGATION DETERMINED THAT QUALITY CONTROL WERE WITHIN RANGE, AND NO ANALYSER OR ASSAY MALFUNCTION WAS IDENTIFIED. ALTHOUGH THERE HAVE BEEN NO REPORTS OF SERIOUS HARM OR INAPPROPRIATE THERAPY, THE DECISION WAS TAKEN TO REPORT AS A MATTER OF CAUTION. IF THE ERROR WERE TO RECUR, THERE IS A POSSIBILITY OF SERIOUS INJURY IF SAMPLE RESULTS WERE CONSIDERED IN ISOLATION. AS PER THE IFU (INS016.OPT.A), RESULTS OF THE FREE LIGHT CHAIN MEASUREMENTS SHOULD ALWAYS BE INTERPRETED IN CONJUNCTION WITH OTHER LABORATORY AND CLINICAL FINDINGS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24574421 ↗MalfunctionMar 10, 2026High Test Results
THE CUSTOMER RELEASED A DISCREPANT, FALSELY ELEVATED KAPPA FREE LIGHT CHAIN (FLC) RESULT (1262.39 MG/L; REF RANGE: 3.30-19.40 MG/L) OBTAINED USING THE FREELITE® ASSAY. THE PATIENT HAD A PRIOR DIAGNOSIS OF AL AMYLOIDOSIS AND WAS UNDER ROUTINE MONITORING. THE ELEVATED RESULT (TAKEN (B)(6) 2026) WAS INCONSISTENT WITH THE PATIENT'S CLINICAL HISTORY AND FOLLOW UP TESTING. A NEW SAMPLE, TAKEN ON (B)(6) 2026, GAVE A KAPPA RESULT OF 11.47 MG/L, CONSISTENT WITH HISTORICAL PATIENT RESULTS. THE CUSTOMER CONFIRMED THAT PATIENT TREATMENT WAS NOT AFFECTED. THE INVESTIGATION DETERMINED THAT QUALITY CONTROL WERE WITHIN RANGE, AND NO ANALYSER OR ASSAY MALFUNCTION WAS IDENTIFIED. ALTHOUGH THERE HAVE BEEN NO REPORTS OF SERIOUS HARM OR INAPPROPRIATE THERAPY, THE DECISION WAS TAKEN TO REPORT AS A MATTER OF CAUTION. IF THE ERROR WERE TO RECUR, THERE IS A POSSIBILITY OF SERIOUS INJURY IF SAMPLE RESULTS WERE CONSIDERED IN ISOLATION. AS PER THE IFU (INS016.OPT.A), RESULTS OF THE FREE LIGHT CHAIN MEASUREMENTS SHOULD ALWAYS BE INTERPRETED IN CONJUNCTION WITH OTHER LABORATORY AND CLINICAL FINDINGS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24556210 ↗InjuryNov 19, 2025Low Test Results; Low Test Results
A CUSTOMER REPORTED TO THE BINDING SITE AN ERRONEOUSLY LOW KAPPA FREE LIGHT CHAIN (FLC) RESULT (28.3 MG/L; REF RANGE: 3.30 - 19.40 MG/L) OBTAINED USING THE OPTILITE® FREELITE® KAPPA FREE KIT (PRODUCT CODE: LK016.OPT.A), ON (B)(6) 2025. THE RESULT WAS INCONSISTENT WITH EARLIER RESULTS (839.6 & 789 MG/L) FROM SAMPLE DATED (B)(6) 2025 (CONFIRMED BY ERROR REPORT ANALYSIS) AND A STRONGER MONOCLONAL PROTEIN BAND BY SERUM ELECTROPHORESIS. LAMBDA FLC LEVELS WERE CONSISTENTLY LOW AND THE KAPPA/ LAMBDA RATIO WAS ABNORMAL THROUGHOUT. THERE WAS NO ALLEGATION OF SERIOUS INJURY, ALTHOUGH TREATMENT WAS TEMPORARILY PAUSED WHILE THE DISCREPANCY WAS INVESTIGATED. QUALITY CONTROLS (QC) WERE WITHIN RANGE AND ACCEPTABLE THROUGHOUT TESTING. A SUBSEQUENT SAMPLE FROM THE (B)(6) SHOWED AN INCREASE IN KAPPA FLC LEVELS (6426 MG/L). ALTHOUGH THERE HAVE BEEN NO REPORTS OF SERIOUS HARM OR INAPPROPRIATE THERAPY, THE DECISION WAS TAKEN TO REPORT AS A MATTER OF CAUTION. IF THE ERROR WAS TO RECUR THERE IS A POSSIBILITY OF SERIOUS INJURY IF SAMPLE RESULTS WERE CONSIDERED IN ISOLATION. AS PER THE IFU, INS016.OPT.A, RESULTS OF THE FREE LIGHT CHAIN MEASUREMENTS SHOULD ALWAYS BE INTERPRETED IN CONJUNCTION WITH OTHER LAB
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23596224 ↗MalfunctionNov 18, 2025Low Test Results; Low Test Results
A CUSTOMER REPORTED TO THE BINDING SITE AN ERRONEOUSLY LOW KAPPA FREE LIGHT CHAIN (FLC) RESULT (28.3 MG/L; REF RANGE: 3.30 - 19.40 MG/L) OBTAINED USING THE OPTILITE® FREELITE® KAPPA FREE KIT (PRODUCT CODE: LK016.OPT.A), ON 22 OCTOBER 2025. THE RESULT WAS INCONSISTENT WITH EARLIER RESULTS (839.6 & 789 MG/L) FROM SAMPLE DATED 17 OCTOBER 2025 (CONFIRMED BY ERROR REPORT ANALYSIS) AND A STRONGER MONOCLONAL PROTEIN BAND BY SERUM ELECTROPHORESIS. LAMBDA FLC LEVELS WERE CONSISTENTLY LOW AND THE KAPPA/ LAMBDA RATIO WAS ABNORMAL THROUGHOUT. THERE WAS NO ALLEGATION OF SERIOUS INJURY, ALTHOUGH TREATMENT WAS TEMPORARILY PAUSED WHILE THE DISCREPANCY WAS INVESTIGATED. QUALITY CONTROLS (QC) WERE WITHIN RANGE AND ACCEPTABLE THROUGHOUT TESTING. A SUBSEQUENT SAMPLE FROM THE 10TH NOVEMBER SHOWED AN INCREASE IN KAPPA FLC LEVELS (6426 MG/L). ALTHOUGH THERE HAVE BEEN NO REPORTS OF SERIOUS HARM OR INAPPROPRIATE THERAPY, THE DECISION WAS TAKEN TO REPORT AS A MATTER OF CAUTION. IF THE ERROR WAS TO RECUR THERE IS A POSSIBILITY OF SERIOUS INJURY IF SAMPLE RESULTS WERE CONSIDERED IN ISOLATION. AS PER THE IFU, INS016.OPT.A, RESULTS OF THE FREE LIGHT CHAIN MEASUREMENTS SHOULD ALWAYS BE INTERPRETED IN CONJUNCTION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23582756 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K150658, K231290. One submission often covers many models, so check the product and lot details in each record.
- Z-1043-2017 · Jul 5, 2016 · TerminatedOptilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.
Other Kappa, Antigen, Antiserum, Control devices
- Optilite Freelite Mx Kappa Free Kit LK016.M.OPT.AThe Binding Site Group Limited · Kappa, Antigen, Antiserum, Control1 recall0 reports
- Freelite® Human Kappa Free Kit for use on the Roche cobas® systems LK016.CBThe Binding Site Group Limited · Kappa, Antigen, Antiserum, Control2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page