Skip to content

SP CAL (LIQ) IT2691Recalls & reported problems

Randox Laboratories Limited · Liquid Protein Calibrators are an in vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4, Ceruloplasmin, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, RF and Transferrin assays. · Class 2

All models of SP CAL (LIQ) (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (K061056); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Mar 1, 2024Open, Classified

    The firm has realigned C3 and Haptoglobin in Liquid Protein Calibrators, IT2691, to reference material ERM-DA470k/IFCC.

    Z-1328-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jul 19, 2023Open, Classified

    As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.

    Z-2448-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K061056. One submission often covers many models, so check the product and lot details in each record.

  • Z-0283-2020 · Aug 9, 2019 · Terminated
    Liquid Protein Calibrators (SP CAL (LIQ)) Catalogue Number: IT2691 GTIN: 05055273204032 Product Usage: Liquid Protein Calibrators are intended for in vitro diagnostic use in the ca

Other Calibrator, Multi-Analyte Mixture devices

All Calibrator, Multi-Analyte Mixture devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page