CAL 3 CAL2351Recalls & reported problems

Randox Laboratories Limited · The device is intended for use as a calibrator in clinical chemistry assays. Randox calibration sera are based on lyophilized human serum. The concentrations and activities are suitable for calibration of clinical chemistry assays on a wide range of automatic analyzers. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
6

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

6 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Mar 5, 2024Class IIOpen, Classified

    Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.

    Z-1759-2024 · Root cause: Process design · Recall details · FDA record ↗

  2. Nov 20, 2023Class IIOpen, Classified

    CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.

    Z-0752-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Nov 20, 2023Class IIOpen, Classified

    Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples

    Z-0753-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Nov 13, 2023Class IIOpen, Classified

    There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human Assayed Multi-Sera Level 2 (HN1530) for TBIC taget value and on IFU of Human Assayed Multi-Sera Level 3 (HE 1532) for Lipase target value.

    Z-0751-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. Mar 29, 2021Class IITerminated

    Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been attributed to bilirubin instability in the calibrator and is evident in both Quality Control (QC) and Patient samples.

    Z-1563-2021 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  6. Jul 12, 2019Class IIITerminated

    Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).

    Z-2490-2019 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K053153 on Jan 13, 2006 (decision: Substantially Equivalent). Submitted by Randox Laboratories, Ltd. as “Randox Calibration Serum”. The FDA review took 64 days. FDA record ↗

The FDA regulates product code JIX (calibrator, multi-analyte mixture) under 21 CFR 862.1150 as a Class II device. 21 CFR 862.1150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CAL 3 CAL2351

Has the CAL 3 CAL2351 been recalled?

Yes. 6 FDA recall records name this device's identifier: 4 still open and 2 closed (terminated) by the FDA. The most recent, Z-1759-2024 (Class II), began on Mar 5, 2024: Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the CAL 3 CAL2351 FDA approved?

It was cleared through 510(k) K053153 on Jan 13, 2006. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the CAL 3 CAL2351, and is it still sold?

The FDA's device database lists Randox Laboratories Limited as the company responsible for it (brand CAL 3, model CAL2351). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Calibrator, Multi-Analyte Mixture devices

All Calibrator, Multi-Analyte Mixture devices →