Enterprise 9611AF101BCUSBRecalls & reported problems
Arjohuntleigh AB · Medical Bed - 9611AF101BCUSB- USA · Class 2
All models of Enterprise (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This ArjoHuntleigh Enterprise bed is an adjustable hospital bed. The 32 reports mostly describe side rails breaking loose or failing to latch, beds raising or lowering on their own, and backrests collapsing. There are no deaths and 5 injuries, including falls and head injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 05055982755030), received between Aug 20, 2019 and Mar 3, 2026 — classified as 5 injuries, 27 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement15
- Detachment of Device or Device Component11
- Device Fell2
- Use of Device Problem2
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Consequences Or Impact To Patient5
- Fall3
- Head Injury2
- No Patient Involvement2
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 3, 2026Detachment of Device or Device Component
ARJO WAS INFORMED ABOUT AN EVENT INVOLVING THE ENTERPRISE 9000X BED. THE CUSTOMER REPORTED THAT THE RIGHT HEAD SIDE RAIL WAS DAMAGED. THERE WAS NO PATIENT INVOLVED. NO INJURY WAS SUSTAINED. BASED ON THE INSPECTION FINDINGS AND THE PHOTOGRAPHIC EVIDENCE PROVIDED, THE CUSTOMER-REPORTED ISSUE WAS CONFIRMED. THE UPPER ARM OF THE SIDE RAIL WAS FOUND DETACHED, AND THE LOWER ARM WAS DISCONNECTED, RESULTING IN A PARTIAL DETACHMENT OF THE SIDE RAIL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24494665 ↗MalfunctionJan 16, 2026Fracture
THE CUSTOMER REPORTED THAT THE BACKREST FUNCTION ON THE ENTERPRISE 9000X BED WAS INOPERATIVE. NO INJURY WAS SUSTAINED. DURING THE EVALUATION, AN ARJO TECHNICIAN FOUND THAT THE BACKREST ACTUATOR AND DAMPENER HAD FAILED, AND THE BACKREST INNER HINGE WAS BROKEN. THE IDENTIFIED DAMAGES INDICATE THAT THE BACKREST LIKELY COLLAPSED WHILE THE BED WAS BEING USED BY A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24091199 ↗MalfunctionAug 26, 2025Device Fell; Device Fell
ARJO WAS INFORMED ABOUT AN EVENT INVOLVING THE ENTERPRISE 9000X BED, WHERE A PATIENT REPORTEDLY FELL FROM THE BED. ACCORDING TO THE CUSTOMER, THE SIDE RAIL DID NOT PREVENT THE PATIENT FROM EXITING THE BED. NO INJURY WAS SUSTAINED. DURING THE DEVICE INSPECTION, THE ARJO TECHNICIAN FOUND THAT THE SIDE RAIL LATCHING MECHANISM WAS NOT FULLY ENGAGING DUE TO INTERNAL DEBRIS OR AGED GREASE. LUBRICATION WAS APPLIED, AND NO SIDE RAIL PARTS REQUIRED REPLACEMENT. AFTER LUBRICATION, THE SIDE RAIL LOCKING MECHANISM OPERATED AS INTENDED.
Patient problems: Fall
MAUDE report 22892407 ↗MalfunctionAug 20, 2025Unintended Movement; Unintended Movement
DURING THE ON-SITE VISIT, THE ARJO TECHNICIAN NOTICED A NOTE LEFT ON THE ENTERPRISE 9000X BED SAYING THAT THE BED WAS GOING UP AND DOWN ON ITS OWN. THE BED WAS LEFT IN THE STORAGE ROOM. THERE WAS NO INDICATION OF PATIENT INVOLVEMENT, AND NO INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22848284 ↗MalfunctionJul 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THAT THE SIDE RAIL WAS BROKEN AND DID NOT LOCK. THERE WAS A PATIENT ON THE BED WHEN THE EVENT OCCURRED. NO INJURY WAS SUSTAINED. DURING THE DEVICE INSPECTION, IT WAS FOUND THAT THE RIGHT BODY SIDE RAIL WAS SEPARATING FROM THE INSIDE COVER. IT WAS CONFIRMED THAT TWO SCREWS HAD BEEN PULLED OUT AND STUCK BETWEEN THE COVER PIECE AND THE BODY OF THE SIDE RAIL. THE SIDE RAIL WAS REPLACED. THE BED WAS TESTED AND WAS FULLY OPERATIONAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22392013 ↗MalfunctionDec 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ARJO WAS INFORMED ABOUT BROKEN SIDE RAIL FROM ENTERPRISE 9000X BED. THE DEVICE WAS INSPECTED BY AN ARJO SERVICE TECHNICIAN. IT WAS CLARIFIED THAT SIDE RAIL LOWER ARM WAS RIPPED OFF FROM THE BED FRAME. IT WAS SUGGESTED BY THE ARJO SERVICE TECHNICIAN THAT THE SIDE RAIL WAS DAMAGED BY A MENTALLY ILL PATIENT. ARJO RECEIVED IN TOTAL 3 COMPLAINTS FROM THE SAME CUSTOMER WITH THE SAME MALFUNCTION (B)(6); 3007420694-2024-00314, (B)(6); 3007420694-2024-00315 AND (B)(6); 3007420694-2024-00321). THIS REPORT REFERS TO (B)(6); (3007420694-2024-00321).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20901128 ↗MalfunctionDec 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ARJO WAS INFORMED ABOUT BROKEN SIDE RAIL FROM ENTERPRISE 9000X BED. THE DEVICE WAS INSPECTED BY AN ARJO SERVICE TECHNICIAN. IT WAS CLARIFIED THAT SIDE RAIL LOWER ARM WAS RIPPED OFF FROM THE BED FRAME. IT WAS SUGGESTED BY THE ARJO SERVICE TECHNICIAN THAT THE SIDE RAIL WAS DAMAGED BY A MENTALLY ILL PATIENT. ARJO RECEIVED IN TOTAL 3 COMPLAINTS FROM THE SAME CUSTOMER WITH THE SAME MALFUNCTION (TW (B)(4) - 3007420694-2024-00314, TW (B)(4) AND TW (B)(4) - 3007420694-2024-00321). THIS REPORT REFERS TO TW (B)(4) .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20901094 ↗MalfunctionDec 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ARJO WAS INFORMED ABOUT BROKEN SIDE RAIL FROM ENTERPRISE 9000X BED. THE DEVICE WAS INSPECTED BY AN ARJO SERVICE TECHNICIAN. IT WAS CLARIFIED THAT SIDE RAIL LOWER ARM WAS RIPPED OFF FROM THE BED FRAME. IT WAS SUGGESTED BY THE ARJO SERVICE TECHNICIAN THAT THE SIDE RAIL WAS DAMAGED BY A MENTALLY ILL PATIENT. ARJO RECEIVED IN TOTAL 3 COMPLAINTS FROM THE SAME CUSTOMER WITH THE SAME MALFUNCTION (TW(B)(4) , TW (B)(4) - 3007420694-2024-00315 AND TW (B)(4) - 3007420694-2024-00321). THIS REPORT REFERS TO TW (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20900969 ↗MalfunctionNov 8, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
DURING THE SERVICE REPAIR, THE ARJO TECHNICIAN FOUND THAT THE HINGES WERE BROKEN AND THE BACKREST ACTUATOR WAS DETACHED FROM THE METAL FRAME. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20639311 ↗MalfunctionSep 17, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
FOLLOWING THE INFORMATION RECEIVED THE HEAD SIDE RAIL WAS PARTIALLY DETACHED FROM THE ENTERPRISE 9000X BED FRAME (TWO OF FOUR SCREWS BECAME DETACHED). IT WAS INDICATED THAT THE USER PULLED STRONGLY ON THE SIDE RAIL TO MOVE THE BED. THERE WAS NO PATIENT IN THE BED AT THE TIME OF THE EVENT. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20235050 ↗MalfunctionJun 19, 2024Device Fell; Device Fell
ARJO WAS NOTIFIED OF AN EVENT INVOLVING AN ENTERPRISE 9000X BED. IT WAS REPORTED THAT A PATIENT LEANED UP AGAINST THE BED SIDERAIL AND THAT THE LATCHING MECHANISM FAILED. THE PATIENT FELL OUT OF BED ONTO THE FLOOR. NO INJURIES WERE REPORTED.
Patient problems: Fall; No Clinical Signs, Symptoms or Conditions
MAUDE report 19565544 ↗MalfunctionJan 24, 2024Unintended Movement; Unintended Movement
ARJO WAS NOTIFIED OF AN EVENT INVOLVING THE ENTERPRISE 9000X BED. IT WAS INDICATED THAT THE HEADBOARD OF THE BED MOVED UNEXPECTEDLY (IT RAISED WITHOUT TOUCHING ANY BUTTONS). NO PATIENT WAS INVOLVED IN THE EVENT AND NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18569125 ↗MalfunctionNov 9, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ARJO RECEIVED A CUSTOMER COMPLAINT FOR ENTERPRISE 9000X BED. FOLLOWING THE INFORMATION PROVIDED ENTERPRISE 9000X BED WAS FOUND WITH DAMAGED SIDE RAIL. BASED ON THE INFORMATION RECEIVED FROM AN ARJO SERVICE TECHNICIAN, THE SIDE RAIL DETACHED PARTIALLY FROM THE BED FRAME DUE TO A MISSING PIVOT PIN. NO PATIENT INVOLVEMENT WAS REPORTED. NO INJURIES OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18099614 ↗MalfunctionJun 30, 2023Unintended Movement; Unintended Movement
FOLLOWING THE INFORMATION PROVIDED, THERE WAS AN INDICATION OF AN UNINTENDED MOVEMENT ON THE ENTERPRISE 9000X BED. IT WAS REPORTED THAT SIDE RAIL CONTROLS WERE NOT WORKING HOWEVER, THE HEAD SECTION OF THE BED RAISED ON ITS OWN AND AN ERROR CODE E300 WAS DISPLAYED ON THE DEVICE. THERE WAS NO ALLEGATION OF PATIENT INVOLVEMENT. NO INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17239219 ↗MalfunctionJun 20, 2023Unintended Movement; Unintended Movement
FOLLOWING THE INFORMATION PROVIDED, THERE WAS AN INDICATION OF AN UNINTENDED MOVEMENT ON THE ENTERPRISE 9000X BED. IT WAS REPORTED THAT THE BED ¿KEEPS ELEVATING UNCONTROLLABLY AND STOPS WORKING¿ AND THAT THE CODES E410 AND E300 WERE DISPLAYED ON THE BED. THERE WAS NO INDICATION OF A PATIENT INVOLVEMENT. NO INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17161351 ↗MalfunctionMay 24, 2023Unintended Movement; Unintended Movement
ARJO WAS INFORMED ABOUT THE EVENT INVOLVING THE ENTERPRISE 9000X BED. IT WAS INDICATED THAT THE BED MOVED UNINTENDED. IT IS NOT KNOWN IF THE PATIENT WAS ON THE BED AT THE TIME OF THE EVENT. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16990323 ↗MalfunctionFeb 8, 2023Unintended Movement; Unintended Movement
ARJO WAS INFORMED ABOUT THE INCIDENT INVOLVING THE ENTERPRISE 9000X BED. THE CUSTOMER REQUESTED ARJO FOR SERVICE REGARDING THE FAULTY ENTERPRISE 9000X BED AND DURING THE DEVICE INSPECTION PERFORMED BY AN ARJO SERVICE TECHNICIAN IT WAS FOUND THAT THE UPPER RIGHT BUTTON OF THE CONTROL PANEL ON THE ENTERPRISE 9000X WAS DEFECTIVE AND WHEN PRESSED, IT JAMMED AND THE BED WAS RAISING. THERE WAS NO INDICATION OF PATIENT INVOLVEMENT. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16320653 ↗MalfunctionOct 26, 2022Unintended Movement; Unintended Movement
FOLLOWING THE INFORMATION PROVIDED THE CUSTOMER REPORTED THAT THE BED WAS DISPLAYING E300 ERROR CODE AND WAS MALFUNCTIONING. AS PER THE INFORMATION PROVIDED, THE BED BACKREST SECTION WENT UP BY ITSELF AND THEN THE E300 CODE WAS DISPLAYED (MEANING THAT THE CONTROL BUTTON WAS PRESSED CONTINUOUSLY FOR OVER 90 SECONDS), DISABLING ALL BED FUNCTIONS. THERE IS NO INDICATION OF PATIENT INVOLVEMENT. NO INJURY WAS SUSTAINED. THE EVALUATION REVEALED THAT THE BACKREST CONTROL BUTTON WAS STUCK IN PUSHED POSITION, WHICH FIRST GENERATED UNINTENDED MOVEMENT AND THEN RESULTED IN E300 ERROR CODE. THE TECHNICIAN REPAIRED THE BED BY REPLACING THE SIDE RAIL WITH BUILT-IN CONTROLLERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15672283 ↗MalfunctionSep 29, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
FOLLOWING THE INFORMATION PROVIDED THE CUSTOMER COMPLAINED THAT THE UNIT NEEDED SERVICE. THE ARJO TECHNICIAN ATTENDED SITE TO EVALUATE THE BED AND DETERMINED THAT THE SIDE RAIL LOWER ARM WAS DETACHED FROM ITS MOUNTING POINTS. THERE IS NO INDICATION OF PATIENT INVOLVEMENT. NO INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15507581 ↗MalfunctionSep 14, 2022Unintended Movement; Unintended Movement
FOLLOWING THE INFORMATION PROVIDED THE CUSTOMER REPORTED INTERMITTENT FUNCTIONING OF ONE OF THE CONTROL PANEL BUTTONS ON ENTERPRISE 9000X RIGHT BODY SIDE RAIL. UPON EVALUATION THE ARJO TECHNICIAN CONFIRMED THAT THE BUTTON ON THE CONTROL PANEL WAS FAULTY AND WHEN THE BUTTON WAS RELEASED, THE BED CONTINUED TO MOVE INTERMITTENTLY. THERE IS NO INDICATION OF PATIENT INVOLVEMENT. NO INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15416705 ↗MalfunctionMay 27, 2022Unintended Movement; Unintended Movement
FOLLOWING THE INFORMATION PROVIDED THE CUSTOMER COMPLAINED THAT THE BED WAS GOING UP AND DOWN ON ITS OWN. NO PATIENT WAS INVOLVED, NO INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14520946 ↗MalfunctionApr 29, 2022Unintended Movement; Unintended Movement
FOLLOWING THE INFORMATION PROVIDED THE BED WAS INTERMITTENTLY LOWERING UNINTENDED, WHEN THE BUTTON WAS PRESSED, THE BED DID NOT STOP MOVING ONCE THE BUTTON WAS RELEASED. THE ARJO ENGINEER ATTENDED THE BED FOR EVALUATION AND REPAIRED THE BED BY REPLACING THE SAFETY SIDE WITH THE BUILT-IN CONTROLLERS. NO PATIENT WAS INVOLVED, NO INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14250314 ↗MalfunctionJul 7, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
FOLLOWING INFORMATION REPORTED BY (B)(6), THERE WAS A PROBLEM WITH ENTERPRISE 9000X. MORE DETAILS WERE NOT PROVIDED BY THE CUSTOMER. THE EVALUATION OF THE DEVICE PERFORMED BY AN ARJO TECHNICIAN REVEALED THAT THE SAFETY SIDE RAIL DETACHED FROM THE BED FRAME. THERE WAS NO INDICATION REGARDING PATIENT INVOLVEMENT. NO INJURY OR OTHER MEDICAL CONSEQUENCES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10238777 ↗InjuryJun 18, 2020Unintended Movement; Unintended Movement
ON (B)(6) 2020 ARJO WAS INFORMED THAT THE BACKREST SECTION OF THE ENTERPRISE 9000X MOVED UNINTENDED WITHOUT ANY ACTIONS GIVEN BY THE USER. THERE WAS NO PATIENT ON THE BED WHEN THIS ISSUE OCCURRED AND NO INJURY WAS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10166984 ↗InjuryApr 3, 2020Unintended Movement; Unintended Movement
FOLLOWING INFORMATION REPORTED BY (B)(6) IN THE US, THE UNINTENDED MOVEMENT OF THE BACKREST SECTION OF THE ENTERPRISE 9000X BED WAS OBSERVED. NO INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9920104 ↗
FDA clearance and classification
The FDA regulates product code FNL (bed, ac-powered adjustable hospital) under 21 CFR 880.5100 as a Class II device. 21 CFR 880.5100 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Enterprise 9611AF101BCUSB
Has the Enterprise 9611AF101BCUSB been recalled?
No FDA recall record in our data names this exact device identifier (DI 05055982755030), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Enterprise 9611AF101BCUSB?
The FDA's MAUDE database lists 32 adverse event reports for this device, received between Aug 20, 2019 and Mar 3, 2026. In those reports, the most common specific device problems are “Unintended Movement” (15), “Detachment of Device or Device Component” (11) and “Device Fell” (2); the most common patient problems are “Fall” (3) and “Head Injury” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Enterprise 9611AF101BCUSB?
None of the reports are classified as deaths, 5 as injuries and 27 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Enterprise 9611AF101BCUSB FDA approved?
Devices of this type (product code FNL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Arjohuntleigh AB.
Who makes the Enterprise 9611AF101BCUSB, and is it still sold?
The FDA's device database lists Arjohuntleigh AB as the company responsible for it (brand Enterprise, model 9611AF101BCUSB). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,831 reports
- Progressa® Bed System P7500AHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death1,320 reports
- VersaCare® Med Surg Bed P3200KHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL3 deaths1,149 reports
- Medline MDR107003EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall3 deaths333 reports
- Medline MDR107002EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death330 reports
- InTouch 2141Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths361 reports
- Citadel Plus FX811B3B4AMABBArjohuntleigh AB · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths354 reports
- TotalCare® Bed P1900QHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL299 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page